← Back
Water-soluble arene ruthenium complexes containing pyridinethiolato ligands: Synthesis, molecular structure, redox properties and anticancer activity of the cations [(η6-arene)Ru(p-SC5H4NH)3]2+
Selection of the Optimum Electrospray Voltage
for Gradient Elution LC-MS Measurements
Ioan Marginean, Ryan T. Kelly, Ronald J. Moore, David C. Prior,
Brian L. LaMarche, Keqi Tang, and Richard D. Smith
Biological Sciences Division, Pacific Northwest National Laboratory, Richland, Washington, USA
Changes in liquid composition during gradient elution liquid chromatography (LC) coupled to
mass spectrometry (MS) analyses affect the electrospray operation. To establish methodologies
for judicious selection of the electrospray voltage, we monitored in real time the effect of the
LC gradient on the spray current. The optimum range of the electrospray voltage decreased as
the concentration of organic solvent in the eluent increased during reversed-phase LC
analyses. These results and related observations provided the means to rationally select the
voltage to ensure effective electrospray operation throughout gradient-elution LC separations.
For analyses in which the electrospray was operated at constant voltage, a small run-to-run
variation in the spray current was observed, indicating a changing electric field resulting from
fouling or degradation of the emitter. Algorithms using feedback from spray current
measurements that can maintain the electrospray voltage within the optimum operating range
throughout gradient elution LC-MS were evaluated. The electrospray operation with voltage
regulation and at a constant, judiciously selected voltage during gradient elution LC-MS
measurements produced data with similar reproducibility. (J Am Soc Mass Spectrom 2009, 20,
682– 688) © 2009 Published by Elsevier Inc. on behalf of American Society for Mass Spectrometry
E
lectrospray ionization (ESI) has enabled the successful routine on-line coupling of liquid chromatography (LC) with mass spectrometry (MS) [1,
2], a combination that has become the de facto standard for proteomics research [3, 4]. Despite the
tremendous technological growth of LC-MS, the electrospray remains a seemingly robust but relatively
neglected and poorly understood intermediate. It is
expected that shifting eluent composition during
gradient LC separations may alter the electrospray
operation; however, essentially all gradient-elution
LC-MS analyses are performed with little or no
feedback with respect to electrospray performance.
During LC-MS analyses, the solvent composition
and flow rate are dictated by the separation conditions
and generally only the applied voltage and the emitter–MS inlet distance can be adjusted. Direct optimization of the analyte signal before data acquisition is not
possible, as for infusion experiments. The voltage is
usually selected according to the instrument manufacturer’s recommendations or previous experience, rather
than careful consideration of the experimental conditions. Emitter–MS inlet distance adjustments induce
large changes in the electric field driving the electrospray
because of the nonlinear dependence between the two
parameters. This may result in different electrospray current or even a different operating regime [5]. When
Address reprint requests to Dr. Richard D. Smith, Pacific Northwest
National Laboratory, Biological Systems Analysis and Mass Spectrometry,
3335 Q. Avenue (K8-98), P.O. Box 999, Richland, WA 99352. E-mail: rds@
pnl.gov
nonmetallic (e.g., fused silica) emitters are used, the voltage is usually applied upstream of the emitter via a
conductive material in contact with the solution. The
voltage drop across the emitter should also be considered
in this case [6]; applying the same voltage on emitters of
different lengths positioned at the same distance in front
of the MS inlet would also result in different electric fields.
The voltage that ensures optimum operation at the
beginning of the analysis may not be appropriate for
the solvent mixture eluting at the end of the gradient. In
previous work, the electrospray characteristic curves measured with several solvent compositions simulating a
typical reversed-phase liquid chromatography (RPLC)
gradient indicated that the voltage required to maintain
the electrospray in a specific operating regime decreases
significantly with increasing organic solvent concentration [5]. The vast majority of LC-MS analyses are
currently performed with constant electrospray voltage;
previous experience with a specific type of analysis may
help in guiding the selection of a reasonable voltage
for the beginning of the gradient that is not too high or
too low for the end of the gradient. The changing eluent
composition can be counterbalanced by post-column
combination with an inverted gradient [7]; this approach provides steady solvent composition throughout the analysis at the cost of decreased sensitivity
because of analyte dilution and increased flow rate.
Only a few attempts to rationally control the electrospray operation during LC-MS analyses have been
described. Valaskovic et al. [8] used feedback from an
electrospray imaging system to automatically adjust the
© 2009 Published by Elsevier Inc. on behalf of American Society for Mass Spectrometry.
1044-0305/09/$32.00
doi:10.1016/j.jasms.2008.12.004
Published online December 13, 2008
Received September 23, 2008
Revised December 3, 2008
Accepted December 7, 2008
J Am Soc Mass Spectrom 2009, 20, 682– 688
applied voltage and control its operating regime. Two
recently reported techniques aimed at developing feedback mechanisms based on spray current measurements [9, 10]. Staats et al. [9] sought to account for
sudden interruptions of the spray arising from emitter
clogging, air bubbles, or changes in the eluent composition by adjusting the distance between the emitter and
the MS inlet in response to changes in spray current.
Gapeev et al. [10] independently built a feedback circuit
very similar to the one we described [11] and used it to
maintain the electrospray current at a certain level by
adjusting the applied voltage. Both techniques disregarded the potential for electrospray switching between
operating regimes.
The present report provides insights into the electrospray operation during LC-MS analyses. It is shown
that changes in eluent composition during a gradient
RPLC analysis are reflected by shifting electrospray
characteristic curves. A slow drift of the current generated by an electrospray operated at constant voltage is
also demonstrated between LC runs. Two approaches
to control the applied voltage through feedback from
spray current measurements are discussed.
Experimental
Sample Preparation
Acetic acid (HOAc), trifluoroacetic acid (TFA), and
acetonitrile were purchased from Sigma–Aldrich (St.
Louis, MO, USA), and water was purified using a
Barnstead Nanopure Infinity system (Dubuque, IA,
USA). Solvents A and B consisted of 0.2% HOAc ⫹
0.05% TFA in water and 0.1% TFA in 90:10 acetonitrile:
water, respectively. The two solvents were degassed
before use by vacuum filtration followed by helium
sparging for 10 min. Proteolytic digestion of bovine
serum albumin (BSA; Pierce Biotechnology, Rockford,
IL, USA) was conducted using sequencing grade trypsin (Promega, Madison, WI, USA) according to established procedures [12]. The BSA tryptic digest was
diluted to a concentration of 0.1 g/L in purified
water and the stock solutions were refrigerated for later
use.
Optimum ES Voltage for LC-MS Measurements
683
(Phenomenex, Torrence, CA, USA) into a 60-cm-long,
360-m o.d. ⫻ 75-m i.d. fused silica capillary tubing
(Polymicro Technologies Inc., Phoenix, AZ, USA) incorporating a 0.5-m retaining screen in a 1/16-in. union
(Valco Instruments Co.) custom bored to 75 m i.d.
(Lenox Laser, Glen Arm, MD, USA). When equilibrated
at 5000 psi with 100% mobile phase A, the HPLC system
delivered a flow of close to 20 L/min, which was split
with a 30-cm-long, 20-m-i.d. fused silica tubing to
provide about 200 nL/min flow through the HPLC
column. Following a 60-min column regeneration with
solvent A and 30-min sample load time, the solvent
selection valve was switched to deliver solvent B into
the mixer, thus creating an exponential gradient.
Spray Current Measurements
A schematic of the instrument setup is presented in
Figure 1. The solution eluting from the chromatographic column was delivered through 20-m-i.d. electrospray emitters prepared by chemically etching fused
silica capillaries [15]. High voltage generated by a
Bertan 205B-03R (Hicksville, NY, USA) power supply
was applied through a previously described ammeter
[11] to the stainless steel union holding the emitter. In
this configuration the ammeter measures both the current delivered by electrospray and the current leaking
to ground through the HPLC system. To minimize the
contribution of the current leaking through the LC
column on the spray current measurements, the unions
at the two ends of the column were maintained at
similar potentials. A USB-6251 Multifunction Data
Acquisition board (National Instruments, Austin, TX,
USA) was used to monitor the ammeter output and to
remotely control the high-voltage power supply.
MS Measurements
Mass spectra were collected using a Micromass Q-TOF
Ultima mass spectrometer (Waters Corporation, Milford, MA, USA) with the standard interface replaced by
an ion funnel interface with a heated capillary inlet.
Decon2LS [16] and MultiAlign [17], open-source pro-
Reversed-Phase Liquid Chromatography
LC systems similar to those used here have previously
been described in detail [13, 14]. Briefly, solvents A and
B were loaded into a pair of 100-mL Isco 100DM syringe
pumps controlled by a series D controller (Isco, Lincoln,
NE, USA). A two-position, four-port manually operated
valve (Valco Instruments Co., Houston, TX) selected
one of the solvents to be delivered into a 2.5-mL mobile
phase dynamic mixer (fabricated in-house). The samples were injected using a two-position, six-port valve
(Valco Instruments Co.) with a 5-L loop. The reversedphase capillary HPLC column was prepared in-house
by slurry packing 3-m Jupiter C18 stationary phase
Figure 1. Instrument setup. See Marginean et al. [11] for a
detailed description of the ammeter circuit.
J Am Soc Mass Spectrom 2009, 20, 682– 688
grams developed at PNNL and freely available at
http://omics.pnnl.gov, were used for data deconvolution and alignment, respectively.
The present study uses a constant-pressure LC system
delivering a flow of about 200 nL/min, which is amenable to high-throughput, automated LC-MS analyses
[18]. At this flow rate, the electrospray characteristic
curve (see Figure 1A in Marginean et al. [5]) can be
divided into three sections, each approximating an
ohmic response to the applied voltage. At relatively low
voltages the liquid wets the outer wall of the emitter;
the current increases and the wetted area decreases
with increasing applied voltage. The slope decreases
slightly when the liquid anchors to the emitter rim; this
point will be referred to as the “anchoring point.” The
second section of the curve resembles the pulsating
regime observed at lower flow rates but does not have
the same self-regulating character. A corona discharge
most likely contributes to the current at larger applied
voltages [11], when another change of slope is visible in
the characteristic curve; the corresponding point will be
referred to as the “discharge point.” The cone-jet regime
could not be established under these experimental
conditions. The emitter length and/or the distance to
the MS inlet change the onset voltage of the characteristic curve, but not its appearance.
The appearance of the characteristic curve can be
significantly altered by emitter aging, an unavoidable
process accompanied by visual and operational clues
[19, 20]. The walls of fused silica emitters—initially
transparent—turn translucent; sometimes residue deposits become visible. Geromanos et al. [21] reported
cavitation and/or cracking of the emitters operated at
excessive field strengths. Emitter aging is generally a
slow process (days), but can be accelerated significantly
(hours) by operating the electrospray at voltages above
the discharge point. Electrospray operation at voltages
below the anchoring point can also promote the deposition of residue on the emitter outer wall; however, the
emitter aging kinetics seems to be less affected. These
physical and/or chemical processes eventually change
the wetting properties of the emitter; the liquid maintains the contact with the outer emitter wall even at
voltages above the anchoring point, swinging the electrospray to an alternative path in the parameter space.
The new path is characterized by poorer reproducibility
and noisier current measurements (not shown) because
of more erratic fluid dynamics at the tip of a wetted
emitter. During this study we tried to prolong the
emitter lifetime by avoiding the electrospray operation
at voltages above the discharge point. With a single
exception (see Figure 2), we also maintained the voltage
above the anchoring point.
The characteristic curves obtained for a blank RPLC
run, shown in Figure 2 in a color-coded representation,
form a gradient map. A base-peak chromatogram, col-
1300
1200
Voltage (V)
Results and Discussion
1400
180.0
100
90.00
75
0
50
1100
1000
Intensity (%)
MARGINEAN ET AL.
Spray Current (nA)
684
25
900
0
0
30
60
90
120
Time (minutes)
Figure 2. Electrospray characteristic curves measured during a
gradient LC-MS dry run superimposed on a base peak ion
chromatogram to illustrate the domain of analytical interest.
lected at constant voltage, is also provided to illustrate
the gradient region of analytical interest. The voltage
was scanned in steps of 10 V over a 300-V range above
the value generating a threshold spray current, which
was set at 30 nA to ensure uninterrupted electrospray
operation. During the first 100 min of the gradient
delivered by the constant-pressure LC system, the flow
rate increases from about 200 to 250 nL/min because of
decreasing eluent viscosity. The gradient map for an LC
system operated at constant flow (⬃200 nL/min) is
expected to be slightly different from that in Figure 2.
The spray current would be similar at the beginning of
the gradient, but would shift to roughly 10% smaller
values at the end of the analysis. The onset voltages are
expected to be insignificantly shifted toward lower
values; however, the discharge voltages are somewhat
smaller at 200 than at 250 nL/min [5]. Along with an
expected slight decrease in ion utilization efficiency, the
increased flow rate is expected to improve the electrospray robustness.
During the time required for the liquid to travel from
the mixer to the emitter tip (⬃15 min) no change in the
characteristic curves was observed. A significant drop
in the threshold voltage (from 1080 to 870 V) is readily
visible after 100 min. This is related to the decreasing
surface tension of the eluent as the concentration of the
organic component increases. A voltage set 300 V above
the threshold generated close to 175 nA at the beginning
of the gradient and only about 135 nA at the end of the
analysis. The decrease of the characteristic curve slope
is a result of the smaller eluent conductivity. Several
minutes after the beginning of the gradient reached the
emitter tip, the characteristic curves deviated from the
theoretical shape discussed earlier. The linear dependence between the spray current and the applied voltage was disrupted by the electrospray switching to a
multi-jet (rim emission) regime at the top of the applied
voltage domain. This regime change produced the
yellow peninsula (current ⬃135 nA) in the gradient
map.
J Am Soc Mass Spectrom 2009, 20, 682– 688
1250
150
Intensity (a.u.)
100
750
500
50
Spray Current (nA)
1000
250
0
0
-60
0
(a)
60
Time (minutes)
150
140
Spray Current (nA)
Given the changing eluent properties during the LC
gradient, operating the electrospray at constant voltage
can be problematic. For example, a voltage that delivers
abundant current at the beginning of the gradient could
later induce a corona discharge regime. Another scenario could follow the horizontal line corresponding to
an applied voltage of about 1275 V, which ensured a
current of close to 135 nA at the beginning of the
gradient. The current would initially increase as the
gradient reaches the emitter tip, then decrease as a
result of the electrospray regime shift, and then increase
again with increasing content of organic solvent in the
eluent. Because of decreasing eluent conductivity, the
current would finally decrease by the end of the gradient. Performing the analysis at a voltage that ensures a
current ⬍135 nA at the end of the gradient (⬃1150 V in
this case) seems to be the safest choice. A voltage above
the anchoring point at the beginning of the gradient that
remains below the corona discharge point at the end of
the gradient would be ideal for RPLC-MS analyses
conducted with constant electrospray voltage.
Characteristic curves measured at increasing/decreasing emitter–MS inlet distance show voltage shifts
to larger/smaller values. Similarly, shorter/longer
emitters shift the voltage range to smaller/larger values
(not shown). These conditions were maintained during
each set of measurements presented (see Figures 3 and
5); however, a new emitter was used for each set of
measurements. It is thus likely that the emitters did not
have exactly the same length and/or were positioned at
slightly different distance from the MS inlet. Consequently, voltage values are strictly comparable only
within the same set of measurements and thus voltage
offsets between sets of measurements are not surprising. The current delivered by the electrospray is a better
indicator of the electric field established at the emitter
tip and is thus a better metric to compare different sets
of measurements.
Figure 3a presents the base peak ion chromatogram
and the spray current measured during an RPLC-MS
analysis performed with the electrospray voltage set at
1250 V. The voltage was selected to maintain a spray
current ⬍140 nA throughout the gradient. In terms of
current generated by the electrospray, these operating
conditions would be equivalent to an applied voltage
of close to 1200 V in Figure 2. The first 60 min of the
spray current trace correspond to the column regeneration phase. The arrow marks the sample injection time;
the introduction of the sample loop into the circuit led
to a slightly smaller spray current. The spray current
decreased significantly when the solvent from the
sample injection loop reached the emitter (⬃13 min
later) because of the low solution conductivity. The
gradient was initiated at time 0.
Figure 3b shows spray current traces corresponding
to four consecutive RPLC-MS analyses performed as
described earlier. Run-to-run variations are visible in
this representation, especially for the black trace. We
suspect that residue accumulation on the emitter, which
685
Optimum ES Voltage for LC-MS Measurements
130
120
run 4
run 3
run 2
run 1
110
0
(b)
30
60
90
Time (minutes)
Figure 3. Base peak ion chromatogram and spray current for
LC-MS analysis performed with constant (1250 V) electrospray
voltage (a). Spray current traces for four consecutive LC-MS
analyses (b).
changes the rim geometry and alters the electric field
driving the electrospray, is responsible for this drift.
Earlier results [22] also indicated spray current decays
during long-term electrospray operation at constant
voltage as well as the need to increase the voltage for
maintaining a specific operating regime.
The spray current measurements can be used as
feedback to actively maintain the electrospray between
the anchoring and the corona discharge points. The
corresponding section of the characteristic curve is
fairly linear and can be characterized by two coefficients: the slope and the threshold voltage (V
Vt). These
parameters can be calculated by linear regression at
any point of the gradient by changing the voltage
within a limited range and measuring the corresponding current.
Considering the change in slope at the anchoring
point, clearly no theoretical significance can be associated with Vt; the practical threshold voltage ensuring
the delivery of minimum current by the electrospray is
always larger than Vt calculated as just described. The
voltage range scanned to collect the data necessary to
calculate Vt should be as small as possible to minimize
MARGINEAN ET AL.
J Am Soc Mass Spectrom 2009, 20, 682– 688
500
50
250
0
0
0
30
60
90
Time (minutes)
Figure 4. LC-MS analysis performed using feedback from spray
current measurements to control the voltage with minimal a priori
knowledge about the LC gradient (see text for details).
spray current variation that may affect the MS signal.
Also, the voltage should be considerably larger than Vt
to maintain a reasonable spray current throughout the
analysis. Consequently, the extrapolation used to calculate Vt is prone to relatively large errors. However,
based on Vt values calculated along the gradient, the
applied voltage may be adjusted to maintain the electrospray at equivalent points on the continuously shifting characteristic curves. For example, Figure 4 summarizes the results of an LC-MS run using this feedback
algorithm. During this experiment, Vt values were
calculated by scanning the voltage in 10-V steps within
a 50-V range and the limits of the voltage range were
continuously adjusted to maintain the electrospray at
least 400 V above Vt.
The voltage decreased from an average of approximately 1250 V at the beginning of the gradient to an
average of about 1000 V at the end of the gradient.
Decreasing eluent conductivity led to the spray current
declining from an average of about 120 to 80 nA during
the analysis. The voltages generating similar currents in
Figure 2 were about 1230 and 1010 V, respectively. This
approach leads to relatively large spray current fluctuations, which could be minimized if the determination
of new Vt values would be triggered only by significant
changes in the spray current or regular time intervals.
The feedback algorithm easiest to implement adjusts
the applied voltage to maintain a given spray current.
In 1972, Evans and Hendricks [23] built a feedback
circuit that regulated the voltage to avoid spontaneous
changes in the emission of a liquid metal ion source.
Gapeev et al. [10] implemented a similar approach
using a proportional-integral-derivative (PID) algorithm, but did not provide any guidance on selecting
the spray current level to be maintained. The results
presented in this report suggest a linear behavior of the
spray current as a function of the applied voltage for
our particular experimental conditions. This system
should be regulated effectively by a PID algorithm;
1250
Intensity (a.u.) & Voltage (V)
100
750
Spray Current (nA)
Intensity (a.u.) & Voltage (V)
1000
however, the electrospray characteristic curves are not
always linear [5, 24]. Even if the assumption of linearity
were correct, air bubbles—which are common in LC
systems—induce significant spray current drops, which
would lead to large voltage overshoots.
Figure 2 suggests that spray currents as high as 135
nA could be safely maintained throughout the analysis;
different solvent combinations or flow rates would
require a re-evaluation of this value. Figure 5a summarizes the results of a typical RPLC-MS analysis in which
the voltage was continuously adjusted to maintain the
current at a conservative level of 120 nA throughout the
experiment. As in Figure 3a, the first 60 min of the spray
current trace correspond to column regeneration. Immediately after sample injection (marked by an arrow),
the voltage increased to compensate for the small
current drop, after which it was maintained constant
during elution of the low-conductivity sample solvent.
The gradient was also initiated at time 0.
Figure 5b presents the voltage applied during four
consecutive RPLC-MS runs. In good agreement with
expectations from Figure 2 (a voltage drop from 1230 to
1130 V), the voltage dropped roughly 100 V by the end
150
1000
100
750
500
50
250
0
0
-60
0
(a)
60
Time (minutes)
1200
run 4
run 3
run 2
run 1
1150
Voltage (V)
150
1250
Spray Current (nA)
686
1100
1050
0
(b)
30
60
90
Time (minutes)
Figure 5. Base peak ion chromatogram, applied voltage and
spray current for LC-MS analysis performed with feedback control of the applied voltage (a). Applied voltage traces for four
consecutive LC-MS analyses (b).
J Am Soc Mass Spectrom 2009, 20, 682– 688
of the gradient (from 1160 to 1060 V). This voltage drop
was smaller than that observed with the previous
algorithm (⬃250 V) and may seem minor when compared to an applied voltage of close to 1200 V; however,
the voltage drop is significant in comparison with the
voltage range that affords the electrospray operation
between the anchoring and the corona discharge points.
To maintain a current of 120 nA, the voltage increased slightly at the end of the gradient, compensating for the lower eluent conductivity. There was no
major difference between the voltage traces during the
first two runs, but the voltage increased nearly 30 V to
generate the same current during the last two runs. As
in the case of the experiments performed at constant
voltage, we suspect that microscopic deposits on the
emitter modified its rim structure, affecting the electric
field at the tip. Running the same set of measurements
at constant voltage would result in slightly lower spray
currents during the last two runs than during the first
two runs.
There is a significant difference between the voltage
values (1230 and 1160 V) that ensured the generation of
similar spray current (120 nA) in Figures 2 and 5; this
may be related to different emitter length and/or positioning relative to the MS inlet. Assuming that operation at constant voltage was desirable for the experiments in Figure 5, a voltage of 1230 V would be selected
using Figure 2 as benchmark and without any hint of
this voltage shift. The resulting electric field would
draw about 155 nA at the beginning of the analysis,
which would be appropriate at the start of the gradient
but excessive at the end.
To evaluate the effect of the small run-to-run spray
current drift observed when the electrospray was operated at constant voltage (Figure 3b), we considered the
species detected in all four measurements to calculate
the reproducibility of the corresponding analytical signal. The coefficient of variance was slightly lower than
30%, a level roughly equivalent to that found for
label-free quantitation in LC-MS studies [25, 26]. The
measurements with feedback control (Figure 5b) resulted in a similar coefficient of variance. It may seem
surprising that compensating for the spray current drift
did not improve the reproducibility of the data. We can
interpret this observation in the light of our recent
results [11], showing that changes in ionization and
transmission efficiencies offset one another for electrosprays operated at relatively large flow rates, offering
greater robustness at the cost of smaller ion utilization
efficiency.
Conclusions
Spray current measurements can be used to monitor the
electrospray operation throughout gradient LC-MS
measurements. Even if not used as feedback for voltage
control, they can still provide a reliable quality control
mechanism and important information regarding electrospray operation by diagnosing emitter clogging, the
Optimum ES Voltage for LC-MS Measurements
687
presence of air bubbles in the eluent, and emitter aging.
Based on gradient maps, the electrospray voltage can be
rationally selected to ensure optimal operation throughout the analysis. We expect that the small current drift
observed during the four LC-MS runs at constant
voltage (Figure 3b) becomes more significant with emitter aging and would eventually alter the quality of the
data collected at constant electrospray voltage. We are
in the process of integrating the spray current monitoring and voltage selection/regulation approaches in our
automated LC-MS systems [18]. This will enable longerterm studies of electrospray behavior that will further
clarify the effect of voltage regulation on the LC-MS
data reproducibility.
We have shown that our particular experimental
conditions afford the electrospray operation at a constant but rigorously selected voltage with no negative
impact on the reproducibility of the MS signal. However, these results should not be interpreted to suggest
that an electrospray can be operated at constant voltage
without unfavorable consequences throughout any
LC-MS experiment. Selection of the electrospray voltage for the vast majority of LC-MS analyses is less
rigorous than that presented here; the voltage optimization at the beginning of the analysis often results in
poor electrospray operation at the end of the gradient.
Changing electrospray operating regimes and/or faster
emitter aging arising from wetting/corona discharges
are expected to induce larger run-to-run spray current
fluctuations, which could affect the analyte ionization
and the MS signal more significantly.
This study is most relevant for experiments in which
the electrospray characteristic curves are virtually linear. Experimental conditions that bring more definition
to the shape of the characteristic curves (different solvent systems, solvents of much lower conductivity,
lower flow rates, flow split through multi-emitter arrays [27]) are expected to benefit significantly from
more sophisticated voltage-control algorithms.
Acknowledgments
We thank Dr. Yehia M. Ibrahim and Heather M. Mottaz for providing bovine albumin samples; Dr. Eric A. Livesay, Rui Zhao, Daniel
Orton, and Dr. Kostantinos Petritis for assistance with LC-MS operation; and Dr. Christina M. Sorensen, Dr. Tyler Heibeck, Dr. Vlad A.
Petyuk, and Dr. Ashoka D. Polpitiya for sharing their experience
with LC-MS data analysis procedures. This research was supported by the NIH National Center for Research Resources
(RR-018522). Experimental portions were performed in the Environmental Molecular Sciences Laboratory, a DOE national scientific user facility located at the PNNL in Richland, Washington.
PNNL is a multiprogram national laboratory operated by Battelle
for the DOE under Contract DE-AC05-76RLO 1830.
References
1. Aleksandrov, M. L.; Gall, L. N.; Krasnov, V. N.; Nikolaev, V. I.;
Pavlenko, V. A.; Shkurov, V. A.; Baram, G. I.; Gracher, M. A.; Knorre,
V. D.; Kusner, Y. S. Direct Coupling of a Microcolumn Liquid Chromatograph and a Mass Spectrometer. Bioorg. Khim. 1984, 10, 710 –712.
688
MARGINEAN ET AL.
2. Whitehouse, C. M.; Dreyer, R. N.; Yamashita, M.; Fenn, J. B. Electrospray Interface for Liquid Chromatographs and Mass Spectrometers.
Anal. Chem. 1985, 57, 675– 679.
3. Aebersold, R.; Mann, M. Mass Spectrometry-Based Proteomics. Nature
2003, 422, 198 –207.
4. Domon, B.; Aebersold, R. Mass Spectrometry and Protein Analysis.
Science 2006, 312, 212–217.
5. Marginean, I.; Kelly, R. T.; Page, J. S.; Tang, K. Q.; Smith, R. D.
Electrospray Characteristic Curves: In Pursuit of Improved Performance in the Nanoflow Regime. Anal. Chem. 2007, 79, 8030 – 8036.
6. Jackson, G. S.; Enke, C. G. Electrical Equivalence of Electrospray
Ionization with Conducting and Nonconducting Needles. Anal. Chem.
1999, 71, 3777–3784.
7. Ramanathan, R.; Zhong, R.; Blumenkrantz, N.; Chowdhury, S. K.;
Alton, K. B. Response Normalized Liquid Chromatography Nanospray
Ionization Mass Spectrometry. J. Am. Soc. Mass Spectrom. 2007, 18,
1891–1899.
8. Valaskovic, G. A.; Murphy, J. P.; Lee, M. S. Automated Orthogonal
Control System for Electrospray Ionization. J. Am. Soc. Mass Spectrom.
2004, 15, 1201–1215.
9. Staats, S. L. T.; Fogiel, A. J.; Suna, A. Active Spray Control with Electric
Field Optimization for Online NanoLC with Polymeric Spray Tips. In
Proceedings of the 56th ASMS Conference on Mass Spectrometry, Denver,
CO, June 1–5, 2008.
10. Gapeev, A.; Berton, A.; Fabris, D. Current-Controlled Nanospray for the
Analysis of Less Than Ideal Samples. In Proceedings of the 56th ASMS
Conference on Mass Spectrometry, Denver, CO, June 1–5, 2008.
11. Marginean, I.; Kelly, R. T.; Prior, D. C.; LaMarche, B. L.; Tang, K.; Smith,
R. D. Analytical Characterization of the Electrospray Ion Source in the
Nanoflow Regime. Anal. Chem. 2008, 80, 6573– 6579.
12. Kinter, M. M.; Sherman, N. E. Protein Sequencing and Identification Using
Tandem Mass Spectrometry. New York: Wiley_Interscience, 2000.
13. Shen, Y. F.; Tolic, N.; Zhao, R.; Pasa-Tolic, L.; Li, L. J.; Berger, S. J.;
Harkewicz, R.; Anderson, G. A.; Belov, M. E.; Smith, R. D. HighThroughput Proteomics Using High Efficiency Multiple-Capillary Liquid Chromatography with On-line High-Performance ESI FTICR Mass
Spectrometry. Anal. Chem. 2001, 73, 3011–3021.
14. Shen, Y. F.; Zhao, R.; Belov, M. E.; Conrads, T. P.; Anderson, G. A.; Tang,
K. Q.; Pasa-Tolic, L.; Veenstra, T. D.; Lipton, M. S.; Udseth, H. R.; Smith,
R. D. Packed Capillary Reversed-Phase Liquid Chromatography with
High-Performance Electrospray Ionization Fourier Transform Ion Cyclotron Resonance Mass Spectrometry for Proteomics. Anal. Chem. 2001,
73, 1766 –1775.
15. Kelly, R. T.; Page, J. S.; Luo, Q. Z.; Moore, R. J.; Orton, D. J.; Tang, K. Q.;
Smith, R. D. Chemically Etched Open Tubular and Monolithic Emitters
J Am Soc Mass Spectrom 2009, 20, 682– 688
for Nanoelectrospray Ionization Mass Spectrometry. Anal. Chem. 2006,
78, 7796 –7801.
16. Jaitly, N.; Mayampurath, A.; Littlefield, K.; Adkins, J. N.; Anderson,
G. A.; Smith, R. D. Decon2LS: An Open-Source Software Package for
Automated Processing and Visualization of High Resolution Mass
Spectrometry Data. BMC Bioinformatics 2009, in press.
17. Jaitly, N.; Monroe, M. E.; Petyuk, V. A.; Clauss, T. R. W.; Adkins, J. N.;
Smith, R. D. Robust Algorithm for Alignment of Liquid Chromatography-Mass Spectrometry Analyses in an Accurate Mass and Time Tag
Data Analysis Pipeline. Anal. Chem. 2006, 78, 7397–7409.
18. Livesay, E. A.; Tang, K. Q.; Taylor, B. K.; Buschbach, M. A.; Hopkins,
D. F.; LaMarche, B. L.; Zhao, R.; Shen, Y. F.; Orton, D. J.; Moore, R. J.;
Kelly, R. T.; Udseth, H. R.; Smith, R. D. Fully Automated Four-Column
Capillary LC-MS System for Maximizing Throughput in Proteomic
Analyses. Anal. Chem. 2008, 80, 294 –302.
19. Nilsson, S.; Svedberg, M.; Pettersson, J.; Bjorefors, F.; Markides, K.;
Nyholm, L. Evaluations of the Stability of Sheathless Electrospray
Ionization Mass Spectrometry Emitters Using Electrochemical Techniques. Anal. Chem. 2001, 73, 4607– 4616.
20. Chen, M. L.; Cook, K. D. Oxidation Artifacts in the Electrospray Mass
Spectrometry of A  Peptide. Anal. Chem. 2007, 79, 2031–2036.
21. Geromanos, S.; Freckleton, G.; Tempst, P. Tuning of an Electrospray
Ionization Source for Maximum Peptide-Ion Transmission into a Mass
Spectrometer. Anal. Chem. 2000, 72, 777–790.
22. Marginean, I.; Kelly, R. T.; Page, J. S.; Tang, K. Q.; Smith, R. D. Toward
a Stable Electrospray. In Proceedings of the 55th ASMS Conference on Mass
Spectrometry, Indianapolis, IN, June 3–7, 2007.
23. Evans, C. A.; Hendricks, C. D. Electrohydrodynamic Ion-Source for
Mass-Spectrometry of Liquids. Rev. Sci. Instrum. 1972, 43, 1527–1530.
24. Chen, D. R.; Pui, D. Y. H.; Kaufman, S. L. Electrospraying of Conducting
Liquids for Monodisperse Aerosol Generation in the 4 nm to 1.8 m
Diameter Range. J. Aerosol Sci. 1995, 26, 963–977.
25. Old, W. M.; Meyer-Arendt, K.; Aveline-Wolf, L.; Pierce, K. G.; Mendoza,
A.; Sevinsky, J. R.; Resing, K. A.; Ahn, N. G. Comparison of Label-Free
Methods for Quantifying Human Proteins by Shotgun Proteomics. Mol.
Cell. Proteomics 2005, 4, 1487–1502.
26. Ono, M.; Shitashige, M.; Honda, K.; Isobe, T.; Kuwabara, H.; Matsuzuki,
H.; Hirohashi, S.; Yamada, T. Label-Free Quantitative Proteomics Using
Large Peptide Data Sets Generated by Nanoflow Liquid Chromatography and Mass Spectrometry. Mol. Cell. Proteomics 2006, 5, 1338 –1347.
27. Kelly, R. T.; Page, J. S.; Marginean, I.; Tang, K. Q.; Smith, R. D.
Nanoelectrospray Emitter Arrays Providing Interemitter Electric Field
Uniformity. Anal. Chem. 2008, 80, 5660 –5665.