CHARACTERISATION OF THE
PERFORMANCE OF AN
ABSORBABLE MAGNESIUM STENT
BY EXPERIMENTAL AND
NUMERICAL ANALYSIS
A thesis submitted for the award of PhD
Emmet Galvin, BEng MSc
Supervisors:
Dr. Bryan MacDonald
Dr. Caítriona Lally
School of Mechanical and Manufacturing Engineering
Dublin City University
September, 2014
Declaration
I hereby certify that this material, which I now submit for assessment on the
programme of study leading to the award of Doctor of Philosophy (PhD) is entirely
my own work, that I have exercised reasonable care to ensure that the work is
original, and does not to the best of my knowledge breach any law of copyright, and
has not been taken from the work of others save and to the extent that such work
has been cited and acknowledged within the text of my work.
Signed: ____________________________
Student ID No.: 98573900
Date: 18th September, 2014
i
Acknowledgements
I would like to thank my supervisors Dr. Bryan MacDonald and Dr. Tríona Lally for
all their guidance and support throughout the project. This research could not have
been achieved without the help of Christy Cummins at Vasorum Ltd. I would also
like to thank Liam, Jim, Chris, Michael, and Keith for their technical support. I would
especially like to thank my family and friends for their support and encouragement.
This study was kindly funded by the Irish Research Council for Science
Engineering and Technology (IRCSET) under the postgraduate research awards
scheme, funded by the Irish Government’s National Development Plan. Scholarship
Reference Number: RS/2006/82.
ii
Publications and Presentations Resulting from this
Study
Papers
Galvin, E., Cummins, C., Yoshihara, S., MacDonald, B., Lally, C. 2014.
Experimental testing of an absorbable magnesium stent demonstrates the critical
influence of deployment strain on corrosion behaviour. Journal of the Mechanical
Behaviour of Biomedical Materials, In Review.
Galvin, E., Morshed, M.M., Cummins, C., Daniels, S., Lally, C., MacDonald, B.J.,
2013. Surface Modification of Absorbable Magnesium Stents by Reactive Ion
Etching. Journal of Plasma Chemistry and Processing. 33(6), 1137-1152
Conferences and Abstracts
Galvin E., Cummins, C., Colombo, A., Lally, C., MacDonald, B.J., “In Vitro Corrosion
Testing of Absorbable Magnesium Stents”, Proceedings of the 28th International
Manufacturing Conference, Dublin City University, September 2011.
Galvin, E., Cummins, C, Coleman, J., Mathews, D., Colombo, A., Lally, C.,
MacDonald, B.J., “An Experimental and Numerical Analysis of a Bioabsorbable
Magnesium Alloy Cardiovascular Stent”, Proceedings of the 26th International
Manufacturing Conference, Trinity College Dublin, September 2009.
Galvin, E., MacDonald, B.J., “Explicit & Implicit Finite Element Analysis of a
Bioabsorbable Magnesium Alloy Cardiovascular Stent”, Proceedings of the Northern
Ireland Biomedical Engineering Society Spring Conference, April 2009.
Galvin, E., MacDonald, B.J., “Experimental and Numerical Analysis of a
Bioabsorbable Magnesium Alloy Cardiovascular Stent”, Bioengineering In Ireland 15
Conference, Castletroy Park Hotel, Limerick, January 2009.
Galvin, E., MacDonald, B.J., “Finite Element Simulation of the Deployment of a
Magnesium
Cardiovascular
Stent”,
Proceedings
of
the
25th
International
Manufacturing Conference, Dublin Institute of Technology, September 2008.
iii
List of Figures
Figure 1.1 Flow chart of the key project steps ...........................................................7
Figure 2.1 (a) Anterior view of the heart showing the coronary circulation (b)
schematic of the layers of a coronary artery..............................................................8
Figure 2.2 Schematic representation of the process of atherosclerosis...................11
Figure 2.3 Stent restenosis: the re-blockage of a stented artery owing primarily to
the formation of neointimal hyperplasia ...................................................................12
Figure 2.4 Schematic representation of the restenosis process after stent
implantation ............................................................................................................12
Figure 2.5 Influence of stenting on the cyclic strain environment in the local coronary
artery ......................................................................................................................15
Figure 2.6 Details of current bioabsorbable stents (BAS) platforms (adapted from
(Zhang et al., 2013)) ...............................................................................................20
Figure 2.7 Magnesium coronary stent platforms manufactured by Biotronik AG: (a)
AMS-1, (b) DREAMS-1 and (c) DREAMS-2 ............................................................21
Figure 2.8 SEM images and EDX mapping of DREAMS-2 degradation products after
28, 90 and 180 days implantation in porcine coronary arteries (Wittchow et al., 2013)
...............................................................................................................................22
Figure 2.9 Phase diagrams (a) Mg-Y (b) Mg-Zr (Li et al., 2013; Zhao et al., 2011)..27
Figure 2.10 Backscattered scanning electron micrograph of wrought WE43 alloy
(Coy et al., 2010) ....................................................................................................28
Figure 2.11 (a) Galvanic corrosion (internal) in magnesium alloys. (b) Galvanic
activity among the primary α, eutectic α and β-phases and (c) the resulting
morphology. (e) a continuous β-phase inhibits corrosion and (f) the resulting
morphology .............................................................................................................32
Figure 3.1 (a) Two-dimensional CAD model of the unfurled stent geometry (b) SEM
image of a tube in which the stent pattern was machined .......................................45
Figure 3.2 (a) Machined stent remaining after the inner sections had fallen out (b)
dross that formed on the inner surfaces as a result of laser cutting process ...........46
Figure 3.3 Schematic diagram of the experimental setup used for the plasma
etching process.......................................................................................................47
iv
Figure 3.4 Free expansion of stents (a) Location of 8 stent OD measurements taken
using a videoextensometer during bench testing (b) 316L stent..............................51
Figure 3.5 Stent expansion in mock coronary arteries (a) WE43 stent (b) 316L stent
...............................................................................................................................52
Figure 3.6 Stented mock coronary artery under the influence of static pressure head
during the experiment performed to measure cyclic strain amplitude ......................52
Figure 3.7 (a) The dogbone specimen geometry (b) Representative stress-strain
response of dogbone specimens during the application of plastic strain .................55
Figure 3.8 Simplified flowchart of the corrosion model process ...............................59
Figure 3.9 The possible values for the element exposure parameter, κe, that were
used to scale the element damage calculated at each time step.............................60
Figure 3.10 (a) slice of dogbone specimen test region exhibiting repetitive and
reflective symmetry (b) boundary and loading conditions used for the FE simulations
...............................................................................................................................62
Figure 3.11 FE material model input used to represent the stress-strain properties of
elements with D values ranging from 0 to 0.9 .........................................................63
Figure 3.12 (a) Image of the full stent geometry with a repeating unit cell (RUC)
indicated (b) The geometry of the FE RUC model prior to deployment (c) Loading
applied to contact elements in order to predict the radial stiffness of the stent ........66
Figure 3.13 (a) Distribution of the elemental εp1 values for the stent model (b) Linear
and non-linear models used to represent the dependency of φe on elemental εp1 ...67
Figure 4.1 Stent surfaces resulting from chemical etching: (a) outer stent surface
with residue (dark areas), and (b) inner surface of stent strut..................................69
Figure 4.2 SEM images of a finished stent following plasma etching and cleaning:
(a) stent mounted on fused silica tube, (b) inner surface, and (c) outer surface ......71
Figure 4.3 Image showing a finished WE43 stent ...................................................71
Figure 4.4 (a) Typical surface morphology of as-received specimen showing spectra
sites S1-S6, and (b) average chemical composition at spectra sites S1-S3 and
spectra sites S3-S6 .................................................................................................72
Figure 4.5 Typical surface morphology of chemically etched specimen: (a) normal
surface appearance, (b) dark area with surface residue, and (c) chemical
composition of surface at spectra sites S1-S4.........................................................74
v
Figure 4.6 (a) Typical surface morphology of plasma etched specimen showing EDX
spectra sites S1-S6, and (b) average chemical composition of surface at spectra
sites S1-S6 .............................................................................................................75
Figure 4.7 Typical surface morphology of specimens generated by AFM: (a) asreceived, (b) chemically etched, and (c) plasma etched ..........................................76
Figure 4.8 Surface roughness (Sa) values (mean ± standard deviation) measured by
AFM for as-received, chemically etched and plasma etched tube specimens .........76
Figure 4.9 SEM images of tube specimens: (a) as-received, (b) chemically etched,
and (c) plasma etched and cleaned (magnification = 1000x) ..................................77
Figure 4.10 (a) Pressure-diameter curves for freely expanded stents. Inset: WE43
stent shown unexpanded, fully expanded and after recoil, (b) performance metrics
for the free expansion and expansion in MCA of stents, (c) WE43 stent (top) and
316L stent (bottom) and (d) cyclic strain amplitude results. Error bars represent one
standard deviation of uncertainty. ...........................................................................79
Figure 4.11 Expansion testing of WE43 stent: (a) unexpanded, (b) fully expanded (c)
after recoil (d) SEM image of a strut section from an expanded stent .....................80
Figure 4.12 Plot of percentage mass lost over time by dogbone specimens based on
the results of experiment A. Error bars represent a single standard deviation from
the mean.................................................................................................................81
Figure 4.13 SEM images of dogbone specimens with 0% PS following immersion in
solution for (a) 0 d, (b) 1 d, (c) 3 d, (d) 3 d (e) 10 d, (f) 10 d and, (g) 10 d ...............82
Figure 4.14 SEM images of dogbone specimens with 2.7% PS following immersion
in solution: (a) 1 d, (b) 6 d, (c) 10 d, (d) 10 d and, (e) 10 d (8.4%PS) .....................83
Figure 4.15 Representative stress-strain curves for dogbone specimens based on
the results of experiment A, (a) 0% PS (b) 2.7% PS and (c) 8.4% PS. Mechanical
properties of dogbone specimens over time: (d) UTS, (e) εu, (f) E and (g) σy. Error
bars represent one standard deviation of uncertainty..............................................85
Figure 4.16 SEM images of typical fractured dogbone specimens after tensile
testing: (a) 0% PS and 0 d (i.e. no corrosion), (b) 0% PS and 10 d (c) 2.7% PS and
10 d, (d) 8.4% PS and 10 d immersion ...................................................................86
Figure 4.17 Typical morphology of the fracture surfaces of dogbone specimens after
tensile testing (a) 0% PS and 0 d – dimple rupture, (b) 0% PS and 10 d – corrosion
pits with ductile features on non-corroded cross section (c) 2.7% PS and 10 d –
vi
predominant intergranular cracking with brittle features and some transgranular
(arrow) cracking, (d) 8.4% PS and 10 d corrosion - predominant intergranular
cracking with brittle features and some transgranular (arrow) cracking ...................87
Figure 4.18 Results plots for experiment B (a) Percentage mass lost over time for
stent, (b) Mass loss per unit area for stent and dogbone specimens (c) Plot of
average strut width over time for stent. Error bars represent a single standard
deviation from the mean .........................................................................................88
Figure 4.19 Typical morphology of a stent section after an immersion time of (a) 3 h,
(b) 20 h, (c) 40 h & (d) 60 h. (e) Representative force versus compression curves for
the WE43 and 316L stent (error bars omitted for clarity), (f) Temporal radial stiffness
of WE43 stent. Error bars represent a single standard deviation from the mean .....89
Figure 4.20 Damage parameter (D) values predicted by the corrosion model at 48 h
intervals for a specimen with 0% plastic strain ........................................................90
Figure 4.21 Predicted mass loss from the calibrated corrosion model compared to
that observed experimentally ..................................................................................91
Figure 4.22 Predicted specimen stress-strain response over time for the plastic
strain corrosion model compared to that observed experimentally for plastic strains
of (a) 0%, (b) 2.7% and (c) 8.4%.............................................................................91
Figure 4.23 Predicted specimen stress-strain response over time for the uniform
corrosion model compared to that observed experimentally for plastic strains of (a)
0%, (b) 2.7% and (c) 8.4% ......................................................................................92
Figure 4.24 Simulated deployment of the RUC stent model: (a) fully expanded (b)
predicted εp1 post recoil ...........................................................................................93
Figure 4.25 Predicted corrosion damage (D) after an immersion time of 10 hours (a)
Plastic strain corrosion model (b) Linear PS corrosion model (c) Uniform corrosion
model......................................................................................................................93
Figure 4.26 Predicted mass loss for the stent over time ..........................................94
Figure 4.27 (a) Predicted stent radial stiffness at 10% compression (b) Predicted
stent radial force versus percentage compression over time ...................................95
Figure 4.28 (a) Predicted stent mass loss for different values of stent expansion ratio
(SER) (b) Predicted stent radial stiffness at 10% compression for different values of
SER ........................................................................................................................95
Figure 5.1 (a) Typical stress-strain response for a specimen induced with 2.7%
vii
plastic strain (curve 1) and, the subsequent response (inset) for the same specimen
prior to corrosion (curve 2) and after corrosion (curve 3), (b) Mass loss curve (green)
that is driven by plastic strain alone ......................................................................108
Figure A.1 Stress-strain curves (terminating at the UTS) for dogbone specimens
based on the results of experiment A, (a) 0% PS (b) 2.7% PS and (c) 8.4% PS.
Error bars represent 95% confidence intervals derived from T-table. ....................134
Figure A.2 (a) Plot of predicted dogbone specimen mass loss over time for varying
average element size (Le) values, (b) Plot of predicted dogbone specimen mass loss
over time for varying time step (∆t) values.............................................................141
Figure A.3 (a) Plot of predicted stent mass loss over time for varying average
element size (Le) values, (b) Plot of predicted stent mass loss over time for varying
time step (∆t) values .............................................................................................142
List of Tables
Table 2.1 Available data for the mechanical properties of WE43 specimens...........29
Table 3.1 Details of the experiments performed (PCM: phase contrast microscopy)
...............................................................................................................................53
Table 3.2 The table of values established in experiment A from which the strain
parameter, φe, was calculated in the FE corrosion model .......................................64
Table 4.1 Optimised laser machining parameters for cutting WE43 tube ................68
Table 4.2 Parameters used in the corrosion models in this work.............................90
Table A.1 Elemental composition (average weight %) of WE43 tube specimen
determined by ICP-MS..........................................................................................134
viii
Nomenclature
General
AFM
Atomic force microscopy
BAS
Bioabsorbable stent
BMS
Bare-metal stent
CABG
Coronary artery bypass graft
CAD
Coronary artery disease
CDM
Continuum damage modelling
CTO
Chronic total occlusion
DB
Dogboning
DES
Drug-eluting stent
EC
Endothelial cell
EDX
Energy dispersive x-ray spectrometry
ELR
Elastic longitudinal recoil
ERR
Elastic radial recoil
FE/FEM
Finite element/Finite element method
FS
Foreshortening
IH
Intimal hyperplasia
ISR
In-stent restenosis
MCA
Mock coronary artery
MI
Myocardial infarction
PCI
Percutaneous coronary intervention
PS
Plastic strain
PTCA
Percutaneous transluminal coronary angioplasty
RE
Rare-earth
RIE
Reactive ion etching
SBF
Simulated body fluid
SEM
Scanning electron microscope
SER
Stent expansion ratio
SMC
Smooth muscle cell
ST/LST
Stent thrombosis/Late stent thrombosis
TVR
Target vessel revascularisation
ix
Material Properties
Unit
E
Young’s modulus
GPa
UTS
Ultimate tensile strength
MPa
εf
Failure strain
-
εu
Strain at UTS
-
σy
Yield stress (0.2% offset)
MPa
Corrosion Modelling
Unit
ߜ
Material characteristic length
mm
ߢ
Element exposure parameter
-
D
Damage parameter
-
ku
Corrosion kinetic parameter
h-1
Le
FE characteristic length
mm
t
Simulation time
h
tmax
Maximum solution time
h
∆t
Time-step
h
εp1
First principal plastic strain
-
φe
Plastic strain parameter
-
x
Contents
Declaration .......................................................................................................... i
Acknowledgements ........................................................................................... ii
Publications and Presentations Resulting from this Study ........................... iii
List of Figures ................................................................................................... iv
List of Tables................................................................................................... viii
Nomenclature .................................................................................................... ix
Abstract .............................................................................................................. 1
1.
2.
Introduction ............................................................................................... 2
1.1
Coronary Artery Disease and Stenting................................................ 2
1.2
Bioabsorbable Stents ......................................................................... 3
1.3
Objectives of this Study ...................................................................... 6
Literature Review ...................................................................................... 8
2.1
Anatomy, Physiology and Biomechanics of Coronary Arteries............ 8
2.2
Coronary Artery Disease .................................................................. 10
2.2.1 Restenosis ............................................................................... 12
2.3
Coronary Stents ............................................................................... 16
2.3.1 Bare-Metal Stents .................................................................... 16
2.3.2 Drug-Eluting Stents.................................................................. 17
2.3.3 Bioabsorbable Stents............................................................... 17
2.4
Magnesium and Magnesium Alloys .................................................. 24
2.4.1 The WE43 Alloy ....................................................................... 26
2.4.2 Mechanical Properties of Magnesium Alloys ............................ 28
2.4.3 Corrosion of Magnesium Alloys ............................................... 29
2.4.4 Influence of Stent Manufacture on Magnesium Alloys .............. 35
2.5
Computational Modelling of Magnesium Stents ................................ 37
2.5.1 Corrosion Modelling ................................................................. 38
2.6
3.
Summary .......................................................................................... 41
Materials and Methods ............................................................................ 44
3.1
Summary .......................................................................................... 44
3.2
Stent Manufacture ............................................................................ 45
3.2.1 Laser Cutting ........................................................................... 45
3.2.2 Chemical Etching ..................................................................... 46
xi
3.2.3 Plasma Etching ........................................................................ 47
3.2.4 Surface Characterisation ......................................................... 48
3.3
Mechanical Testing........................................................................... 49
Test 1: Free Expansion .................................................................... 51
Test 2: Expansion in Mock Coronary Arteries ................................... 51
Test 3: Cyclic Strain Amplitude ......................................................... 52
3.4
Corrosion Experiments ..................................................................... 53
3.5
Corrosion Modelling.......................................................................... 57
3.5.1 Corrosion Model Calibration..................................................... 61
3.5.2 Stent Model ............................................................................. 65
4.
Results ..................................................................................................... 68
4.1
Stent Manufacture ............................................................................ 68
4.1.1 Laser Cutting ........................................................................... 68
4.1.2 Chemical and Plasma Etching ................................................. 69
4.1.3 Surface Characterisation ......................................................... 72
4.2
Mechanical Testing........................................................................... 78
4.3
Corrosion Experiments ..................................................................... 80
4.4
Corrosion Modelling.......................................................................... 89
4.4.1 Corrosion Model Calibration..................................................... 89
4.4.2 Stent Model ............................................................................. 92
5.
6.
Discussion ............................................................................................... 96
5.1
Stent Manufacture ............................................................................ 96
5.2
Mechanical Testing........................................................................... 97
5.3
Corrosion Experiments ................................................................... 100
5.4
Corrosion Modelling........................................................................ 106
5.5
Summary and Limitations ............................................................... 109
Conclusions .......................................................................................... 111
6.1
Main Findings ................................................................................. 111
6.2
Future Work.................................................................................... 114
References ..................................................................................................... 116
Appendix A..................................................................................................... 134
A.1 Compositional analysis ..................................................................... 134
A.2 Additional Data from Corrosion Experiment A ................................... 134
A.3 APDL Implementation of Corrosion Model ........................................ 135
A.4 Corrosion Model Convergence Studies ............................................. 141
xii
Abstract
Characterisation of the Performance of an Absorbable
Magnesium Stent by Experimental and Numerical Analysis
Emmet Galvin
Magnesium stents are a promising candidate in the emerging field of bioabsorbable
stents and could obviate many of the limitations of current permanent stent
platforms such as stent thrombosis and in-stent restenosis. In this study, viable
magnesium stents were manufactured from WE43 magnesium alloy precursor tubes
by a combination of laser cutting and chemical and plasma-based surface
treatments. The mechanical and corrosion performance of the stent was elucidated
via experiments. It was shown that the plastic strain that is induced in the stent
struts during the deployment procedure has a critical influence in directing both the
kinetics and the nature of subsequent corrosion processes in the alloy. The
deployment and scaffolding characteristics of the stent were determined and
contrasted with those of a commercial stainless steel stent. The magnesium stent
was shown to support significantly higher levels of cyclic strain amplitude which
plays a crucial role in reducing in-stent restenosis. This study provided new insights
into the performance of a current magnesium stent design and material and
demonstrated the influence of deployment induced corrosion on the long-term
scaffolding ability of the stent. A finite element based phenomenological corrosion
model was developed and calibrated based on the results of the corrosion
experiments. A plastic strain parameter was introduced in the modelling framework
to account for the effects of plastic strain on corrosion behaviour. The model was
found to be capable of predicting the experimentally observed mass loss profile and
the temporal corrosion-induced reduction in the radial stiffness of the stent.
1
1. Introduction
1.1
Coronary Artery Disease and Stenting
Coronary artery disease (CAD) is the largest cause of mortality in the western world
and is responsible for over seven million deaths worldwide each year (Go et al.,
2013). The pathogenesis of CAD involves the formation of stenotic lesions (plaque)
in the vessel walls leading to narrowing and hardening of the coronary arteries
(Yutani et al., 1999). These lesions may decrease the lumen of the artery and the
resulting restriction in blood supply (ischemia) can lead to myocardial infarction (MI).
The treatment of CAD often involves the deployment of an endovascular
stent at the stenosis in order to restore patency and prevent arterial recoil (Lally et
al., 2006). However, stenting inflicts injury in the intimal and medial layer of the
arterial wall which can induce a partial re-blocking of the artery known as restenosis
(Scott, 2006). In-stent restenosis (ISR) results in the formation of intimal hyperplasia
(IH) around the stent struts and represents a major limitation to the success of
current stent procedures (Dangas et al., 2010). Bare-metal stents (BMS), which
have typically been produced from corrosion-resistant materials such as stainless
steel (316L) and cobalt chromium, are associated with complications such as
elevated rates of ISR (~25%) and late stent thrombosis (LST) (<1%) (Kandzari et al.,
2013; Stone et al., 2004).
In order to overcome these limitations, drug-eluting stents (DES) were
developed. DES normally consist of a BMS platform with a drug-eluting coating that
is intended to reduce IH (Gogas et al., 2014). Although DES have demonstrated
lower rates of ISR (~3%) and target vessel revascularisation (TVR) (<10%)
compared to BMS, stent thrombosis (ST) (<1%) and ISR remain a significant
limitation (Dangas et al., 2013; Kandzari et al., 2013). Permanent BMS and DES
have other major limitations such as restrictions to non-invasive imaging and future
2
revascularisation procedures, long term disruption of native haemodynamics and
vasomotor function, chronic inflammation and delayed endothelialisation (Sammel et
al., 2013). DES patients require longer dual antiplatelet therapy (DAPT) courses (~3
months) compared to BMS (~1 month) in order to minimise the risk of LST (Kim et
al., 2009; Nishinari et al., 2013). Variations in ST, TVR and ISR outcomes have
been observed among existing stent platforms but the optimum treatment strategy
after stenting has yet to be determined (Dangas et al., 2010; Hoffmann et al., 2004).
Hence, a new generation of bioabsorbable stents (BAS), which are gradually
absorbed in the body, is attracting significant interest (Gogas et al., 2012; Haude et
al., 2013; Nishio et al., 2012; Zhang et al., 2013).
1.2
Bioabsorbable Stents
There are a number of limitations associated with current BAS including relatively
large strut thicknesses (~150 µm) and crossing profiles (~1.4 mm) (Gogas et al.,
2014), limited extensibility and a higher risk of strut fracture (Zhang et al., 2013).
Several polymer BAS have been proposed (Gogas et al., 2012; Nishio et al., 2012)
but they have yet to achieve the mechanical and surface properties of current BMS
and DES (Jabara et al., 2009). Other concerns with polymers include toxicity and
endarteritis (Palmerini et al., 2014). Compared to polymers, metals generally have
superior mechanical properties for coronary stents. The potential of iron (Moravej et
al., 2010a, 2010b; Peuster et al., 2006, 2001) and iron-manganese alloys
(Hermawan et al., 2010a, 2008) has been studied, though the corrosion rate of iron
appears too slow for coronary BAS (Peuster et al., 2006, 2001) and unwanted
degradation products are retained in the arterial wall (Pierson et al., 2011).
Recently, absorbable magnesium stents based on the WE43 alloy have
shown promise in clinical trials (Erbel et al., 2007; Haude et al., 2013) and, could
potentially obviate many of the limitations of traditional permanent stents. However,
3
limitations associated with current magnesium stents include relatively high rates of
late lumen loss due to excessive IH formation and premature loss of scaffolding
(Erbel et al., 2007; Haude et al., 2013). Accordingly, the widespread adoption of
magnesium stents depends on further developments in the technology such as
characterisation and optimisation of materials and stent designs, and improved
understanding of the adsorption of magnesium in vascular tissues (Hermawan et al.,
2010; Kitabata et al., 2014; Zhang et al., 2013).
To date, stents based on WE43 derivatives represent the only magnesium
platform to go through clinical trials for coronary stents (Haude et al. 2013; Erbel et
al. 2007). Despite this, only a limited amount of data regarding their mechanical and
corrosion performance characteristics has been reported in animal (Di Mario et al.,
2004; Heublein et al., 2003; Waksman et al., 2006) and clinical (Bosiers et al., 2009;
Erbel et al., 2007; Haude et al., 2013) trials. There is little information available in
the literature with regard to the manufacture of magnesium stents (Demir et al.,
2012; Di Mario et al., 2004; Hassel et al., 2006; Wu et al., 2012). Fibre laser cutting
and surface treatment of balloon-expandable AZ31 stents has been described
(Demir et al., 2012; Wu et al., 2012); though no such study has been reported for
WE43. Also, few studies have reported mechanical and corrosion performance data
for WE43 precursor tubes (Pachla et al., 2012; Witte et al., 2008).
When investigating the performance of metallic stents, it is desirable to
conduct testing on the stent or on a sample of the precursor tube in the final form
and finish, due to the influence of materials processing, size effects and surface
treatments on mechanical and corrosion behaviour (Connolley et al., 2005; Jain et
al., 2008). It is noted that few experimental studies have been reported to this end
for magnesium stents, with the study of Wu et al. (Wu et al., 2012), on an AZ31
stent, being the most pertinent to date. Many of the existing studies employ
specimens that were prepared using materials processing and surface treatments
4
that differed from those of coronary stents (Bowen et al., 2013; Coy et al., 2010;
Grogan et al., 2011; Kannan et al., 2008; Zhang et al., 2010). Furthermore, while a
number of studies have examined the effects of mechanical loading on the corrosion
of magnesium for near-constant (Grogan et al., 2011) and non-constant (Kannan et
al., 2008; Winzer et al., 2008) loads, the influence of the loading that is induced
during stent deployment on corrosion behaviour has not been extensively studied.
A number of experimental studies have provided correlations between the
magnitude of specimen stress and corrosion behaviour in magnesium alloys (Dietzel
et al., 2010; Kannan et al., 2008; Winzer et al., 2008). However, plastic strain, and
the accompanying deformation-induced changes in electrochemical behaviour have
been shown to play a crucial role in the micro-galvanic corrosion processes that are
observed in magnesium alloys (Snir et al., 2012; Wittke et al., 2014). Since the stent
deployment procedure is known to induce large plastic strain in the struts (Murphy et
al., 2003), it is hypothesised in this work that deployment-induced plastic strain plays
a critical role in directing subsequent corrosion phenomena in magnesium stents
and thus effects the scaffolding ability of these devices in the long-term.
In addition to experimental corrosion testing, computational modelling of the
corrosion process can provide important insights into the fundamental corrosion
processes that lead to the degradation of magnesium stents. To date, only a small
number of corrosion models have been developed for absorbable metallic stents,
which were based on the finite element method (FEM) and used either a continuum
damage mechanics (CDM) approach (Gastaldi et al., 2011; Grogan et al., 2011) or
adaptive meshing (Grogan et al., 2014a). Typically, these models adopted a stressmediated approach to simulate the corrosion of magnesium stent geometries
(Gastaldi et al., 2011; Grogan et al., 2014a). However, in this work, a plastic strain
mediated phenomenological corrosion model of the corrosion behaviour of a
magnesium alloy stent was developed to facilitate prediction of stent behaviour.
5
While previous studies involved large and computationally-expensive FE models
that required high-performance computing capabilities (Gastaldi et al., 2011; Grogan
et al., 2014a), the focus of this work was the development of a compact and
computationally-efficient corrosion model that facilitated virtual testing of magnesium
stents and other implants on a standard desktop computer. Such models provide
short solution times which are especially useful in the preliminary stages of stent
design where many cycles of design iteration are necessary (Xia et al., 2007).
These models may find utility in the design of stents through device evaluation
simulations and the optimisation of device geometries for prolonging structural
integrity during the absorption process.
1.3
Objectives of this Study
The key aim of this study was the determination of the mechanical and corrosion
performance of a WE43 alloy absorbable magnesium stent using a combination of
experimental and numerical analysis. The mechanical performance of the stent was
determined and compared to that of a commercial 316L stent in terms of a range of
deployment, radial stiffness and cyclic strain amplitude metrics. The influence of
deployment-induced plastic strain on the subsequent corrosion behaviour of the
stent was elucidated. The corrosion rates, the dominant corrosion processes and
the temporal corrosion-induced reduction in the mechanical integrity of the stent
were revealed. Since the specimens were laser cut from precursor tubes and were
prepared using surface treatments suited to processing magnesium stents, it was
believed that such an approach provided a more accurate characterisation of the
mechanical and corrosion behaviour of the stent than previous experimental studies
on magnesium test specimens with processing histories, surface treatments and
loading conditions that differed from those of coronary stents.
A finite element (FE) based plastic strain mediated phenomenological
6
corrosion model, which was capable of modelling the corrosion of a threedimensional (3-D) stent structure, was developed in order to assess the
performance of a corroding magnesium stent. The development of such a model,
which was calibrated based on the results of corrosion experiments and solved on a
standard desktop computer by means of a computationally-efficient algorithm,
provided a fast, convenient and inexpensive preclinical tool for testing the design of
stents and other absorbable metallic implants.
In order to achieve the aforementioned goals, the following tasks first needed
to be completed (see Figure 1.1);
•
A suitable processing route had to be identified by which viable stents could
be manufactured from precursor tubes of the WE43 alloy.
•
A range of experiments had to be developed so that the mechanical and
corrosion performance of stents and other specimens could be determined.
•
A corrosion algorithm had to be developed in a suitable computer language
for use in a commercial implicit FE software package.
Figure 1.1 Flow chart of the key project steps
7
2. Literature Review
2.1 Anatomy, Physiology and Biomechanics of Coronary
Arteries
The coronary circulation supplies oxygenated blood to the myocardium with the
majority of flow occurring in diastole, when the heart is relaxed (Nichols et al., 2005).
The left and right coronary arteries (LCA and RCA) originate at the base of the
ascending aorta, at the aortic sinuses, where blood pressure is at its greatest in the
systemic circuit (Martini, 2006), as shown in Figure 2.1(a).
Figure 2.1 (a) Anterior view of the heart showing the coronary circulation (b)
schematic of the layers of a coronary artery
The RCA supplies blood mainly to the right atrium, portions of both ventricles
and the conducting system of the heart including the sinoatrial and atrioventricular
nodes. The LCA supplies the left ventricle, left atrium and interventricular septum.
Towards the anterior surface of the heart it gives rise to a circumflex branch (LCX)
and the left anterior descending (LAD) artery. The LCX curves around the coronary
sulcus where it finally joins small branches of the RCA. The much larger LAD swings
8
around the pulmonary trunk and runs along the surface within the interventricular
sulcus. The LAD supplies small tributaries continuous with those of the posterior
descending artery through interconnections known as arterial anastomoses (Martini,
2006). LAD arteries are of central importance in CAD pathophysiology as ISR after
stenting is most common in the LAD (Holzapfel et al., 2005).
The wall of a healthy coronary artery consists of three distinct concentric
layers; intima, media, and adventitia, as shown in Figure 2.1(b). The intima is the
innermost layer and consists of a single layer of endothelial cells (ECs) that grow on
the internal elastic lamina (Humphrey, 2002). ECs align with the direction of blood
flow and act as a selective barrier between blood constituents and the extravascular
space while also preventing thrombosis (Hunt et al., 2002). The intima may also
contain a thin layer of sub-endothelial connective tissue and axially-orientated
smooth muscle cells (SMCs) (Humphrey, 2002). The media is the thickest and most
significant layer in terms of mechanical compliance at physiological pressures and
consists of SMCs in an extracellular network of elastin fibres and collagen fibrils as
well as an aqueous ground substance (Holzapfel et al., 2000). The media is
separated from the adventitia by the external elastic lamina. The adventitia is the
outermost layer and consists largely of thick bundles of collagen fibrils with admixed
elastin, fibroblasts, and nerves (and in some cases, a network of blood vessels
called the vasa vasorum) (Holzapfel and Ogden, 2006).
Coronary arteries display both active and passive mechanical behaviour
(Fung, 1993). Active behaviour is determined by the contraction of SMCs that are
arranged spirally around the vessel wall allowing them to function as control points
for blood distribution. The highly non-linear passive mechanical properties of the
multi-layered artery are dependent on many factors such as the relative amounts of
elastin and collagen present, the age of the artery and the level of disease present
(Holzapfel et al., 2005).
9
2.2
Coronary Artery Disease
CAD (also known as coronary atherosclerosis or ischemic heart disease) is a
condition involving the accumulation of plaque (an atheroma or lesion) in the walls of
the coronary circulation leading to narrowing and hardening of the arteries
(Holzapfel and Ogden, 2006), as shown in Figure 2.2. Low density lipoprotein (LDL)
particles accumulate in the intima causing ECs to produce adhesion molecules that
latch onto blood monocytes. The monocytes mature into active macrophages which
ingest modified lipoprotein particles, ending up as fat-laden foam cells. SMCs
proliferate and migrate to the intima promoting extracellular matrix accumulation in
the growing atherosclerotic plaque (Zohdi et al., 2004). As the atheroma develops it
increases in volume and may occlude the lumen of the artery thus restricting the
blood supply to the myocardium leading to myocardial ischaemia. The mismatch in
the mechanical properties of the vessel wall and the stiffer atheroma may also lead
to rupture or detachment of the atheroma, creating a downstream clot (embolism) in
the blood flow (Holzapfel et al., 2004).
Risk factors for CAD include hyperlipidaemia, tobacco smoking, obesity,
diabetes mellitus, hypertension, psychosocial and dietary factors, physical inactivity
and a genetic disposition (Yusuf et al., 2004). The risk of CAD increases
progressively with adult age and with the level of systolic blood pressures
(Lewington et al., 2003). CAD may be treated by a number of interventional
techniques including coronary artery bypass grafting (CABG), atherectomy, laser
vaporisation techniques, percutaneous transluminal coronary angioplasty (PTCA)
and stent deployment (Topol, 2008). CABG involves diverting the blocked blood flow
through an alternative path using a vein or artery from elsewhere in the body or a
prosthetic vessel in order to restore blood flow to the effected cardiac muscle.
Specific advantages of CABG include the ability to revascularise chronic total
occlusions (CTOs) and the ability to treat anatomic subsets that are not
10
approachable with stents (Topol, 2008). However, the CABG procedure is highlyinvasive and requires a considerable recovery period. It normally requires the heart
to be stopped and therefore a cardiopulmonary bypass module is required. Although
CABG has a low death rate it has associated complications and so is more
commonly used for high-grade stenosis or occlusion (Drenth et al., 2002).
Figure 2.2 Schematic representation of the process of atherosclerosis
PTCA, also known as balloon angioplasty, involves making a small incision
in the femoral artery and guiding a balloon catheter to the blockage site in the
coronary arteries (Agema et al., 2004). The balloon is expanded thereby pushing the
atheromatous plaque into the arterial wall, thus restoring vessel patency. Although
PTCA is non-invasive, it is associated with high rates of restenosis (30-60%) due to
elastic recoil of the arterial wall and/or the formation of IH (Mercado et al., 2001;
Topol, 2008). Therefore, PTCA is often combined with stent implantation in order to
reduce the incidence of restenosis (Serruys et al., 1994; Svilaas et al., 2007).
11
2.2.1 Restenosis
Restenosis is the re-blocking of a stented arterial segment causing a reduction in
luminal size and a restriction in blood flow. It primarily involves the formation of a
neointimal layer that occurs as a result of an exaggerated healing response
prompted by vascular injury (EC denudation, vessel wall stretching and sometimes
plaque rupture) consequent to high pressure (10-20 bar) balloon expanded stent
implantation (Kraitzer et al., 2008), as shown in Figure 2.3.
Figure 2.3 Stent restenosis: the re-blockage of a stented artery owing primarily to
the formation of neointimal hyperplasia
The biological response to stenting may consists of four phases (Kraitzer et
al., 2008; Welt and Rogers, 2002), as shown in Figure 2.4.
Figure 2.4 Schematic representation of the restenosis process after stent
implantation
a) Platelet Aggregation: The arterial injury that occurs during the stenting
12
procedure (endothelial denudation and medial dissection) causes platelet
aggregation and activation, thus initiating an inflammatory cascade and
releasing adhesion molecules that trigger thrombus formation. The
thrombosis that develops may be a major cause of vessel re-occlusion in the
first month after stenting (Edelman and Rogers 1998).
b) Inflammatory phase: Over days to weeks, a variety of white cells may gather
at the injury site, secrete factors and influence the healing tissue. The
inflammatory phase may persist for months.
c) Proliferation phase: Between one and three months after stent implantation,
SMC migration and proliferation and IH formation is the key mechanism of
restenosis (Topol, 2008). SMCs migrate from the vessel media (and possibly
adventitia) through the fracture of the internal elastic membrane to the intima
where they may continue to proliferate and synthesise extra-cellular matrix
(ECM); the major component of the restenotic lesion.
d) Late remodelling phase: involves alterations in vessel wall geometry through
redistribution
of
cells
through
processes
such
as
selective
cell
proliferation/apoptosis, or matrix synthesis/degradation (Hunt et al., 2002) in
order to adapt to the new strain environment imposed by the stent. Initial
deformation consists of elastic recoil following balloon deflation. Later,
geometric remodelling occurs as the artery shrinks, owing to collagen
deposition and fibrosis. Although remodelling in a BMS and DES stented
vessel is relatively small, the arterial wall may be squeezed through the stent
strut gaps from without. The artery opposes the strain imposed by the stent
struts through increased collagen deposition, marked destruction of elastin,
and persistent inflammation (Edelman and Rogers 1998). As the amount of
scar develops, blood flow is gradually reduced.
13
The classic binary definition of restenosis is an angiographically-measured
obstruction of 50% diameter stenosis or more at the site of a previously treated
coronary vessel (Briguori et al., 2000). When the lumen diameter is reduced to 50%
or less, coronary flow reserves become impeded. Late loss is a continuous
angiographic measure of lumen deterioration and is usually calculated by
subtracting the minimum lumen diameter (MLD) at follow-up from the postprocedural MLD (irrespective of the locations of MLD measurements). Late loss
measurements in stented segments offer a surrogate marker of IH. Target vessel
revascularisation (TVR) is defined as any repeat percutaneous coronary intervention
(PCI) of the treated coronary segment or CABG of the target vessel. ISR can be
characterised as focal (≤10mm long), diffuse (>10mm), proliferative (>10mm and
extending outside the stent) or total occlusion (Dangas et al., 2010).
The nature of restenosis appears to be related to the stent platform used
(Sangiorgi et al., 2007). Restenosis with sirolimus-eluting stents is mostly focal
(>90%) and located at the stent edges, whereas diffuse or total occlusion accounts
for about 50% of restenosis cases after implantation of polymer-coated paclitaxeleluting stents (Topol, 2008). Studies have shown that many factors associated with
the stent deployment procedure also contribute to restenosis outcomes, such as
procedural complications (Castagna et al., 2001), stent-induced injury or stretch
which is a function of maximum balloon pressure (Arakawa et al., 1998; Hoffmann et
al., 2004; Schwartz et al., 1992) and the stent-to-artery ratio. Stent design variables
also influence the biologic response to stent implantation (Hara et al., 2006), such
as the number of struts (Briguori et al. 2002), strut thickness and profile (ideally <75
µm) (Kastrati et al., 2001) (Palmaz, Benson et al. 1999), the release of toxic metals
(e.g. Ni, Mo) (Köster et al. 2000; Saito et al. 2009) and haemodynamic factors
(Duraiswamy et al., 2007). The largest alteration of shear stress in the blood occurs
at the entry and exit segment of stents where IH is often focused (LaDisa 2006).
14
Stenting introduces significant and chronic alterations in the biomechanical
environment of the cells within the local coronary arterial wall. The levels of stresses
and strains are elevated compared to normal physiological conditions and these
levels have been used as determinants in ISR development (Arakawa et al. 1998;
Zahedmanesh et al. 2010). The proliferative vascular response in the restenosis
process is exacerbated by the persistent stimuli of rigid metallic struts within the
arterial wall (Topol, 2008). Vessel stiffness is increased by the presence of the
expanded stent. Consequently, the cyclic strain environment in the artery is changed
from a low-mean, high-amplitude strain environment to a high-mean, low-amplitude
strain environment (Vernhet et al., 2000), as shown in Figure 2.5. In a stented artery
the mean strain can increase by as much as 20-50% and the cyclic strain amplitude
can fall as low as 1-2% (Vernhet et al., 2001, 2000). Cyclic strain amplitude is an
important regulator of SMC growth capacity within a stented artery (Colombo et al.
2012). Therefore, compliant stent designs that allow cyclic strain amplitude levels to
remain close to the normal physiological range (5-8%) (Draney et al., 2002; Vernhet
et al., 2000), and thus maintain the anti-proliferative effect of cyclic strain on SMCs,
could offer an effective strategy to reduce the rates of ISR.
Figure 2.5 Influence of stenting on the cyclic strain environment in the local coronary
artery
15
2.3
Coronary Stents
Coronary intravascular stents are cylindrical mesh structures which are expanded at
the site of stenosis in order to scaffold the vessel walls and prevent elastic arterial
recoil (Lally et al., 2006). Coronary stents are traditionally manufactured from
materials such as stainless steel (316L), cobalt chromium, tantalum, titanium, nitinol
or polymers (Sangiorgi et al., 2007). Stent deployment is normally achieved by
expanding an angioplasty balloon (pressure >16 bar) onto which the stent has been
crimped. Alternatively, deployment may be achieved by releasing a self-expanding
stent or a coiled stent from a restraining sheath. Stents are used to mechanically
support the vessel wall, to treat balloon or atherectomy device-induced coronary
dissections, and to minimise post-procedure vessel recoil and late negative vascular
remodelling (Savage et al., 1994; Topol, 2008). The stent diameter is normally
selected to approximate a 1.1:1 ratio compared to the angiographically normal
adjacent vessel, while stent length should cover the entire lesion length (Topol,
2008). Advantages of stent procedures include relatively mild morbidity, fast
reperfusion and rapid stabilisation of patients with MI. Limitations of stents include
the inability to treat a CTO (Topol, 2008).
2.3.1 Bare-Metal Stents
The results of a number of clinical trials reported that bare-metal stents (BMS)
reduced ISR rates from 35% with PTCA to 19% (Kandzari et al., 2002; Mercado et
al., 2001). However, ISR and ST remain a major limitation of BMS implantation. In
general, an event that occurs within 30 days of stenting is called an early stent
thrombosis (EST). An event that occurs more than 30 days and 12 months after
stent implantation is called a late stent thrombosis (LST), and a very late thrombosis
(VLST), respectively. Studies have shown that LST is uncommon with BMS, and
that events have a mean time of 2.4 to 3.6 months. Although the incidence of ST is
low with BMS (<2% of cases) there is a high risk of MI or even death (Ong et al.,
16
2005). In order to address the complications associated with using BMS, new types
of drug-eluting stents (DES) were developed.
2.3.2 Drug-Eluting Stents
Typically, DES are composed of three principal components: the stent backbone
(often a BMS platform), the pharmacologic agent (typically anti-proliferative drugs
such as sirolimus, paclitaxel, taxus, zotarolimus or everolimus) which is intended to
reduce IH and sometimes other effects, and a polymer designed to mediate
abluminal drug release over a two-to-four week period (Bailey 2009) in order to
reduce ISR (Garg & Serruys, 2010). Several clinical trials (CYPHER, RAVEL,
SIRIUS, TAXUS, ENDEAVOUR, and SPIRIT) have been performed that
demonstrate the ability of DES to significantly reduce ISR, LST and TVR rates
(typically <3%, <1% and <10% respectively) compared to BMS (Dangas et al., 2013;
Kandzari et al., 2013). Despite these improvements, limitations remain with regard
to the safety and efficacy of DES in terms of late adverse clinical events, such as
LST and ISR. The suggested mechanisms for ISR and ST with DES include in-stent
neo-atherosclerosis, neointimal hyperplasia, persistent endarteritis, uncovered
struts, stent malapposition (Colombo and Latib 2008), strut fracture (Alqahtani et al.,
2013; Aoki et al., 2007; Lim et al., 2008) and the persistent presence of the device
(Sammel et al., 2013). Patients implanted with DES normally require DAPT for about
3 months in order to minimise the risk of ST (Kim et al., 2009; Nishinari et al., 2013);
though a number of side-effects are associated with DAPT (Albaladejo and
Samama, 2010). Many experts suggest that continued development of permanent
stents (BMS and DES) is not likely to eliminate these problems and therefore radical
alternatives to conventional stents, such as bioabsorbable stents (BAS), must be
developed (Di Mario and Ferrante, 2008).
2.3.3 Bioabsorbable Stents
After stent implantation, the local vascular healing response (re-endothelisation,
17
remodelling etc.) and the majority of cardiac events including ISR normally occurs
within six-to-nine months (Edelman and Rogers, 1998), beyond which the utility of
permanent stents is unclear (Colombo and Karvouni, 2000; Jabara et al., 2009;
O’Riordan, 2006). Permanent stents interfere with normal vessel function and can
induce adaptive changes and abnormal reactions in the local tissue (Hoffmann et
al., 1996). Owing to the drawbacks of BMS and DES, intravascular stent technology
has progressed toward the development of bioabsorbable stents (BAS) that
temporarily scaffold the vessel during the initial period of high risk for recoil, and
then degrade in the long term (Erne et al., 2006). The gradual loss of mechanical
integrity due to the bioabsorption of the stent should reduce the stimuli for
endothelial irritation, endarteritis, IH, and ISR (Williams, 2006) while facilitating the
restoration of normal local vascular compliance and permitting positive late
remodelling (Anis and Karsch, 2006). BAS may help avoid many of the long term
problems of permanent stents, such as LST and ISR. BAS do not preclude repeat
PCI or surgical revascularisation, may be used repeatedly in a single vessel, may be
useful in paediatric patients where growth adaption and repeat PCI is often required
and may provide new treatment options for diffuse disease and unstable or
vulnerable atherosclerotic plaques (Gundogan et al., 2014). As the degradation of
BAS and their polymer coatings is not expected before six months, it is unlikely that
the minimum duration of DAPT for coated BAS will differ greatly from current
metallic DES (Kim et al., 2009; Nishinari et al., 2013); though uncoated BAS could
potentially minimise DAPT to one month or less in line with BMS guidelines (Levine
et al., 2011).
Stack et al. developed the first BAS, composed of poly-L-lactide (PLLA) and
tested it in animal models (Stack et al., 1988). The Igaki-Tamai stent (Kyoto Medical
Planning Co. Ltd, Kyoto, Japan) was the first coronary BAS implanted in humans in
1998. The PLLA stent demonstrated safety and efficacy at 6 months (Tamai et al.,
18
2000) and in a long-term (>10 years) study (Nishio et al., 2012). Currently, there are
more than fourteen BAS available with the majority of these still under preclinical or
clinical evaluation, as shown in Figure 2.6. A review of all current BAS platforms is
beyond the scope of this work but may be found in (Gundogan et al., 2014; Zhang et
al., 2013). Instead, the sections that follow focus on absorbable magnesium stents.
Current BAS technology has a number of shortcomings that limit their
application in current clinical practice such as relatively large strut thicknesses and
crossing profiles (Gogas et al., 2014), limited extensibility and higher risk of strut
fracture (Zhang et al., 2013). BAS pose significant development challenges. The
ideal period that a coronary stent needs to remain in place is unknown, though some
clinicians suggest six-to-nine months may be sufficient (Colombo and Karvouni,
2000). Also, stent manufacturers are proposing different BAS solutions such as
polymeric, bare-metal, hybrid, and drug-eluting BAS platforms. An effective BAS
should possess a number of characteristics such as similar or better acute and long
term performance compared to current generation metallic DES, adequate radial
strength, steerable biodegradation kinetics, minimal or zero inflammatory or
thrombotic response with normal endothelial function and non-toxic degradation
products that leave no residual substances in arterial tissues (Bailey, 2009; Topol,
2008; Weiß et al., 2009). While PTCA, BMS, and DES are regarded by many as the
first, second, and third revolution of interventional cardiology, BAS are being
heralded by many as the fourth revolution (Onuma and Serruys, 2011). The global
coronary stent market, which had a value of €5.5 billion in 2012, is expected to grow
to €7.75 billion by 2016, with BAS projected to account for about €1.3 billion of this
annual figure (Transparency Market Research, 2014).
19
Figure 2.6 Details of current bioabsorbable stents (BAS) platforms (adapted from
(Zhang et al., 2013))
Magnesium Stents
The efficacy of an absorbable magnesium stent was first investigated by Heublein et
al. who implanted a prototype stent made from the AE21 alloy in porcine coronaries
(Heublein et al., 2003). The first absorbable magnesium stent (AMS-1, Biotronik AG)
implanted in human coronaries was manufactured from the WE43 alloy, as shown in
Figure 2.7(a). However, the corrosion rate and loss of scaffolding was too rapid and
20
resulted in relatively high rates of late vessel recoil. The absence of an antiproliferative drug on the AMS-1 also led to high rates of IH and ISR (Erbel et al.,
2007). In order to address these limitations, the DREAMS-1 stent (AMS-3, Biotronik
AG) had a revised alloy (non-commercial WE43 derivative) and stent design which
incorporated a paclitaxel-eluting polylactide-co-glycolide (PLGA) coating, as shown
in Figure 2.7(b). Lower recoil and IH were reported and the corrosion profile was
extended with restoration of normal vasomotor function reported at six months. The
stent was shown to be safe and demonstrated modest efficacy with regard to late
recoil and TVR rates (Haude et al., 2013). A number of design iterations were made
to the DREAMS-2 stent including the use of tantalum markers on a revised stent
design and a sirolimus-eluting PLA coating, as shown in Figure 2.7(c).
Figure 2.7 Magnesium coronary stent platforms manufactured by Biotronik AG: (a)
AMS-1, (b) DREAMS-1 and (c) DREAMS-2
Compared to DREAMS-1, DREAMS-2 showed increased flexibility, larger
deployment
diameter
capability,
improved
endothelialisation
and
reduced
inflammation in preclinical trials (Guy, 2013). A PLGA coating that used a mixture of
85% lactide and 15% glycolide provided the lowest intimal area at 28 days and 180
days, and the highest endothelialisation score at 28 days in a porcine model
(Wittchow et al., 2013). The alloy used in DREAMS-2 (WE43 derivative) degraded
(0.036-0.072 mg cm-2 day-1) into a magnesium-rich compound containing a large
21
amount of oxygen, most likely in the form of magnesium carbonate (MgCO3) and
Mg(OH)2 after 28 days in a porcine artery model, as shown in Figure 2.8. These
compounds were converted into a soft and gel-like amorphous calcium phosphate
product after 90-to-180 days implantation. The accumulation of calcium products in
the arterial wall may be of concern as it could potentially act as a precursor to
atherosclerotic plaque which is calcium-rich (Holzapfel et al., 2004); though this
requires further investigation. The performance of DREAMS-2 is being assessed in
the BIOSOLVE-II trial which commenced in late 2013.
Figure 2.8 SEM images and EDX mapping of DREAMS-2 degradation products after
28, 90 and 180 days implantation in porcine coronary arteries (Wittchow et al., 2013)
Investigations are on-going to assess the utility of various other magnesium
alloy systems for stent applications such as Mg-Y-Zn (Hänzi et al., 2010), Mg-Zn-Zr
(Hong et al., 2013), Mg-Zn-Mn (Fang et al., 2013) Mg-Zn-Y-Nd (Wu et al., 2012),
Mg-Nd (Seitz et al., 2012) and Mg-Li (Zhou et al., 2013), amongst others. The
ZMED stent (Zorion Medical, IN, USA) is a wire-form, balloon-expandable fullyabsorbable drug-eluting BAS composed of a (RE-free) proprietary magnesium alloy.
Preclinical studies showed that the device was completely absorbed with minimal
22
inflammatory tissue response within ninety days. Further preclinical evaluations are
needed before a first-in-human study which will begin in early 2015 (Zorion Medical,
2014). Other companies with absorbable magnesium stent programs include
Medtronic Inc. MN, USA (magnesium alloy stent with sirolimus coating), Boston
Scientific Corporations MA, USA (magnesium alloy) and QualiMed Innovative
Medizinprodukte GmbH, Winsen, Germany (magnesium alloy and polymer hybrid).
The aforementioned trials have shown that magnesium stents are safe and
feasible within small study groups and for simple lesion types. Despite this, only a
limited amount of data regarding their mechanical and corrosion performance
characteristics has been reported in animal (Di Mario et al., 2004; Heublein et al.,
2003; Waksman et al., 2006) and clinical (Bosiers et al., 2009; Erbel et al., 2007;
Haude et al., 2013) trials. Therefore, studies that can reveal the mechanical and
corrosion performance of magnesium stents are needed. Further, the elucidation of
the biomechanical environment that is supported by magnesium stents may prove
crucial in helping to solve the problem of ISR owing to the influence of cyclic strain
amplitude and stent radial stiffness on the development of restenosis (Colombo et
al., 2012; Vernhet et al., 2001).
WE43 was selected for this study as it was used in multiple generations of
the Biotronik AMS platform (AMS-1 to AMS-4) and currently represents the only
magnesium stent platform to successfully go through clinical trials for coronary
(Haude et al. 2013; Waksman et al. 2009; Erbel et al. 2007) and peripheral (Bosiers
et al. 2005; Bosiers et al. 2009) applications. However, limitations associated with
current magnesium stents include low mechanical properties compared to
commercial BMS and DES, relatively high rates of late lumen loss due to excessive
IH formation and premature loss of scaffolding (Erbel et al., 2007; Haude et al.,
2013), and biocompatibility concerns associated with alloying elements that may
remain in vascular tissues after stent degradation (Feyerabend et al., 2010).
23
These limitations can be addressed in future through the development of
high-purity magnesium alloys (particularly with minimal impurities such as Fe, Cu, Ni
and Si) alloyed with high-purity hardeners (non-misch metal) and alloying elements
(such as Zn, Nd, Zr, Y) that are biocompatible and safely eliminated from arterial
tissues following device degradation (Li and Zheng, 2013). Ultra-fine or nanocrystalline magnesium alloys offer significantly improved mechanical and corrosion
performance and may be obtained by a number of severe plastic deformation (SPD)
techniques that are the subject of on-going research and development (Zheng et al.,
2014). Further improvements in corrosion performance and stent biocompatibility
(improved endothelialisation, reduced inflammation and IH, ISR and ST rates) will
be achievable through surface modification techniques, such as surface treatments
and coating and drug elution strategies specific to magnesium alloys (Hornberger et
al., 2012).
2.4
Magnesium and Magnesium Alloys
Magnesium is essential to human health and is involved in numerous metabolic
reactions and biological mechanisms (Hruby and McKeown, 2013). Magnesium
deficiency is common in industrialised nations and can increase our susceptibility to
many illnesses, including CAD and stroke (Hartwig, 2001). Oral magnesium
supplementation can improve endothelial health (Fuentes et al., 2006), offer a
modest reduction in CAD risk (Mathers and Beckstrand, 2009) and can aid skeletal
growth and development. The risk of hypermagnesemia is low as the kidneys very
efficiently excrete excess magnesium (Birrer and Shallash, 2002). Absorbable
magnesium implants were first explored in the early twentieth century but were later
abandoned due to the inherent rapid corrosion and evolution of hydrogen (H2) gas
subcutaneously (Witte, 2010). Recent progress in magnesium metallurgy, further
understanding of corrosion processes, and the development of corrosion protection
techniques has seen a burgeoning interest in magnesium biomaterials (Song, 2011).
24
Magnesium alloys possess many physical and chemical properties that
make them highly-suited for use in coronary stents such as low thrombogenicity and
high biocompatibility (Erbel et al., 2007), suitable mechanical properties (Zheng et
al., 2014) favourable corrosion kinetics (Kirkland et al., 2010) and zero
magnetisation (MRI and CT compatible) (Staiger et al., 2006). The metabolic
conversion of magnesium to its chloride, oxide, sulphate, and phosphate salts is well
tolerated in vivo and the resulting by-products are not retained in the arterial wall
(Bowen et al., 2014). Intravenous administration of magnesium has been shown to
be safe and acts as a systemic and coronary vasodilator (Fuentes et al., 2006). In
vitro testing involving human cells has shown that magnesium can control SMC
proliferation and stimulate ECs (Sternberg et al., 2011). Extensive preclinical testing
has shown that bare-metal and drug-eluting magnesium-based absorbable stents
are both safe and feasible (Erbel et al., 2007; Haude et al., 2013).
Although the mechanical properties of pure magnesium are too low for loadbearing implants such as stents, the strength, ductility and corrosion resistance of
magnesium can be significantly improved by developing high-purity alloys (Zainal
Abidin et al., 2013) and by using alloying elements which aid precipitation and solidsolution hardening (Yang et al., 2008). Micro-alloying with RE elements such as Y,
Zr, Nd and Ce produces ductile alloys with high fatigue strength and fine
microstructural features that can restrict grain growth during solidification and pin
grain boundaries during later thermo-mechanical processing (Song, 2011).
Therefore, RE alloys such as WE43 are of particular utility in stent manufacturing
where the material is repeatedly subjected to thermo-mechanical processing steps
such as tube drawing and stress-relief by annealing (T-5 and T-6 heat treatments)
(Hänzi et al., 2010)), and in-service where the stent is subject to cyclic loading under
the influence of pulsatile blood flow in the coronary artery (Hänzi, Gerber et al.
2010).
25
2.4.1 The WE43 Alloy
WE43 typically contains (by weight %) approximately 93% magnesium, 3.7-4.3%
yttrium (Y), >0.4% zirconium (Zr), with the remainder (2.4-4.4%) consisting of RE
elements, principally Neodymium (Nd) (2.0-2.5%) and heavy RE elements such as
Ytterbium (Yb), Erbium (Er), Dysprosium (Dy) and Gadolinium (Gd) (Magnesium
Elektron, 2007). Zr is a potent grain-refining agent in magnesium alloys and
contributes to strengthening due to the formation of fine grains by grain boundary
strengthening (Avedesian and Baker, 1999). Y and Nd are used mainly to increase
strength and service temperature, and to improve creep and corrosion resistance
(Zhang et al., 2009). Y is commonly added to magnesium alloys in the form of Y-rich
misch metal which contains impurities (such as Fe, Ni and Si) that may be
detrimental to corrosion performance (Song and Atrens, 2003). Moreover, Y-rich
misch metal contains approximately 20% by weight of other RE elements (such as
Nd, La, Dy, Ho, Er, Tm, Yb and Lu) (Zhang et al., 2009). The effects of these RE
elements in vivo is not completely understood though in vitro studies involving
human SMCs suggest that some RE elements may have cytotoxic effects and
negative influences on inflammatory reactions (Drynda et al., 2009; Feyerabend et
al., 2010).
One of the notable features of Mg-Y alloys is that they exhibit both solution
and age-strengthening effects; these being strongly linked to the solubility of Y in αMg matrix and the volume of intermetallic phases (Zhao et al., 2011). The Mg-Y and
Mg-Zr phase diagrams are shown in Figure 2.9(a) and (b), respectively. From these
phase diagrams it can be seen that a small change in composition can result in
different phases and hence different material properties occurring. Therefore, control
of the composition is important. The process of grain refinement using Zr may be
described with reference to the phase diagram shown in Figure 2.9(b). For alloys
such as WE43 which are saturated with Zr in the liquid phase, cooling allows some
26
precipitation of α-Zr particles as the freezing temperature is approached. At the
peritectic temperature (~654°C), suspended Zr particles can form nuclei for the
peritectic reaction with Zr-saturated Mg liquid to produce solid α-Mg with high Zr
content. Owing to the close similarity in Mg and Zr lattice parameters, on final
solidification a relatively homogenous distribution of fine hexagonal grains forms
which reflects the growth mode of the two components. As maximum solubility of Zr
is necessary for obtaining the highest grain refinement, some elements which form
stable high melting point compounds with Zr should ideally be removed and not be
present in Mg-Zr alloys such as Al, Si, Fe, Ni, Mn Sn, Co and Sb (Friedrich and
Mordike, 2006).
Figure 2.9 Phase diagrams (a) Mg-Y (b) Mg-Zr (Li et al., 2013; Zhao et al., 2011)
The microstructure of wrought and heat-treated WE43 normally consists of
recrystallised grains with a large volume fraction of precipitates within the αmagnesium matrix. Typically, the precipitates present include (1) Mg12(RE,Y)
polyhedral precipitates located within the matrix at grain boundaries, (2) spherical
Zr-rich and cubic Y-rich precipitates, both containing Fe and Si impurities and (3)
needle-like β-phase precipitates, as shown in Figure 2.10 (Coy et al., 2010). The
relative Volta potential differences between micro-constituent phases and the α27
magnesium matrix indicate the corresponding role of the micro-constituent phases in
potential galvanic interactions. In the case of WE43, Zr-rich precipitates exhibit
greater volta potential differences (+180 mV) when compared to Y-rich precipitates
in the presence of Fe (0.2 at.%) (+90 mV), Mg12(Nd,Y) (+25 mV) and β-phase
precipitates (+15 mV) (Coy et al., 2010).
Figure 2.10 Backscattered scanning electron micrograph of wrought WE43 alloy
(Coy et al., 2010)
2.4.2 Mechanical Properties of Magnesium Alloys
A useful review of the mechanical properties of many current absorbable
magnesium alloys was presented by Li and Zheng (Li and Zheng, 2013). Studies
have reported moderate to strong asymmetry in the mechanical properties (ratio of
yield stress in tension and compression) of magnesium alloy tubes with consistently
lower properties observed in compression when compared to tension (Gunde et al.,
2011; Vedani et al., 2012). This has been attributed to the crystallographic texture of
samples and to the occurrence of preferred {10–12} <10–11> twinning in
compression. Heat treatment provides an effective method to reduce asymmetry,
which is advantageous with respect to the fatigue behaviour of stents where the
lower of the tensile or compressive yield stress value determines the fatigue
strength (Frey et al., 2005). A number of studies have reported mechanical
28
properties for extruded WE43 derived from heat-treated bulk test specimens (Gu
and Zheng, 2010; Magnesium Elektron, 2007; Witte et al., 2008). However, bulk
specimens have very different materials processing histories to stents and the
presence of size effects means that the mechanical behaviour may not capture the
true behaviour of the device (Connolley et al., 2005; Murphy et al., 2003; Savage et
al., 2004; Wiersma and Taylor, 2004). The processing of magnesium alloy stent
precursor tubes has been described in a number of studies (Fang et al., 2013; Ge et
al., 2013, 2011; Hassel et al., 2006). To date, few studies have reported mechanical
property data for stent precursor tubes of the WE43 alloy (Witte et al., 2008) or other
magnesium alloys (Hassel et al., 2006; Kutniy et al., 2009); though the surface
treatments used in these studies differed to those used for coronary stents.
Generally, these studies did not report on the geometry of the specimens used and
either employed no surface finishing or simple mechanical polishing of specimens.
The mechanical properties for WE43 reported in these studies is summarised in
Table 2.1. Material property data that is derived from WE43 precursor tubes and
from specimens with similar size scale and surface treatments as coronary stents
are necessary. As well as revealing the true mechanical behaviour of finished WE43
stent strut materials, such experimental data would serve as input in future
computational models of magnesium stent performance.
Table 2.1 Available data for the mechanical properties of WE43 specimens
2.4.3 Corrosion of Magnesium Alloys
Magnesium is one of the most anodic metals having a standard electrode potential
29
of -2.37 V. Therefore, with magnesium there is often a high chance of galvanic
corrosion relative to other metal elements (Song, 2011). The corrosion processes
and reaction kinetics of magnesium alloys depend on parameters including
composition of the alloy, metallurgical treatment, surface treatment, composition of
the interacting electrolytes, chemistry and biology of the environment, and the
transport and metabolic phenomena associated with the reactants and the corrosion
products (Bowen et al., 2012). The body electrolyte contains many anions (mainly
chloride, phosphate, and bicarbonate ions), cations (mainly Na+, K+, Ca2+ and Mg2+),
organic substances and dissolved gases such as oxygen (Ratner et al., 1996). The
high corrosion rates of magnesium alloys in vivo and in simulated body fluid (SBF)
are generally attributed to the influence of aggressive ions such as chlorides,
phosphates, sulphates and carbonates (Xin et al., 2008). Magnesium surfaces do
not naturally passivate in aqueous chloride-containing solutions at physiological pH
levels (Song, 2011) and so accelerated corrosion can be expected (Zhao et al.,
2008). The corrosion attack in aqueous environments often involves galvanic
coupling between anodic and cathodic areas that result in preferential dissolution of
the (anodic) magnesium matrix (alpha (α) phase) and cathodic reduction of water
(Zeng et al., 2006). The chemical process associated with the attack can be
summarised as follows:
Anodic reaction: Mg → Mg 2 + + 2e −
(3.1)
−
−
Cathodic reaction: 2H 2 O + 2e → H 2( gas) + 2OH
(3.2)
Combined reaction: Mg + 2 H 2 O → Mg (OH ) 2 ( solid ) + H 2 ( gas )
(3.3)
Generally, the dissolution rate of the α-phase exceeds the formation rate for
the magnesium hydroxide (Mg(OH)2) surface layer and so corrosion is not easily
self-inhibited. Chlorides can transform the partially protective Mg(OH)2 into soluble
30
MgCl2, thus accelerating magnesium dissolution. The evolved H2 gas causes further
detachment of the surface layer and increases the surface alkalinity (Kainer, 2003).
Several studies have shown that magnesium alloys with fine-grained
microstructures exhibit significantly higher corrosion resistance in chloridecontaining environments as compared to coarse-grained microstructures (AlvarezLopez et al., 2010; Amira et al., 2008; Argade et al., 2012; Kannan, 2010; Song et
al., 1999). This has been attributed to the higher fraction of continuous β-phase that
is more finely distributed around finer α-grains and acts as a barrier to galvanic
corrosion (Song et al., 1999). Fine-grained microstructures show lower porosity,
higher polarisation resistance and the most positive pitting potentials and repassivation potentials (Argade et al., 2012). The corrosion products (mainly
magnesium oxide (MgO)) formed on the surface of fine-grained alloys are more
protective against the action of chloride ions (Alvarez-Lopez et al., 2010). A one
order of magnitude decrease in corrosion rate was observed between a coarsegrained (70 µm) and fine-grained (~0.7 µm) Mg-Y-RE alloy (Argade et al., 2012).
Crystallographic orientation in magnesium alloys has a strong influence on surface
processes such as corrosion, dissolution and oxidation (Song and Xu, 2012). The
influence of crystallographic orientation was shown to be about 10% of the total
corrosion depth for pure magnesium (Liu et al., 2008). Also, the highest corrosion
resistance was observed for grains near (0001) orientation which are commonly
found on the processed surfaces of precursor tubes.
A comprehensive discussion of all corrosion processes affecting magnesium
alloys is outside the scope of this work but can be found in (Atrens et al., 2011;
Song and Atrens, 1999). The main corrosion processes expected to influence the
behaviour of magnesium alloy stents in vivo (Gastaldi et al., 2011) are discussed
here, including micro-galvanic and localised (pitting and filiform) corrosion, intergranular corrosion (IGC), stress corrosion cracking (SCC) and fatigue corrosion.
31
Micro-galvanic Corrosion
The non-uniformity in the composition, microstructure, and crystalline orientation of
magnesium alloys can induce various chemical and electrochemical activities within
the material resulting in micro-galvanic corrosion which leads to dissolution and
thinning of the metal (Coy et al., 2010). Some areas, grains, phases, particles and
impurities may act as anodes while others act as cathodes, as shown in Figure
2.11(a). Magnesium is susceptible to corrosion owing to galvanic activity among its
primary constituents namely, primary α, eutectic α and β-phases, as shown in Figure
2.11(b). The magnesium α-matrix is an anode and is preferentially dissolved due to
its lower content of alloying elements (Kalb et al., 2012) leading to undermining and
fallout of the secondary phase particles, as shown in Figure 2.11(c).
Figure 2.11 (a) Galvanic corrosion (internal) in magnesium alloys. (b) Galvanic
activity among the primary α, eutectic α and β-phases and (c) the resulting
morphology. (e) a continuous β-phase inhibits corrosion and (f) the resulting
morphology
The β-phase serves as a galvanic cathode and accelerates the corrosion of
the α-phase if the volume fraction of β-phase is small. However for a high volume
fraction, a continuous β-phase can act as a barrier to inhibit corrosion (Atrens et al.,
32
2011), as shown in Figure 2.11(d and e). Galvanic corrosion processes at grains
and grain boundaries can also be accelerated by deformation (plastic strain) in
metal parts (Krawiec et al., 2012). In the case of a deformed stent strut, the tensile
side of the strut may be anodic with respect to the compression side, leading to
increased galvanic corrosion (Ratner et al., 1996).
Localised Corrosion
Pitting Corrosion: Pitting corrosion is a localised electrochemical reaction that
leads to the formation of small pits on the affected surface. It can be initiated on
stent surfaces by several processes such as damage of the passive film, localised
plastic yielding during crimping and expansion, the production of stents with
improperly polished surfaces, particle inclusions, inhomogeneous crystallinity or
crystal defects causing dislocations to emerge on the surface and nonhomogeneous
environments containing aggressive chemicals, such as chloride, causing localised
dissolution of the passive film (Song, 2011). Pitting corrosion tends to initiate at
irregular pits that spread laterally over the whole surface but may not form deep pits
owing to the self-limiting nature of localised corrosion in magnesium (Song and
Atrens, 2007).
Filiform Corrosion: Filiform corrosion is instigated by an active corrosion cell that
migrates across the metal surface; the head is the anode and the tail is the cathode
(Song and Atrens, 1999). Filiform corrosion does not occur on uncoated pure
magnesium though it has been reported for anodized and coated magnesium
materials (Lamaka et al., 2008) as well as for alloys with relatively resistant surface
oxides (Lunder et al., 1994) such as the AZ series.
Inter-Granular Corrosion (IGC)
The grain boundaries are preferred sites at which precipitation and segregation of
the β-phase occurs in magnesium alloys (Zeng et al., 2006). Magnesium alloys
33
rarely exhibit inter-crystalline attack because the grain boundary phases are
invariably cathodic to the grain interior; instead, corrosion is normally concentrated
in the area adjoining the grain boundary and may proceed until the grain is undercut
and falls out (Song, 1999). IGC with rapid material dissolution and grain detachment
has been reported for WE43 owing to selective attack of eutectic shaped Nd-based
compounds at grain boundaries which act as anodic sites due to their low
electrochemical potential with respect to grain areas (Valente, 2001). Additions of Zr
may also lead to IGC in WE43 (Song, 2011).
Stress Corrosion Cracking
Stress corrosion cracking (SCC) results from a combination of static stress (either
applied or residual) and a corrosive environment (as opposed to corrosion-fatigue
where loading is dynamic) (Ratner et al., 1996). With magnesium alloys, corrosion
may be initiated at a microscopic crack tip that does not re-passivate (Winzer et al.,
2007). This leads to brittleness at the tip of the crack, because the H2 gas evolved
during the corrosion process is absorbed (Winzer et al., 2005b). Incremental crack
growth may then occur resulting in implant failure. SCC is more marked in wrought,
rather than cast magnesium alloys and is mainly transgranular in nature. SCC does
not occur in pure magnesium and only occurs in Zr-containing alloys such as WE43
as stresses approach the yield stress (Song and Atrens, 1999).
Fatigue Corrosion
Stents are subject to cyclic (fatigue) mechanical loading due to the pulsatile nature
of blood flow. The resulting strain, abrasion or wear may continuously rupture the
surface passivation layer thereby reducing the corrosion resistance of the exposed
material, thus increasing the number of metallic ions released (Ratner et al., 1996).
Further, if a crack is present, anodic dissolution may occur at the tip of a crack.
Corrosion fatigue is highly-significant as there is no limiting stress (endurance limit)
34
below which corrosion-induced cracking will not occur. In their study on WE43, Gu
et al. (Gu et al., 2010) showed that the fatigue strength of samples in air was
significantly reduced in static SBF and further reduced in dynamic SBF. Under the
influence of cyclic loading the corrosion rate of specimens was shown to increase
and a toughness fracture mode was identified where a crack initiated at a corrosion
pit and propagated through the material.
2.4.4 Influence of Stent Manufacture on Magnesium Alloys
To date, a detailed study of the influence of laser cutting on the performance of
WE43 stents has not been reported. The laser cutting of magnesium alloys requires
the use of an inert shield gas (e.g. Ar or Nitrogen) rather than the use of O2 which
results in the formation of a refractory layer of MgO in the kerf which reduces cutting
efficiency (Demir et al., 2013). Laser processing has been shown to increase the
mechanical properties and reduce the micro-hardness of WE43 by influencing
factors such as solid–solution strengthening, dislocations, second phase particles
and geometrical constraints posed by grain boundaries that obstruct dislocation
motion (Santhanakrishnan et al., 2013). A number of studies report significant
increases in the corrosion resistance of magnesium alloys owing to the influence of
the rapid heating and cooling induced in the HAZ during laser processing (Abbas et
al., 2005; Guo et al., 2005; Liu et al., 2005). This has been attributed to
morphological changes in the material microstructure such as significant grain
refinement and the redistribution of intermetallic phases (Abbas et al., 2005). Higher
concentrations of alloying elements on the machined surface, which aid the
formation of more protective oxide films, have been reported for laser-processed
magnesium alloys (Liu et al., 2005). Laser cutting stents from magnesium precursor
tubes exposes higher index crystal planes of lower corrosion resistance (Liu et al.,
2008). Hence, in the absence of sufficient surface treatment and HAZ removal,
reduced corrosion resistance might be expected on these surfaces.
35
There are numerous surface treatment techniques available for the surface
modification of magnesium alloy implants (Hornberger et al., 2012; Ma et al., 2014;
Zheng et al., 2014). Chemical etching is often used with magnesium alloys in order
to replace the native oxide layer with a more passive and corrosion resistant oxide
layer (conversion coating) of lower porosity and lower surface roughness (Nwaogu
et al., 2009). The chemical conversion layers,
which are formed in a complex
interaction of metal dissolution and precipitation in the aqueous solution, mainly
consist of magnesium phosphate, MgO, Mg(OH)2 and mixtures of other metal oxides
and hydroxides, which arise from the dissolved ions in the bath (Gray-Munro et al.,
2009). As the layers are grown in situ, adhesion to the substrate is good and the
surfaces obtained facilitate the adherence of any subsequently deposited coating
(Hornberger et al., 2012). In the case of AZ31 stents, chemical etching in a nitric
acid solution has been used to remove the heat-affected zone (HAZ), dross and
spatter that were formed on surfaces during laser cutting (Demir et al., 2012). GrayMunro et al. described chemical etching in phosphoric acid solution in order to form
a non-toxic coating on AZ31 specimens (Gray-Munro et al., 2009). The increased
corrosion resistance observed was attributed to improved surface uniformity and
crystallinity of the protective Mg(OH)2 and magnesium phosphate layer that was
produced by etching. Zhou et al. described the formation of a magnesium phosphate
coating on an AZ91D specimen in three main phases: (1) corrosion of the
magnesium matrix, (2) mass transposition in the liquid phase and the metal-solution
interface, and (3) formation and growth of crystal nucleus (Zhou et al., 2008). Layer
growth was attributed mainly to formation of Mg(OH)2 at the substrate-layer interface
by anodic reaction and dissolution mainly of the magnesium matrix material.
In medical device manufacturing, gas plasma processes offer useful surface
modification (Yang et al., 2010) and sterilisation (Halfmann et al., 2007) techniques
for devices with complex shapes, such as coronary stents. Plasma etching can be
36
used to remove manufacturing residues, foreign materials and bio-burden and offers
a clean, safe and environmentally-friendly alternative to traditional wet chemical
processes. Materials are removed from the surface by chemical reactions (to form
volatile products) and physical sputtering using inert gases such as argon or neon.
Due to the low energy of the gas plasma, layers of atoms are gradually removed
from the surface without significantly modifying the microstructure of the substrate
material (Yang et al., 2010). It has been shown that O2 plasma processing of
biomaterials leads to increased surface wettability (Yi et al., 2004) and enhanced
adhesion properties for any subsequently deposited coating (Nakamura et al., 1996)
which are of particular utility in the manufacture of coronary stents.
2.5
Computational Modelling of Magnesium Stents
The deployment of magnesium alloy stents by simple pressure and displacement
loads have been simulated in a number of FE studies (Sabir et al., 2012; Wang et
al., 2009). Grogan et al. (Grogan et al., 2012) compared the mechanical
performance of magnesium, iron, 316L and Co-Cr stents, based on multiple stent
platforms and predicted a higher fracture risk for magnesium stents owing to the
lower plasticity of the material. Computational models based on crystal plasticity
theory have been used to predict the influence of size effects in stent struts and
highlight the improvement in deformation performance of magnesium stents for
decreasing material grain size (Grogan et al., 2013). Similar model have been used
to investigate the failure risk and load-bearing capacity of magnesium stent struts
undergoing plastic deformations typical of stent expansion (Grogan et al., 2014b).
Wu et al. (Wu et al., 2010) applied an iterative morphing procedure in order
to reduce the predicted concentrations of stress and strain in a 2-D RUC strut model
during a stent deployment simulation. This procedure was combined with the
corrosion model of Gastaldi et al. (Gastaldi et al., 2011) in a shape optimisation
37
study of a 3-D stent structure in order to identify a stent design that displayed lower
mass loss and hence a prolonged scaffolding profile. The model was validated
based on preliminary corrosion experiments that involved the optimised AZ31 stent
design (Wu et al., 2012). A similar shape optimisation model was reported by Li et
al. (Li et al., 2014) where model validation was based on experimental corrosion
testing of unfurled stent geometries cut from AZ31 sheets. Although not discussed
here, a number of corrosion models have also been developed and applied in
computational studies of magnesium stents. This is discussed in section 2.5.1.
With the exception of the crystal plasticity models described above, many of
the aforementioned computational studies employ FE modelling based on
continuum plasticity theory where the material model parameters were derived from
mechanical testing of specimens that did not have the same processing history as
can be expected for magnesium stents (Gastaldi et al., 2011; Grogan et al., 2012,
2011; Wu et al., 2011, 2010). Therefore, improvements in the predictive capabilities
of future FE models, based on continuum plasticity theory, may be obtained through
the use of mechanical properties derived from specimens of precursor tube that
have surface treatments similar to magnesium stents.
2.5.1 Corrosion Modelling
Several numerical models for external galvanic corrosion, based on the boundary
element method (BEM), in which different galvanic couples are studied have been
proposed for fixed (Jia et al., 2006) and moving mesh (Deshpande, 2010; Thamida,
2012) techniques. A mathematical model for micro-galvanic corrosion that
accounted for the negative difference effect (NDE) that is observed in magnesium
alloys was proposed by Liu and Schlesinger (Liu and Schlesinger, 2009). A
numerical model for micro-galvanic corrosion within a binary phase metal alloy was
developed by Deshpande (Deshpande, 2011) using the Arbitrary Langrangian
38
Eulerian (ALE) method. The model used the non-linear polarization behaviour of the
alloy phases and tracked the moving boundary of the degrading phase. A
mathematical model based on potential theory and calibrated based on in vivo
corrosion data was used to show that the thickness of the electrolyte influenced the
galvanic corrosion of magnesium (Montoya et al., 2014). A macroscale corrosion
damage internal state variable (ISV) model that captured the effects of general
corrosion, pit nucleation, pit growth, pit coalescence, and intergranular corrosion
was developed by (Walton et al., 2014) and applied to the AZ31 magnesium alloy.
Models of other corrosion processes have been developed and applied to
magnesium alloys such as stress corrosion cracking (SCC) (Dietzel et al., 2010)
fatigue cracking (Xue et al., 2007) and crevice corrosion (Sun et al., 2014).
Many of the aforementioned modelling approaches focus on specific
corrosion processes that are not easily applicable to coronary stents due to the
difficulty of establishing the model parameters by experimental calibration
procedures, as well as the high computational costs associated with 3-D
geometries. Therefore, a number of alternative models been developed specifically
for modelling the corrosion of absorbable metallic stents. As numerical modelling of
the various complex phenomena affecting the corrosion performance of metallic
stents in vivo represents a significant challenge and as such this represents a
relatively new field of research, only a small number of phenomenological (Gastaldi
et al., 2011) and physical (Grogan et al., 2014a) corrosion models have been
developed and applied to magnesium stents.
A phenomenological corrosion model was developed by Gastaldi et al.
(Gastaldi et al., 2011) using the CDM approach and implemented within a FE
framework in order to account for the effects of stress corrosion and uniform microgalvanic corrosion processes on the behaviour of a 3-D magnesium alloy stent
geometry. The model was calibrated from experimental corrosion studies performed
39
on mechanically polished cylindrical specimens of extruded magnesium alloys
(AZ31, AZ61, AZ80, ZK60 and ZM21). The same corrosion model was later
calibrated from experimental studies performed on specimens that were cut from
magnesium alloy sheets and extruded bars and used to perform optimisation studies
for magnesium stents in which outputs such as stress, strain, acute and corrosioninduced stent recoil and mass loss were examined (Wu et al., 2011, 2010). In a
combined experimental and numerical study, Wu et al. (Wu et al., 2012) produced
an AZ31 stent and compared experimentally observed elastic recoil and mass loss
with those predicted by the corrosion model.
Grogan et al (Grogan et al., 2011) added an element-specific pitting
corrosion parameter to the damage law developed previously by Gastaldi et al.
(Gastaldi et al., 2011) in order to incorporate the effects of heterogeneous or pitting
corrosion in the modelling framework. The model was calibrated based on corrosion
experiments involving mechanically polished AZ31 foil specimens and was used to
predict the corrosion-induced recoil and mass loss of an AZ31 stent that was
deployed in an idealised coronary artery model.
A physical corrosion model, based on the FEM and adaptive meshing, was
developed by Grogan et al. (Grogan et al., 2014a) to model the diffusion-controlled
corrosion of a 3-D (pure magnesium) stent strut section. The study was predicated
on a transport-controlled process, based on the assumption that metallic ion
transport, particularly diffusion, was slower than the electrochemical reaction rate for
the stent surrounded by layers of corrosion product and tissue in the body. The
model predicted that the mass loss rate was proportional to the saturation
concentration of magnesium ions in solution. The diffusivity of the surrounding
environment was also shown to influence the corrosion rate.
The aforementioned corrosion models of magnesium stent performance
40
(Gastaldi et al., 2011; Grogan et al., 2011) involve stress-mediated corrosion
processes. However, since the stent deployment procedure is known to induce large
plastic strain (20-30%) (Murphy et al., 2003) corrosion models that explore the
interaction of plastic strain and plastic damage with corrosion damage are required.
Indeed, this has been recognised as one of the current challenges in degradable
material modelling (Gastaldi et al., 2011). Further, these corrosion models were
calibrated from corrosion experiments that used magnesium specimens that had
materials processing, surface treatments, and loading regimes that were dissimilar
to those of coronary stents which are all known to bear a significant influence on the
mechanical and corrosion behaviour of magnesium materials (Gastaldi et al., 2011;
Grogan et al., 2011). This is also true of many of the experimental studies of the
corrosion performance of absorbable magnesium materials and devices that are
currently available in the literature (Bowen et al., 2013; Coy et al., 2010; Grogan et
al., 2011; Kannan et al., 2008; Zhang et al., 2010). Therefore, mechanical and
corrosion experiments that use specimens with appropriate processing and loading
histories relevant to magnesium stents are required in order to reveal the true
behaviour of these devices. Such an approach should provide improvements in the
quality of the experimental data obtained. Such data would also prove useful in the
calibration and validation of numerical corrosion models of magnesium stent
behaviour.
2.6
Summary
The optimum treatment strategy to resolve the problem of ISR following stenting
with BMS and DES platforms has yet to be determined. Emerging strategies include
bioabsorbable stents (BAS) which provide temporary arterial scaffolding and are
absorbed in the body after six-to-thirty six months (Zhang et al., 2013). BAS
materials under investigation include polymers and absorbable metals, such as iron
and magnesium. Absorbable magnesium stents have shown promise in extensive
41
clinical trials (Erbel et al., 2007; Haude et al., 2013). However, these trials typically
involved small patient cohorts and simple lesions and therefore performance
comparison with commercially-available stents is difficult. Only limited performance
data has been reported for magnesium stents in general and in particular for WE43
alloy stents which, to date, represent the only magnesium stent platform to go
through coronary trials. It is known that the stent deployment characteristics, radial
stiffness, and cyclic strain amplitude in a stented artery segment influence ISR
outcomes (Colombo et al., 2009; Sangiorgi et al., 2007; Topol, 2008). As yet, these
performance characteristics have not been reported for WE43 stents.
Ideally, in vitro performance assessment should be conducted on the
finished stent or on samples of the precursor tube in the final form and finish owing
to the influence of processing, surface treatment and size effects on device
behaviour (Connolley et al., 2005; Murphy et al., 2003; Savage et al., 2004;
Wiersma and Taylor, 2004). It is noted that while a number of experimental studies
have reported on the mechanical and corrosion performance of magnesium test
specimens (Bowen et al., 2013; Coy et al., 2010; Grogan et al., 2011; Kannan et al.,
2008; Zhang et al., 2010) and simplified unfurled stent geometries that were cut
from sheets (Li et al., 2014; Wu et al., 2010), few studies have employed finished
magnesium stents as specimens (Wu et al., 2012) and no such studies have been
reported for WE43 stents. Many of the experimental studies of absorbable
magnesium materials in the literature employed specimens that had materials
processing, surface treatments and loading regimes that differed to those of
coronary stents (Bowen et al., 2013; Coy et al., 2010; Grogan et al., 2011; Kannan
et al., 2008; Zhang et al., 2010); Therefore, the mechanical and corrosion behaviour
of these specimens may not be representative of finished magnesium stents. Only a
limited about of information regarding the mechanical and corrosion behaviour of
WE43 stent precursor tubes is available (Witte et al., 2008). Fibre laser cutting and
42
surface treatment of balloon-expandable stents has been reported for the AZ31 alloy
(Demir et al., 2012; Wu et al., 2012); though no such study has been reported for
WE43.
While a number of studies have investigated the effects of mechanical
loading on the corrosion of magnesium specimens for near-constant loads (Grogan
et al., 2011) and non-constant loads (Kannan et al., 2008; Winzer et al., 2008), the
influence of the stent deployment procedure on the corrosion performance of
magnesium stents has not been extensively investigated. A number of experimental
studies have attempted to correlate the magnitude of stress in the magnesium
specimen to corrosion behaviour and have subsequently developed stress-mediated
corrosion models of magnesium stent performance (Gastaldi et al., 2011; Grogan et
al., 2011). Plastic strain, and the accompanying deformation-induced changes in
electrochemical measurements in magnesium alloys have been shown to have a
significant effect on galvanic corrosion processes, which are prominent in
magnesium alloys (Snir et al., 2012; Wittke et al., 2014).
Since the stent deployment procedure is known to induce large plastic strain
(20-30%) in stent struts (Murphy et al., 2003), we hypothesise that the plastic strain
that is induced in magnesium stents during the deployment procedure plays a
critical role in directing subsequent corrosion phenomena in the stent and therefore,
has an influence on the scaffolding ability of the stent in the long-term. Therefore, in
this work, plastic strain levels consistent with the deployment procedure were
induced the stents and specimens produced and the resulting effects on the
mechanical and corrosion performance of the specimens were explored in
experiments. The data were also used in the development and calibration of a
plastic strain mediated phenomenological corrosion model capable of simulating the
corrosion behaviour of magnesium alloy stents.
43
3. Materials and Methods
3.1
Summary
Absorbable magnesium stents were laser cut from precursor tubes of the WE43
alloy. Surface treatments were performed using a combination of chemical etching
and plasma etching processes. The manufacture and surface treatment of the
WE43 stents used in this study is discussed in detail in (Galvin et al., 2013).
The mechanical performance of the magnesium stent was compared to that
of a commercially-available 316L stent in three separate tests that determined stent
deployment, radial strength and cyclic strain amplitude metrics.
The corrosion behaviour of stent and dogbone specimens was determined
via immersion corrosion experiments performed in SBF under pseudo-physiological
conditions. The influence of the deployment procedure on the corrosion behaviour of
the stent was investigated in terms of temporal mass loss and corrosion-induced
loss of mechanical integrity.
A corrosion algorithm was written in ANSYS parametric design language
(APDL) and was developed for use with the ANSYS Implicit FE code. The corrosion
model, which was calibrated from the experimental data, was used to simulate the
corrosion performance of the magnesium stent. The results predicted by the model
were compared to those observed experimentally.
44
3.2
Stent Manufacture
3.2.1 Laser Cutting
Coronary stents were manufactured from stent precursor tubes of the WE43
magnesium alloy. An initial compositional analysis of the tube was performed to
confirm that the content of magnesium and the three main alloying elements (Y, Zr
and Nd) was consistent with that specified by the manufacturer of WE43
(Magnesium Elektron, 2007), as shown in Appendix A.1. The tubes had an outer
diameter of 2.1 mm and a wall thickness of 0.15 mm. The stent design was
approximated from published images of the Biotronik AMS-1 platform (Di Mario et
al., 2004). The unfurled stent geometry was first modelled in a computer-aided
design software package (SolidWorks Corp., MA. U.S.A), as shown in Figure 3.1(a).
Figure 3.1 (a) Two-dimensional CAD model of the unfurled stent geometry (b) SEM
image of a tube in which the stent pattern was machined
The computer-aided design data were then used to produce a computer
numerical control (CNC) program that defined all cutting paths necessary to sculpt
the stent by pulsed fibre laser machining (Starcut Tube, Rofin Baasel, UK Ltd.). The
laser beam had a wavelength of 1070 ± 5 nm and diameter of 13 µm. Nitrogen was
employed as an inert shield gas. A SEM image of a tube, in which the stent pattern
was machined, is shown in Figure 3.1(b).
45
Figure 3.2(a) shows the stent after laser cutting with the sections of waste
material removed. A significant amount of dross and spatter adherence was
observed on the inner surface of the struts, as shown in Figure 3.2(b). This dross
was subsequently removed by a combination of chemical and plasma etching
processes. The finished stent had a length of 11.94 mm and an outer diameter of
2.09 mm. The thickness of struts in the radial and longitudinal direction was 140 µm
and 120 µm, respectively. The average stent mass was 4.5 mg. The total surface
area of the stent was 78 mm2, as estimated from the computer model.
Figure 3.2 (a) Machined stent remaining after the inner sections had fallen out (b)
dross that formed on the inner surfaces as a result of laser cutting process
3.2.2 Chemical Etching
A chemical etching process was employed to remove the bulk dross, spatter and
heat-affected zone (HAZ) that were formed during laser cutting. This process also
rounded-off strut edges and improved the surface finish and stent deployment
characteristics. The stents were chemically etched in 85.2% phosphoric acid
(P6560, Sigma Aldrich) at 55°C for 20 seconds and then neutralised immediately by
immersion in sodium hydroxide solution (100g/L, S8045, Sigma Aldrich) for 30
seconds at 20°C. The stents were sonicated in ultrapure H2O for 2 minutes, followed
by alcohol for a further 2 minutes. Specimens were dried for 2 hours in a cleanroom
environment (class 100) using a glass drying oven (B-585, BUCHI Labortechnik AG)
at an air temperature of 25°C (Morshed et al., 2011).
46
3.2.3 Plasma Etching
In order to remove residues remaining from the chemical etching process the stents
were placed in the chamber of a reactive ion etching (RIE) instrument (PlasmaLab
80 Plus, Oxford Instruments). A schematic diagram of the experimental setup used
for the plasma etching process is shown in Figure 3.3.
Figure 3.3 Schematic diagram of the experimental setup used for the plasma
etching process
As RIE acts mainly as a “line-of-sight” process, etching was performed in
three stages where the specimens were manually rotated in increments of 120° on a
quartz plate in order to minimise shadowing effects. A capacitively-coupled plasma
was generated by a 13.56 MHz radio-frequency (RF) power source. The RFpowered electrode was made from stainless steel and had an outer diameter of 240
mm. The grounded electrode was made from stainless steel and acted as a
showerhead to supply the process gases. The distance between the powered and
grounded electrodes was 46 mm. Plasma etching was performed in two separate
47
process steps using two different process gases. In the first step, the surfaces of the
stents were plasma etched in a mixture of 90% argon (Ar) and 10% oxygen (O2) gas
for a total of 10 minutes. The comparatively heavy Ar ions removed material by a
micro-sandblast effect upon striking the substrate (Chu et al., 2002). In the second
step, the stents were cleaned and sterilised in 100% O2 gas for a further 10 minutes.
O2 plasma reacts with molecules deposited on the substrate breaking them down
and turning them into volatile compounds (Chu et al., 2002). In both cases the
process conditions were kept constant where the chamber pressure was 100 mTorr,
the power was 250 W and the flow of gas was 100 sccm. A bias voltage of ~495 V
was recorded.
3.2.4 Surface Characterisation
Specimens of the WE43 tube having a length of 10 mm were laser cut, then
degreased by sonication in alcohol and finally placed in a glass drying oven for 2
hours at 25°C. The tube specimens were subjected to the same chemical and
plasma etching processes that were used to produce the stents. Specimens were
prepared in three different conditions, namely as-received, chemically etched and
plasma etched. Scanning electron microscopy (SEM), Atomic Force microscopy
(AFM) and Energy Dispersive X-ray Spectrometry (EDX) analyses were performed
on the tube specimens in order to characterise the morphology, roughness and
chemical composition of surfaces for the three different conditions.
EDX Analysis
Prior to EDX analysis all specimens were sputter-coated with a thin film of carbon in
order to render them conductive. EDX analysis was performed using an INCA
Energy 350 XT microanalysis system (Oxford Instruments) in conjunction with a
SEM (Leo 440 Stereo Scan, Leica Cambridge Ltd). The vacuum in the chamber was
2.0 x 10-6 mbar. The beam current and spot size of the electron gun was set to 1.0
48
nA and 520 nm, respectively. The working distance between the specimen and
electron gun was maintained between 8.5 and 9 mm and the collection time for the
EDX detector was 120 seconds. Quantitative EDX analysis was performed using the
INCA Microanalysis Suite (v4.09) software package. Background correction was
performed using the background modelling method where the detector background
spectrum was subtracted from the observed spectral distribution by mathematical
filtering using a top hat function. The intensity of spectral peaks corresponding to
specimen elements were measured and compared to standards measured under
known operating conditions by the software. Intensity ratios (k values) were
calculated and converted into chemical concentration (weight %) for the elements
detected within the specimen.
Surface Roughness Analysis
The surface texture of treated and untreated specimens was scanned using high
resolution contact mode AFM (Nano-R™ O-020, Pacific Nanotechnology). A
cantilever AFM tip (SICON-10P, Applied NanoStructures Inc.) was drawn along the
outer surface of each specimen in the longitudinal direction. Three randomly
selected regions, measuring 80 µm x 80 µm, were scanned on each specimen. The
scan rate and number of lines in each scan were kept constant at 0.5 Hz and 256,
respectively. The surface roughness parameter (Sa) was calculated without a
filtering process using the SPM CockpitTM (V3.3.383) software package according to
the formula listed in ASME B46.1. A one-way analysis of variance (ANOVA) was
used to test for differences among the surface roughness values (Sa) of the three
specimen groups (as-received, chemically etched and plasma etched).
3.3
Mechanical Testing
Mechanical testing was carried out in order to determine deployment and cyclic
strain amplitude metrics for the stent. A commercially-available generic modular
49
316L stent design (outer diameter (OD) = 3.5 mm, length = 9 mm, strut thickness =
0.09 mm) was also tested for comparison, as shown in Figure 3.4(b). Prior to
deployment, each WE43 stent was mounted (uncrimped) on a 3.5 x 18 mm
angioplasty balloon. The 316L stents were already crimped to the balloon (3.5 x 18
mm) on their delivery system when supplied. Stent performance was assessed in
three different bench tests as described below. The tests were captured by
videoextensometer and calibrated image sequences were used to calculate stent
performance metrics. In each one of the three tests, three WE43 stents (n=3) and
three 316L stents (n=3) were used.
Non-uniform deployment may be encountered with manually crimped stents
(Morton et al., 2004) and was likely to occur with the WE43 stent due to its highlycompliant nature. Therefore, in this work, all WE3 stents were expanded without
manual crimping to the angioplasty balloon in order to ensure a uniform deployed
stent configuration. The crimping process induces significant compressive loading
and deformation in stents and is known to change the subsequent stress-strain
behaviour of the strut material during the deployment procedure, resulting from the
tension-compression reversal behaviour (Bauschinger Effect) (Oberhofer et al.,
2006). Thus, compared to uncrimped stents, crimped magnesium stents would likely
exhibit slightly reduced mechanical properties and altered fatigue and corrosion
behaviour owing to the influence of material deformation and residual compressive
stresses (Zeng et al., 2008). The use of crimping machines may be necessary with
magnesium stents in order to reduce asymmetries in the deployed configuration
which have been associated with increased vascular injury and higher rates of IH
and ISR (Schulz et al., 2000). In a preliminary FE study carried out in this work, a
crossing profile of 1.3 mm was predicted for the WE43 stent which was comparable
with that observed for the Biotronik AMS-1 (1.2 mm) (Di Mario et al., 2004).
50
Test 1: Free Expansion
Stents were freely expanded to a maximum balloon pressure (15 atm) and then
permitted to undergo elastic recoil. Expansion behaviour was quantified through the
determination of a pressure-diameter curve where the balloon pressure and the
corresponding stent OD were tracked during the free expansion procedure. The
stent diameter was calculated from the average of eight separate OD
measurements made along the length of the stent, as shown in Figure 3.4(a). In
order to account for differences in the strut thickness of the two stent types, stent
performance metrics were calculated using stent ID data. This was derived from the
OD measurements by subtracting the strut thickness in the radial direction.
a
b
Figure 3.4 Free expansion of stents (a) Location of 8 stent OD measurements taken
using a videoextensometer during bench testing (b) 316L stent
Test 2: Expansion in Mock Coronary Arteries
Stents were expanded inside mock coronary arteries (MCAs) that were prepared
using a technique described in an earlier work (Colombo et al., 2010). The MCAs
had an inner diameter of 3 mm and a wall thickness of 0.4 mm. A pre-stretch ratio of
1.1 was applied to each MCA using a custom fixture fitted with a linear translation
stage and micrometer thereby mimicking the arterial tethering observed in vivo.
Elastic longitudinal recoil (ELR), foreshortening (FS) and dogboning (DB) metrics
were calculated using previously described formulae (Migliavacca et al., 2002).
Elastic radial recoil (ERR) was calculated by determining the average recoil at eight
51
separate locations along the stent length, as shown in Figure 3.4. It was not possible
to accurately measure the stent ID or OD by visualisation owing to the opacity of the
MCA material. Therefore, ERR was calculated from measurements of MCA OD.
Figure 3.5 Stent expansion in mock coronary arteries (a) WE43 stent (b) 316L stent
Test 3: Cyclic Strain Amplitude
The cyclic strain amplitude supported by expanded stents in MCAs was assessed,
as shown in Figure 3.6. A static pressure head was applied to each stented MCA by
elevating a tank of non-corrosive solution to a predetermined height. The tank was
connected to the stented MCA using silicone tubing and luer lock fittings. Each
stented MCA was subjected to three pre-conditioning pressure cycles (0 to 120 mm
Hg). The OD of the stented MCA was recorded at eight separate locations while
under the influence of three different static pressure loads, namely: 0, 80 and 120
mm Hg. Cyclic strain amplitude was calculated as the % change in the average OD
of the stented MCA between a pressure of 80 mm Hg and 120 mm Hg.
Figure 3.6 Stented mock coronary artery under the influence of static pressure head
during the experiment performed to measure cyclic strain amplitude
52
3.4
Corrosion Experiments
The corrosion behaviour of dogbone and stent specimens was characterised
through two independent immersion corrosion experiments performed in SBF, as
detailed in Table 3.1. A total of eighty four dogbone specimens and twenty one
stents were prepared for corrosion testing by laser cutting and finished using the
surface treatments described in section 3.2.
Table 3.1 Details of the experiments performed (PCM: phase contrast microscopy)
The main purpose of Experiment A was the determination of the influence of
plastic strain on the corrosion behaviour of the alloy. The corrosion rates, the
dominant corrosion processes and the temporal corrosion-induced reduction in the
mechanical integrity of the specimen were determined. Specimen mass loss was
obtained by gravimetric measurements performed using a semi-microbalance
having a resolution of 0.01 mg (Sartorius RC210P). The results of a preliminary
corrosion experiment indicated a maximum specimen mass loss of almost 100%
after 10 days of immersion. Assuming the requirement of having 500 increments (1
increment = 0.01 mg) of measurement available within the expected experimental
mass loss range, a specimen gauge region with a mass of approximately 5 mg was
deemed suitable. As WE43 had a relative density of 1.84, the corresponding volume
53
in the specimen gauge region was 2.73 mm3. The wall thickness of the specimen
was the same as that for the finished stent. In accordance with ASTM standard E8,
which requires a minimum gauge length (L) to width (W) ratio of four for pin-loaded
tensile specimens, an L/W ratio greater than five was selected in this case. Thus,
the finished dogbone specimen, which is shown in Figure 3.7(a), had a gauge region
with a length of 10.05 mm, (unfurled) width of 1.85 mm, wall thickness of 0.14 mm
and a total surface area of 40 mm2. It is noted that the L/W ratio used in this work is
similar or greater to those of previous studies involving tensile testing of sub-size
metal specimens (Jain et al., 2008; Raulea and Goijaerts, 2001). Thickness
measurements of the gauge region, performed using a SEM, verified that a uniform
specimen wall thickness was obtained from the surface treatments. The average
roughness (Sa) of outer specimen surfaces was 0.17 µm, as measured using AFM.
In order to investigate the influence of plastic strain on corrosion behaviour,
three groups of dogbone specimens were prepared, each subjected to a different
magnitude of plastic strain, namely: 0%, 2.7% and 8.4%. Specimens having 0%
plastic strain required no deformation. The other specimens were mounted in
custom grips in a tensile testing machine (Z005, Zwick Roell) and extended for a
time of 10 seconds by either 4% or 10% of their original length using a crosshead
speed of 0.5 mm/min, as indicated by the stress-strain curves shown in Figure
3.7(b). At peak extension the direction of crosshead travel was reversed in order to
permit elastic recoil of the specimen and the test was terminated when the load cell
recorded zero force. In this way, the specimens were subjected to a loading cycle
analogous to that induced in stent struts (loading and elastic recoil) during the
deployment process. After recoil, the plastic strain (PS) for the specimen groups
(mean ± SD) was 2.66 ± 0.03% and 8.38 ± 0.02%; for simplicity these specimen
groups were named 2.7% PS and 8.4% PS, respectively. The tab regions on each
specimen were coated with an epoxy resin to limit corrosion to the test region.
54
Figure 3.7 (a) The dogbone specimen geometry (b) Representative stress-strain
response of dogbone specimens during the application of plastic strain
The main purpose of Experiment B was to determine the corrosion rate for
expanded WE43 stents and to determine the influence of corrosion on the radial
stiffness of the stent. Each stent was mounted (uncrimped) on a 3.5 x 18 mm
angioplasty balloon, freely expanded to a maximum balloon pressure (15 atm)
consistent with clinical deployment of magnesium stents (Erbel et al., 2007) and
then permitted to undergo elastic recoil.
Prior to corrosion testing, stent and dogbone specimens were stored in a
desiccator for 24 hours to ensure consistent oxidisation of all surfaces and then
sterilised by exposure to ultraviolet irradiation for 2 hours. Specimens were mounted
on a stage (Multiclip, Struers) so that all test region surfaces were exposed to the
solution. Each specimen was placed in a separate container with 25 ml Hank’s
balanced salt solution (H8264, Sigma Aldrich) and the screw cap of the container
was left slightly open in order to allow gaseous exchange with the environment. The
ratio of solution volume (ml) to exposed specimen surface area (cm2) was 63:1 and
32:1 for dogbone and stent specimens, respectively; well above the minimum ratio
55
recommended for magnesium alloys (Yang and Zhang, 2009). Since corrosion
phenomena in magnesium alloys are sensitive to variations in environmental
conditions such as temperature (Kirkland et al., 2010), solution pH (Ng et al., 2010)
and biological contaminants, the corrosion studies were performed under sterile
conditions in an incubator (Nuaire NU-5500E) that provided temperature uniformity
(37 ± 0.2°C) and a humidified gas environment (5% CO2 and 95% air) for increased
pH buffering capacity.
The pH of the SBF was recorded at regular time intervals (pH 213, Hanna
Instruments) and maintained within a physiologically-relevant range of values (6.3–
7.8) as described in previous studies involving similar alloys and solutions (Bowen et
al., 2013). Specimens were removed at various intervals of time, cleaned in a
solution of 10% chromic acid (27081 Sigma-Aldrich) at 55°C in order to remove
corrosion products and weighed using a semi-microbalance. For each interval of
time four dogbone specimens (n=4) and three stents (n=3) were used. Corrosion
surfaces were imaged under the SEM. The thickness of corroded struts was
measured at 30 separate locations under the SEM, from which the mean strut width
was calculated. EDX analysis was used to determine the elemental composition of
the corrosion products and surfaces. Specimens were imaged in solution using a
phase contrast microscope (PCM) (Olympus CK30) in order to observe the evolution
of H2 gas bubbles.
Mechanical strength testing was performed to determine the influence of
corrosion on the mechanical integrity of the specimens. For dogbone specimens,
uniaxial tensile testing was carried out to failure in one loading cycle using a
materials testing machine (Z005, Zwick Roell) with a 5 kN load cell. A pre-load of 0.5
N and a crosshead speed of 0.5 mm/min were used. Tests were terminated when
the load dropped below 70% of the maximum load value. Specimen elongation was
determined from crosshead travel, rather than from local deformation at the test
56
region of the specimen. Specimen stresses were calculated based on the original
specimen cross-section and so the changes in cross-section due to necking and
corrosion were not accounted for; this approach having been used previously in
similar studies (Bowen et al., 2012). Engineering stress-strain curves were produced
for each specimen. The elastic modulus (E), 0.2% offset yield strength (σy) and
ultimate tensile strength (UTS) were calculated in accordance with ASTM E111-04
using the TestXpert software package (V11.02). The value of strain at the UTS and
at failure was termed the ultimate strain (εu) and failure strain (εf), respectively. The
radial stiffness of stents, defined as the change in resistive force for an applied 10%
reduction in stent diameter, was evaluated by compressing freely expanded stents
between the platens of a materials testing machine (Zwick Z005) using a 500 N load
cell, a preload of 0.02 N and a cross-head speed of 0.2 mm/min. Again, the 316L
stent was tested for comparison. Data were normalised for the 316L stent in order to
account for the difference in original length between the 316L and WE43 stent and,
curves were presented as stent reaction force versus stent compression.
3.5
Corrosion Modelling
A FE based plastic strain mediated phenomenological corrosion model, which was
capable of modelling the uniform micro-galvanic corrosion of a 3-D metallic stent
structure, was developed in order to assess the performance of a corroding WE43
alloy stent. The model was used to predict the effects of diffusion-controlled
corrosion in elastic-plastic metals and, was calibrated through corrosion experiments
that established the plastic strain-mediated input corrosion rates for the material of
interest. The corrosion model was written in ANSYS parametric design language
(APDL), as shown in Appendix A.3 and, was developed for use with the commercial
ANSYS Implicit FE code. APDL is a scripting language specific to ANSYS that
facilitates automation of common tasks, parameterisation of models and, allows the
use of other sophisticated features, such as optimisation techniques (Ansys, 2010).
57
The corrosion model developed represents a modification of the CDM model
proposed by Gastaldi et al. (Gastaldi et al., 2011) which allows the effects of
corrosion-induced micro-scale geometric discontinuities on overall specimen
mechanical integrity to be accounted for, without explicitly modelling their
progression. However, in this study, a discrete rather than continuum modelling
approach was adopted. A flowchart outlining the operation of the corrosion model is
shown in Figure 3.8. After the solution to an initial analysis was obtained (stent
deployment and recoil), the results data, which included the magnitude of the first
principal plastic strain, εp1, in each element, was written to arrays in preparation for
corrosion simulation. Element connectivity was determined and the elements on the
corrosion surface were identified. The material behaviour of each element was
initially modelled by a single elastic-plastic material model that was representative of
the mechanical properties of the material prior to the onset of corrosion.
The loss of mechanical integrity due to the corrosion process, denoted by the
damage parameter, D, was calculated on an element-by-element basis through a
modified version of the damage evolution law described by Gastaldi et al (Gastaldi
et al., 2011). The temporal value of the damage parameter, D, was calculated over a
discrete volume of material governed by the element size, Le and, over a discrete
time-step, ∆t. When Le and ∆t were sufficiently small; the damage parameter at a
fixed time, De, was given by the relation:
where ߜ and Le were respective material and FE model characteristic lengths, in
units of mm, and ku was a corrosion kinetic parameter, in units of h-1, as described
previously in (Gastaldi et al., 2011).
58
Figure 3.8 Simplified flowchart of the corrosion model process
59
An element-specific dimensionless plastic strain parameter, φe, was
introduced in this work in order to account for the effects of plastic strain-induced
corrosion in the modelling framework. The value of φe for each element was
dependent on both the magnitude of elemental εp1 and the simulated corrosion time,
t, and was estimated from a table of values that were established in the
experimental corrosion tests described in section 3.4.
Depending on the number of element faces (NEF) that were exposed to the
corrosion environment, the element damage calculated at each time step was
modulated by an element-specific exposure parameter, κe. κe was a simple scaling
parameter that was derived from a linear approximation of the slope of the curve
that described the rate of diffusion through a brick element when NEF varied from
one to six, as shown in Figure 3.9. κe was included in the modelling framework in an
attempt to model the rounding of initially sharp strut corners which can be expected
for a diffusion-dominated corrosion process (Ulvrová et al., 2012).
Figure 3.9 The possible values for the element exposure parameter, κe, that were
used to scale the element damage calculated at each time step
When the damage parameter, D, reached or exceeded certain threshold
values, the mechanical integrity of the element was reduced by modifying its
60
material model. In this way, the gradual corrosion-induced reduction in the
mechanical integrity of the elements was modelled. When the damage parameter
reached a critical value (D=1), the element was assumed to have completely
corroded and was eliminated from the model. The newly exposed element surface
beneath the eliminated element became part of the exposed surface and was thus
subject to corrosion in the next iteration of the model. The model iterated until the
simulated corrosion time, t, reached the user-defined maximum solution time, tmax.
3.5.1 Corrosion Model Calibration
The corrosion model was calibrated based on the results of experiment A. As the
corrosion-induced material loss observed in the experiments was relatively uniform
in both the longitudinal and radial directions of the test region, repetitive and
reflective symmetry simplifications were assumed and only a one-quarter slice of the
dogbone specimen was modelled, as shown in Figure 3.10(a). A representation of
the test region in the specimen, which was modelled as being flat rather than
curved, was meshed with enhanced strain formulation brick elements, as shown in
Figure 3.10(b). Symmetry boundary constraints were applied to the nodes on the
two symmetry edges of the model. Nodes on the upper and lower surfaces of the
model were constrained only in the direction of loading which permitted lateral
constrictions as would be expected under uniaxial tensile loading. All nodes on the
top surface were coupled in the direction of loading to a master node and a force
was applied to this master node in order to simulate the tensile test.
In representing the deformation of the material in the FE model, finite
deformation kinematics was assumed. The initial mechanical properties were based
on those of non-corroded WE43 dogbone specimens with 0% plastic strain, as
determined from the uniaxial tensile tests described in section 3.4. Accordingly,
elasticity was considered linear and isotropic in terms of finite deformation quantities
61
with E = 13 GPa. Material plasticity was described by a multi-linear isotropic
hardening material model, in accordance with previous FE studies which used
similar material models (Gastaldi et al., 2011; Wu et al., 2011), with σy = 160 MPa,
UTS = 245 MPa and εu = 0.17. Poisson’s ratio had a value of 0.3 (Wu et al., 2010).
Figure 3.10 (a) slice of dogbone specimen test region exhibiting repetitive and
reflective symmetry (b) boundary and loading conditions used for the FE simulations
The material properties of damaged elements were controlled through the
evolution of the damage parameter, D. The corrosion-induced reduction in the
mechanical integrity of the material, observed in the experiments, was characterised
in the model by modifying the material model of the elements at certain threshold
values of D. Ten material models were used to represent the material properties of
elements with D values ranging from 0 to 1, in increments of 0.1, as shown in Figure
3.11. For the stress-strain curves shown, it was assumed that the reduction in the
load-bearing capacity of the element was directly proportional to the mass lost to
corrosion. Therefore, the curves were produced by scaling the stress component of
62
curve for the WE43 material by factors ranging from 0.9 to 0.1 in increments of 0.1.
Figure 3.11 FE material model input used to represent the stress-strain properties of
elements with D values ranging from 0 to 0.9
Experiment A was modelled by simulating tensile loading and elastic recoil of
the test region in order to induce the required magnitude of εp1 (0%, 2.7% and
8.4%). Corrosion was then permitted to occur on exposed surfaces in the test
region. The rate of mass loss over time was determined based on the average
damage value (D) over all of the model elements. The maximum force withstood by
the model was taken as a measure of loss in mechanical integrity. In calculating the
stress-strain response of the model to tensile loading, the engineering stress was
calculated based on the original cross-section of the specimen in order to facilitate
comparison with the experimentally measured response.
The material characteristic length, ߜ , was given a value of 0.015 mm,
consistent with grain sizes observed in metallographic analysis of WE43 tubes. The
slope of the mass loss curves from experiment A was used to generate a table of φe
values for the three plastic strain magnitudes examined, as shown in Table 3.2. By
interpolating between the data points in the table, a φe value was calculated for each
63
element based on the magnitude of εp1 in the element and simulated corrosion time.
Table 3.2 The table of values established in experiment A from which the strain
parameter, φe, was calculated in the FE corrosion model
In calibrating the FE corrosion model, the values of the independent model
parameters ku, Le and ∆t were determined for given values of ߜ , φe and κe, based on
the simulation of experiment A. These values were determined by imposing three
conditions. First, the predicted temporal mass loss curves should match that
observed in experiment A. Second, the simulated corrosion-induced reduction in
specimen mechanical integrity should match that observed experimentally. Third,
the model should predict the corrosion rates observed in experiment A, thus
allowing the parameter kU to be determined. These conditions allowed the
determination of suitable element length (Le = 0.012 mm) and time step (∆t = 1 h)
values through an iterative calibration process, as shown in the convergence study
in Appendix A.4.1.
In order to validate the predictive capabilities of the corrosion model for the
calibrated parameters, the corrosion of the dogbone specimens observed in
experiment A was simulated. Loading and elastic recoil of the sample in uniaxial
tension was simulated followed by the subsequent corrosion of the specimen under
the influence of 0%, 2.7% and 8.4% plastic strain. Model predictions were compared
to the experimental results.
64
3.5.2 Stent Model
The calibrated corrosion model was used to predict the corrosion performance of the
absorbable magnesium stent. The design and geometry of the stent were taken
directly from the experiments, as shown in Figure 3.12(a). The stent geometry,
which exhibited rotational and circumferential symmetry, as well as periodicity in the
longitudinal direction, was represented by a repeating unit cell (RUC) model that
was meshed with 91,320 brick elements, as shown in Figure 3.12(b). The RUC
method is an established technique which can be used to reduce model size and
solution time in the numerical analysis of models balloon-expandable stents (Ju et
al., 2008; Li et al., 2009; Xia et al., 2007). The element type and material models
used for the stent were the same as those described earlier for the dogbone
specimen model in section 3.5.1.
In order to reduce the computational cost of modelling the interaction of the
balloon and stent, spring elements were incorporated to simulate the constraint of
the balloon using an established technique (Early et al., 2009; Zahedmanesh et al.,
2010). The springs had multi-linear stiffness and were attached to the nodes on the
inner surface of the stent. The springs had low stiffness (1 x 10-7 N/m) until the stent
reached nominal diameter at which point the stiffness increased sharply (0.1 N/m) to
inhibit further stent expansion. Periodic boundary conditions were imposed on the
symmetry planes of the unit cell in the rotational and longitudinal directions using
nodal displacement constraint equations, as outlined in (Xia et al., 2007). A suitable
average element size (Le = 0.012mm) and time step (∆t = 1 hour) was established by
means of a convergence study in which predicted nodal displacement, element
stress and strain, mass loss and stent radial stiffness results were all converged to
within 2%, as shown in Appendix A.4.2. Based on the selected Le, the corrosion
model parameters were scaled accordingly. Stent material properties were taken to
be the same as those used in the dogbone specimen models (see Figure 3.11).
65
Figure 3.12 (a) Image of the full stent geometry with a repeating unit cell (RUC)
indicated (b) The geometry of the FE RUC model prior to deployment (c) Loading
applied to contact elements in order to predict the radial stiffness of the stent
The deployment of the uncrimped stent, by way of free expansion, was
simulated through the application of a uniform linearly-increasing radial pressure at
the internal surface of the stent. When the outer diameter of the stent reached a
value similar to that measured experimentally, the pressure load was reduced to
zero in order to simulate elastic recoil. The rate of mass loss was determined based
on the average damage value (D) over all of the model elements. At 20 hour
intervals of simulated corrosion time, a flat analytical surface comprising of contact
elements, was brought into contact with the outer surface of the stent and was used
to compress the stent in the same manner as that used in the experiments, as
shown in Figure 3.12(c). Again, the radial stiffness of the stent was calculated as the
change in resistive force for an applied 10% reduction in stent diameter.
66
Stent Model Calibration
Based on the mass loss curves obtained experimentally for plastic strain values of
0%, 2.7% and 8.4%, the corrosion model predicted the value of φe for each element
by interpolating, for the εp1 value in the element, between the data points of the
experimental curves. However, preliminary stent deployment simulations showed
that a small proportion of the elements (typically less than 4%) had εp1 values,
ranging from 8.4% to 27%, that were outside the range measured experimentally, as
shown in the distribution plot in Figure 3.13(a). In order to investigate the
relationship between εp1 and φe for these elements, two different corrosion models
were developed. In the linear plastic strain (PS) corrosion model, a linear
relationship was assumed between εp1 and φe above εp1 values of 8.4%. In the plastic
strain corrosion model, a non-linear relationship was assumed, where φe increased
sharply for elements with εp1 values above 8.4%; with substantially higher φe applied
when εp1 exceeded the engineering strain observed experimentally at the UTS
(~20%). For both models, the results predicted for mass loss and radial stiffness
were compared to the experimental data in order to determine the model that gave
the best correlation.
Figure 3.13 (a) Distribution of the elemental εp1 values for the stent model (b) Linear
and non-linear models used to represent the dependency of φe on elemental εp1
67
4. Results
4.1
Stent Manufacture
4.1.1 Laser Cutting
Laser cutting of WE43 tubes produced parts with substantial dross and spatter
attachment (see Figure 3.2). Dross may form during laser cutting when a significant
temperature gradient exists between the top and bottom surface of the work-piece.
This results in a greater amount of material being melted on the underside of the
work-piece. The low viscosity and low surface tension of melted magnesium also
may have contributed to the spatter attachment observed. Excessive amounts of
dross and spatter attachment on the inner surface of the tube prevented the normal
fallout of cut sections. Therefore, machining parameters were optimised for cutting
the material including the laser power, pulse repetition frequency, pulse width, laser
wavelength and cutting speed, as shown in Table 4.1.
Table 4.1 Optimised laser machining parameters for cutting WE43 tube
Nitrogen was selected as an inert shield gas in order to minimise oxidation
and gas pressure was increased to improve the quality of the cut and to reduce
dross accumulation. The focal position of the laser beam was adjusted to increase
the kerf width and hence aid fallout of cut sections. Optimisation of laser parameters
successfully reduced dross and spatter adherence. The remaining dross was
removed by a combination of chemical and plasma etching processes.
68
4.1.2 Chemical and Plasma Etching
There was an observable difference in the surface texture of inner and outer
surfaces which appeared to originate from the extrusion process used to produce
the tubes. Outer surfaces had the smoothest appearance while inner surfaces
appeared marginally rougher. Surfaces produced by laser cutting had the roughest
appearance. Chemical etching with phosphoric acid removed bulk dross and surface
scale from all stent surfaces. A layer of material measuring approximately 5 µm was
removed, as measured under SEM. The texture of all stent surfaces was
significantly improved to a more uniform morphology, as shown in Figure 4.1.
Significant variation was evident in the surface texture of inner, outer and laser cut
surfaces after chemical etching. A significant amount of residue and contamination
remained on the stent surfaces following the chemical etching process, as shown in
Figure 4.1.
Figure 4.1 Stent surfaces resulting from chemical etching: (a) outer stent surface
with residue (dark areas), and (b) inner surface of stent strut
Plasma etching and cleaning was employed in order to remove the surface
deposits and contamination that remained from the chemical etching process. SEM
and photographic images of a finished stent are shown in Figure 4.2 and Figure 4.3,
respectively. It can be seen that plasma etched surfaces appeared lighter in colour
compared to chemically etched surfaces and residues were completely removed.
69
Plasma processing reduced surface roughness of all surfaces compared to
chemically etched surfaces. The plasma etching process was most effective on the
outer surface and less effective on shadowed surfaces and laser cut edges, as
shown in Figure 4.2. Plasma etching was performed in three stages where the
specimens were manually rotated in increments of 120° on the quartz plate in order
to minimise shadowing effects. Although RIE acts mainly as a “line-of-sight”
process, it is likely that a significant number of ions would have reached the inner
stent surfaces owing to the small strut width (120 µm), the comparatively large
spacing between the struts, and the high sheath potential (~495 V). The specimen
rotation technique employed here would have limited efficacy in reducing the
shadowing effects. A planetary rotating system would provide a more effective
method to minimise shadowing effects through continuous specimen rotation during
plasma processing.
The etch rate of the 90% Ar and 10% O2 plasma process was not measured
as it was beyond the scope of this work. However, one would expect a significantly
higher etch rate for the WE43 alloy when compared to relatively harder traditional
stent materials such as stainless steel (316L), cobalt-chromium alloys and titanium
alloys. Etch rates of 20 to 30 nm/min in 100% Ar plasma have been reported for
pure magnesium (Nishimoto, 2003) and MgO thin films (Kim and Kim, 2007). These
studies used higher power settings (300 - 500 W) and higher proportions of Ar
compared to those used in this work. The use of comparatively lower power settings
and a lower proportion of Ar in this study suggest etch rates well below 20 nm/min
for the WE43 alloy material. Therefore, one may expect that a layer of material no
more than 0.3 µm thick was removed from the stent with a processing time of 10
minutes.
70
Figure 4.2 SEM images of a finished stent following plasma etching and cleaning:
(a) stent mounted on fused silica tube, (b) inner surface, and (c) outer surface
Figure 4.3 Image showing a finished WE43 stent
71
4.1.3 Surface Characterisation
Surface Chemistry
Figure 4.4(a) shows that the as-received specimens had a relatively rough surface
morphology that consisted mainly of a dark-coloured magnesium (Mg) matrix area
(α-phase) interspersed with lighter-coloured β-phase areas (Hänzi et al., 2009). The
chemical compositions of both the α-phase and β-phase regions of the specimen
were measured by EDX at three different sites. Figure 4.4(a) shows the sites where
spectra were taken for α-phase (S1-S3) and β-phase (S4-S6) measurements. The
weight percent (%) of each element detected at the three sites was averaged
(mean) in order to produce the plot shown in Figure 4.4(b). The error bars were
calculated as the standard deviation in order to represent the overall distribution of
the data. It should be noted that, as with many EDX systems, the system used in
this work had inherent difficulty measuring light elements owing to low fluorescence
yield, absorption and peak overlaps with heavier elements (Berlin, 2011). Therefore,
the readings for C and O obtained by EDX in this work should be disregarded.
Figure 4.4 (a) Typical surface morphology of as-received specimen showing spectra
sites S1-S6, and (b) average chemical composition at spectra sites S1-S3 and
spectra sites S3-S6
72
EDX analysis confirmed the α-phase had high concentrations of Mg (62%)
and low concentrations of distributed alloying elements Y (5.6%), Nd (6.2%) and Zr
(1.0%) along with O2 (19.4%) and C (6.1%). Conversely, the surface of the
secondary phase areas had low Mg (16.7%) and relatively higher concentrations of
alloying elements Y (19.6%), Nd (30.3%) and Zr (1.3%) along with O2 (28.3%) and
C (4.0%). A detailed characterisation of the microstructure of a similar wrought
WE43 alloy is given in (Coy et al., 2010).
Chemical etching in phosphoric acid resulted in significant changes in the
appearance and surface chemistry of the specimens. The native oxide layer was
replaced with a more passive oxide layer with lower porosity and increased
corrosion resistance. The surface morphology was more homogenous owing to the
lower porosity of the new oxide layer, as shown in Figure 4.5(a). EDX analysis
showed that the new surface layer consisted of Mg (91.3%), Y (0.8%), Zr (0.1%), O2
(6.0%) and C (0.6%) and small amounts of Na (1.0%) and P (0.1% – 0.85%). It is
believed that the presence of P in the surface layer can be attributed to the
formation of magnesium phosphate during etching (Gray-Munro et al., 2009). The
presence of Na in the surface layer may be attributed to the formation of a thin layer
of sodium hydroxide that is adsorbed to the specimen during the neutralisation
treatment.
A number of rectangular-shaped entities were distributed within the surface
layer of the chemically etched specimens, as shown in Figure 4.5(a). An EDX
spectrum at S2 showed that these entities were rich in the RE elements Y (80.3%),
Nd (9.9%) and Zr (0.8%). The dark surface deposits shown in Figure 4.5(b)
consisted mainly of O2 (46-48%), Na (19.7-28.5%), P (18-25.1%) and Mg (1-10%).
These heavy deposits were likely to have been oxides with P and NaOH content as
this would account the high O2 content observed at these sites.
73
Figure 4.5 Typical surface morphology of chemically etched specimen: (a) normal
surface appearance, (b) dark area with surface residue, and (c) chemical
composition of surface at spectra sites S1-S4
Plasma etching successfully removed unwanted surface deposits and
residue, as shown in Figure 4.6(a). EDX analysis showed lower concentrations of P
and Na on plasma etched surfaces compared to chemically etched surfaces, as
shown in Figure 4.6(b). These results suggest that the heavy Ar ions in the plasma
may have removed some of the phosphor-rich coating. The amount of material
removed from the surface can easily be reduced by shortening the treatment time or
by increasing the ratio of O2 to Ar in the gas mixture.
74
Figure 4.6 (a) Typical surface morphology of plasma etched specimen showing EDX
spectra sites S1-S6, and (b) average chemical composition of surface at spectra
sites S1-S6
Surface Roughness
Figure 4.7 shows the surface morphologies measured by AFM for the as-received,
chemically etched and plasma etched specimens. Figure 4.8 shows average surface
roughness values (Sa) for the same three specimen conditions. It was found that
plasma etched specimens had the lowest surface roughness (172 nm) compared to
chemically etched (191 nm) and as-received (223 nm) specimens. A one-way
analysis of variance (ANOVA) showed that the mean surface roughness amongst
the three specimen groups (as-received, chemically etched and plasma etched) was
statistically significant (F(2,6) = 5.14, p = 0.006).
75
Figure 4.7 Typical surface morphology of specimens generated by AFM: (a) asreceived, (b) chemically etched, and (c) plasma etched
Figure 4.8 Surface roughness (Sa) values (mean ± standard deviation) measured by
AFM for as-received, chemically etched and plasma etched tube specimens
76
The tubes used in this work were subjected to an annealing process. The
average surface roughness (Sa) was relatively high in the as-received condition
owing to the non-uniform morphology during the tube extrusion process. Figure
4.9(a) shows the parallel lines and defects that ran longitudinally on the surface
indicating the direction of the tube extrusion process. These defects were potential
sites of stress concentration and material failure due to crack growth during stent
expansion. The chemical and plasma processing steps ware intended to minimise
these defects. The surface texture was considerably improved by chemical etching
which reduced the Sa value by 14.7% from 223 nm to 191 nm. No surface defects
were evident on the surface layer that was produced by chemical etching, as shown
in Figure 4.9(b). The plasma etched surface was considerably brighter in
appearance compared to the surfaces of the other two specimens, as shown in
Figure 4.9(c). The average surface roughness of plasma treated specimens (172
nm) was 10% lower compared to chemically etched specimens (191 nm).
Figure 4.9 SEM images of tube specimens: (a) as-received, (b) chemically etched,
and (c) plasma etched and cleaned (magnification = 1000x)
77
4.2
Mechanical Testing
In the first test, free expansion of the WE43 stent occurred mainly between balloon
pressures of 1 atm and 3 atm, while the equivalent range for the 316L stent was
between 2 atm and 6 atm, as shown in Figure 4.10(a). Owing to high stent
compliance, the proximal and distal ends of the WE43 stent flared significantly at
low balloon pressure thus contributing to the rapid increase in ID observed. At the
maximum balloon pressure of 15 atm, the average ID for WE43 (4.01 mm) was 9%
greater than for 316L (3.68 mm). The inset in Figure 4.10(a) shows the typical
deformed geometry of a WE43 stent at three different phases of the free expansion
procedure, namely: unexpanded, fully expanded and after elastic recoil. ERR for
WE43 (4.4%) was higher compared to the 316L stent (3.6%), as shown in Figure
4.10(b). FS was significantly lower for WE43 (0.84%) compared to 316L (3.55%).
Negative DB was greater for WE43 (-2.4%) compared to 316L (-0.52%), thereby
indicating a larger central ID than distal ID at maximum balloon pressure, for both
stent types.
In the second test, WE43 and 316L stents were expanded inside MCAs, as
shown in Figure 4.10(c). Stent metrics were calculated based on measurements of
MCA OD as the stent ID could not be accurately measured by visualisation.
Compared to free expansion results in the first test, higher ERR and lower DB
values were recorded for both stent types, as shown in Figure 4.10(b), owing to the
presence of the MCA which applied an inward-acting radial force to the stent.
Compared to the first test, FS values were higher for WE43 and lower for 316L
stents when deployed in MCAs.
In the third test, under the influence of sequential static physiological
pressure loads (80 & 120 mm Hg), the MCA that was stented with a WE43 stent
exhibited a cyclic strain amplitude (0.72 %) that was approximately 2.5 times greater
78
than for the 316L stented MCA (0.29%), as shown in Figure 4.10(d). This finding
suggested that the MCA exhibited significantly greater compliance when stented
with the WE43 stent compared to the 316L stent.
The surfaces of deployed stents were examined by SEM and there was no
evidence of strut failure or cracking, as shown in Figure 4.11(d).
Figure 4.10 (a) Pressure-diameter curves for freely expanded stents. Inset: WE43
stent shown unexpanded, fully expanded and after recoil, (b) performance metrics
for the free expansion and expansion in MCA of stents, (c) WE43 stent (top) and
316L stent (bottom) and (d) cyclic strain amplitude results. Error bars represent one
standard deviation of uncertainty.
79
Figure 4.11 Expansion testing of WE43 stent: (a) unexpanded, (b) fully expanded (c)
after recoil (d) SEM image of a strut section from an expanded stent
4.3
Corrosion Experiments
Based on the results of experiment A, the mass lost over time by the three groups of
dogbone specimen is shown in Figure 4.12. It can be seen that the rate of mass loss
increased for specimens with greater levels of plastic strain. Specimens with 0%
plastic strain exhibited a relatively low rate of mass loss for the first 1 d of
immersion; however, it increased thereafter to give a steady rate of mass loss from
which an average corrosion rate of 0.021 mg cm-2 h-1 was derived. For specimens
with plastic strain, higher rates of mass loss were observed for the first 1 d
immersion, which slowed significantly between 2 d and 4 d. Beyond 4 d, specimens
with 2.7% and 8.4% plastic strain exhibited relatively stable rates of mass loss from
which average corrosion rates of 0.025 mg cm-2 h-1 and 0.027 mg cm-2 h-1 were
derived, respectively.
SEM images, displaying typical surfaces on the inner walls of dogbone
specimens with 0% plastic strain, are shown in Figure 4.13. The surface morphology
produced by the surface treatments on a finished specimen, prior to immersion
80
testing, is shown in Figure 4.13(a). It can be seen that the magnesium matrix
material was uniformly removed by chemical etching but some intermetallic
precipitates were visible in the surface layer. After 1 d immersion, microscopic pits
were observed on this surface layer, as shown in Figure 4.13(b). Wide, shallow
corrosion pits, surrounded by elevated regions of darker material, were observed on
the surface after 3 d immersion, as shown in Figure 4.13(c). The pits increased in
size and number over time and thus coalesced and advanced over large regions of
the surface, as shown in Figure 4.13(d).
Figure 4.12 Plot of percentage mass lost over time by dogbone specimens based on
the results of experiment A. Error bars represent a single standard deviation from
the mean.
The composition of the material in the corrosion pits, as determined by EDX
analysis, was consistent with that of the magnesium matrix of the alloy. The
elevated regions were rich in RE elements (Zr, Y and Nd) and had compositions that
were consistent with the β-phase and other intermetallic compounds for the WE43
alloy. Pitting corrosion of the magnesium matrix, at the interface with the β-phase,
81
led to the undermining and fall-out of β-phase particles, as shown in Figure 4.13(e).
Corrosion attack was observed to progress in a layer by layer manner and thus, a
relatively uniform specimen morphology resulted after 10 d immersion, as shown in
Figure 4.13(f). When the inner and outer walls of the specimen were compared,
corrosion attack on laser cut edges resulted in a different morphology, as shown in
Figure 4.13(g).
Figure 4.13 SEM images of dogbone specimens with 0% PS following immersion in
solution for (a) 0 d, (b) 1 d, (c) 3 d, (d) 3 d (e) 10 d, (f) 10 d and, (g) 10 d
The corrosion processes observed for specimens with plastic strain was
similar to that described above for specimens with 0% plastic strain. However,
specimens with plastic strain exhibited significantly more advanced pitting corrosion
attack after just 1 d, and some cracks were observed, as shown in Figure 4.14(a).
Further, the corrosion layer on these specimens displayed many cracks and had a
significantly rougher appearance, as shown in Figure 4.14(b). As a result, pitting
corrosion of the magnesium matrix and the undermining of β-phase particles
progressed at an accelerated rate, as shown in Figure 4.14(c). Compared to 0%
plastic strain specimens, corrosion attack after 10 d immersion was greater for
82
specimens with 2.7% plastic strain, as shown in Figure 4.14(d) and significantly
greater for specimens with 8.4% plastic strain, as shown in Figure 4.14(e).
Corrosion attack on the laser cut edges was also more pronounced for specimens
with plastic strain. This may be attributed to a number of influences at the specimen
edge surfaces such as deformation-induced residual stresses and the influence of
processing (laser cutting and surface treatments) on the composition, microstructure
and crystallographic orientation of the material.
Figure 4.14 SEM images of dogbone specimens with 2.7% PS following immersion
in solution: (a) 1 d, (b) 6 d, (c) 10 d, (d) 10 d and, (e) 10 d (8.4%PS)
83
Based on the SEM images and the uniform nature of the corrosion observed,
it is suggested that localised micro-galvanic corrosion may be the dominant
corrosion processes in this case. The differences in the surface morphology and
mass loss rates for specimens with plastic strain suggests that plastic strain had a
significant influence on the corrosion behaviour. Plastic strain likely resulted in
significant damage to the brittle surface oxide thus allowing corrosion attack to occur
immediately after immersion in the solution and to progress at an accelerated rate.
The cracking observed on the surface might be attributed to material damage
induced by the application of plastic strain, and the influence of hydrogen
embrittlement and SCC. The application of plastic also likely led to deformationinduced acceleration in galvanic corrosion processes within the alloy.
Stress-strain curves with error bars representing 95% confidence intervals
are shown in Appendix A.2 for specimens with 0%, 2.7% and 8.4% plastic strain.
For simplicity, representative stress-strain curves are shown in Figure 4.15(a-c) for
specimens with 0%, 2.7% and 8.4% plastic strain, respectively. Specimen strength,
as determined by dividing the maximum load withstood by the specimen in tension
by the original specimen cross-sectional area, was seen to decrease significantly for
both immersion time and increasing plastic strain, as shown in Figure 4.15(d).
Similarly, there was a consistent reduction observed in E, σy and εu for increasing
immersion time and for increasing plastic strain, as shown in Figure 4.15(e-g). The
plastic strain induced prior to corrosion testing led to work-hardening and hence an
increase in the initial σy of the 2.7% and 8.4% plastic strain specimen series which
decreased with immersion time, as shown in Figure 4.15(g).
84
Figure 4.15 Representative stress-strain curves for dogbone specimens based on
the results of experiment A, (a) 0% PS (b) 2.7% PS and (c) 8.4% PS. Mechanical
properties of dogbone specimens over time: (d) UTS, (e) εu, (f) E and (g) σy. Error
bars represent one standard deviation of uncertainty.
SEM images of fractured dogbone specimens following tensile testing are
shown in Figure 4.16 (a-d). It can be seen that, compared to the non-corroded
specimen, the corroded specimens display more brittle type of fracture.
85
Figure 4.16 SEM images of typical fractured dogbone specimens after tensile
testing: (a) 0% PS and 0 d (i.e. no corrosion), (b) 0% PS and 10 d (c) 2.7% PS and
10 d, (d) 8.4% PS and 10 d immersion
Typical fracture surfaces for specimens are shown in Figure 4.17. Noncorroded specimens with 0% plastic strain presented dimple (ductile) rupture, as
shown in Figure 4.17(a). After 10 d immersion, specimens with 0% plastic strain
displayed corrosion pits along with ductile features on the non-corroded crosssection, as shown in Figure 4.17(b). After 10 d immersion, specimens with plastic
strain (2.7% and 8.4%) exhibited relatively greater pitting corrosion along with
intergranular cracking and some transgranular cracking (arrows), as shown in Figure
4.17(c-d).
86
Figure 4.17 Typical morphology of the fracture surfaces of dogbone specimens after
tensile testing (a) 0% PS and 0 d – dimple rupture, (b) 0% PS and 10 d – corrosion
pits with ductile features on non-corroded cross section (c) 2.7% PS and 10 d –
predominant intergranular cracking with brittle features and some transgranular
(arrow) cracking, (d) 8.4% PS and 10 d corrosion - predominant intergranular
cracking with brittle features and some transgranular (arrow) cracking
Based on the results of experiment B, the percentage mass lost over time by
freely expanded WE43 stents, as shown in Figure 4.18(a), was relatively high for the
first 10 h but decreased significantly between 10 h and 30 h. Beyond 30 h, the rate
of mass loss was largely steady from which an average corrosion rate of 0.021 mg
cm-2 h-1 was derived. The initial and steady state mass loss per unit area, observed
for the stent was greater than that observed for dogbone specimens with 0% PS, as
shown in Figure 4.18(b). Further, the shape of the curve for the stent was similar to
that for dogbone specimens with plastic strain in experiment A; suggesting that the
elevated rate of mass loss observed for the stent, may be attributed to the influence
87
of the plastic strain. The average strut width, in the longitudinal direction, decreased
with immersion time for the stent, as shown in Figure 4.18(c), to a final value of 103
µm after 60 h.
Figure 4.18 Results plots for experiment B (a) Percentage mass lost over time for
stent, (b) Mass loss per unit area for stent and dogbone specimens (c) Plot of
average strut width over time for stent. Error bars represent a single standard
deviation from the mean
Images of the typical corrosion surfaces of the stent are shown in Figure
4.19(a-d). The liberation of H2 gas bubbles into the solution, which indicated
corrosion activity, was visualised using phase contrast microscopy for a number of
stents and, was more pronounced at plastic hinge regions after 3 h immersion, as
shown in Figure 4.19(a). After 20 h, the stent surfaces were rougher in appearance
owing to preferential pitting corrosion of the magnesium matrix, as shown in Figure
4.19(b). A significant amount of material was corroded from the stent surface due to
the localised micro-galvanic attack after 40 h, as shown in Figure 4.19(c). The
morphology of stent struts after 60 h immersion is shown in Figure 4.19(d). It can be
seen that increased localised corrosion attack occurred at regions of the stent that
experienced high plastic strain during the deployment procedure, as indicated by the
arrows. Representative force-compression curves, derived from radial stiffness tests
88
performed on stents, are shown in Figure 4.19(e). The radial stiffness was
significantly lower for the WE43 stent compared to the 316L stent. The radial
stiffness of the WE43 stent decreased significantly with immersion time, as shown in
Figure 4.19(f), with a loss of 46 % in radial stiffness after 60 h immersion.
Figure 4.19 Typical morphology of a stent section after an immersion time of (a) 3 h,
(b) 20 h, (c) 40 h & (d) 60 h. (e) Representative force versus compression curves for
the WE43 and 316L stent (error bars omitted for clarity), (f) Temporal radial stiffness
of WE43 stent. Error bars represent a single standard deviation from the mean
4.4
Corrosion Modelling
4.4.1 Corrosion Model Calibration
Three simulations were performed using the corrosion model in order to predict the
corrosion behaviour of the dogbone test section for induced plastic strain values of
89
0%, 2.7% and 8.4%. A uniform plastic strain state was predicted in all simulations
thereby confirming that suitable boundary conditions and loading were used for the
model. Corrosion was simulated by the progression of the damage parameter, D,
and the removal of elements from the FE mesh. The progression of D for the case of
a dogbone model with 0% plastic strain is shown in Figure 4.20.
Figure 4.20 Damage parameter (D) values predicted by the corrosion model at 48 h
intervals for a specimen with 0% plastic strain
The parameters for the corrosion model, which were calibrated based on the
results of experiments A, are shown in Table 4.2. Parameters are shown for two
cases, that of an idealised uniform corrosion model and that of the plastic strain
model that most closely captured the experimentally observed corrosion behaviour.
The calibrated rate of mass loss from the corrosion model was compared to that
observed experimentally in Figure 4.21. It can be seen that the plastic strain model
was capable of capturing the experimentally observed rate of mass loss over time
for the three investigated values of plastic strain. However, the mass loss over time
predicted by the uniform corrosion model showed only moderate correlation with
that observed experimentally.
Table 4.2 Parameters used in the corrosion models in this work
90
Figure 4.21 Predicted mass loss from the calibrated corrosion model compared to
that observed experimentally
The plastic strain corrosion model was further able to capture the
experimentally observed specimen stress-strain response and, the non-linear
corrosion-induced reduction in specimen mechanical integrity, as shown in Figure
4.22, whereas the uniform model failed to predict the experimentally observed trend,
as shown in Figure 4.23.
Figure 4.22 Predicted specimen stress-strain response over time for the plastic
strain corrosion model compared to that observed experimentally for plastic strains
of (a) 0%, (b) 2.7% and (c) 8.4%
91
Figure 4.23 Predicted specimen stress-strain response over time for the uniform
corrosion model compared to that observed experimentally for plastic strains of (a)
0%, (b) 2.7% and (c) 8.4%
4.4.2 Stent Model
The simulated free expansion and recoil of the RUC stent model is shown in Figure
4.24(a) and (b), respectively. It can be seen that the predicted magnitude of εp1
varied from 0% to 27% for the deployed model. The stent outer diameters pre- and
post-recoil were 4.3 mm and 3.9 mm, respectively. Following expansion and recoil,
the corrosion of the stent was simulated using the plastic strain corrosion model, the
linear plastic strain corrosion model and the uniform corrosion model, as shown in
Figure 4.25(a-c), respectively. It can be seen that the predicted corrosion attack led
to a non-uniform degradation in the stent geometry for all three corrosion models,
with the largest amount of corrosion attack predicted at the plastic hinge regions of
the stent. Compared to the other corrosion models, the plastic strain corrosion
model predicted the greatest corrosion attack at the plastic hinge regions, as shown
in Figure 4.25(a).
92
Figure 4.24 Simulated deployment of the RUC stent model: (a) fully expanded (b)
predicted εp1 post recoil
Figure 4.25 Predicted corrosion damage (D) after an immersion time of 10 hours (a)
Plastic strain corrosion model (b) Linear PS corrosion model (c) Uniform corrosion
model
93
The plastic strain corrosion model predicted the greatest stent mass loss
over time and best characterised that observed experimentally, when compared to
the other corrosion models, as shown in Figure 4.26.
Figure 4.26 Predicted mass loss for the stent over time
The plastic strain corrosion model was further able to predict the influence of
plastic strain on the stent radial stiffness both quantitatively and qualitatively, while
the other two corrosion models failed to describe the non-linear relationship, as
shown in Figure 4.27(a). The force compression curves predicted by the plastic
strain corrosion model were in good agreement with those observed experimentally,
as shown in Figure 4.27(b).
The plastic strain corrosion model was used to predict the influence of three
different values of stent expansion ratio (SER) on the performance of the stent. It
was found that higher values of SER led to significant increases in the predicted
stent mass loss and in the temporal corrosion induced loss of radial stiffness for the
stent, as shown in Figure 4.28(a) and (b), respectively.
94
Figure 4.27 (a) Predicted stent radial stiffness at 10% compression (b) Predicted
stent radial force versus percentage compression over time
Figure 4.28 (a) Predicted stent mass loss for different values of stent expansion ratio
(SER) (b) Predicted stent radial stiffness at 10% compression for different values of
SER
95
5. Discussion
5.1
Stent Manufacture
Magnesium alloys possess many inherent characteristics that render them difficult to
cut by laser (Abderrazak et al., 2009). The use of an inert nitrogen shield gas,
appropriate laser cutting parameters and suitable surface treatments greatly
reduced the effects of the HAZ produced by laser cutting. Machining parameters
were optimised for the WE43 material in order to produce consistent strut
geometries and clean cuts with minimal dross and spatter adherence. This enabled
the normal fallout of waste sections from the machined stents.
Chemical etching provided a simple and inexpensive surface treatment
which removed dross and spatter remaining from laser cutting, improved surface
finish, imparted a non-toxic biocompatible phosphor-rich corrosion-resistant layer
and provided a coating pre-treatment in order to functionalise the stent surface.
SEM, AFM and EDX studies showed that plasma etching and cleaning in a
mixture of Ar-O2 gas plasma and pure O2 plasma successfully reduced surface
roughness by 10% compared to chemically etched surfaces and completely
removed surface contaminants remaining from the chemical etch process. Plasma
processing yielded improvements in biocompatibility through increased surface
sterility and wettability (Yi et al., 2004) and, also enhanced the surface adhesion
properties for subsequently deposited coating (Nakamura et al., 1996).
The combination of chemical and plasma etching surface treatments
produced stent surfaces with low roughness which have been associated with
decreased thrombogenicity and IH formation (De Scheerder et al., 2001) and
reductions in ISR in porcine models (Halwani et al., 2010). Also, the pitting tendency
and corrosion rate are reduced with decreasing surface roughness for magnesium
96
alloys (Walter and Kannan, 2010) thus, permitting the mechanical integrity of the
device to be maintained for a longer period of time.
The suitability of the processing route used in the manufacture of the
magnesium stent was confirmed in deployment testing where all stents were
expanded successfully and exhibited a uniform deployed geometry. During stent
deployment, the elevated strain levels that are induced at plastic hinge sites may
induce micro-cracks and other defects in poorly processed stents. Such defects can
propagate through the material leading to strut fracture which may contribute to
overall device failure and has been associated with higher rates of focal ISR (Celik
et al., 2008; Sianos et al., 2004). In this study, stents were deployed with no
cracking or fractures observed under SEM (Figure 4.11).
It should be acknowledged that cracks in the surface oxide layer are difficult
to detect in post-deployment analysis (i.e. under SEM) as they are likely to close
upon balloon deflation. Also, magnesium has high affinity to oxygen and therefore a
new surface oxide is likely to quickly form over cracks and other defects making
their detection difficult. Sub-surface defects would likely become evident with the
progression of corrosion in the substrate, though no evidence of this was observed
in the experiments. Non-destructive test (NDT) methods such as ultrasonic testing,
eddy current or phased array could be used to determine the severity of the material
damage induced in the struts by the stent deployment process.
5.2
Mechanical Testing
The ability of the WE43 stent to deploy at low balloon pressures, compared to
commercial 316L stents (Figure 4.10(a)) could potentially be beneficial in terms of
reducing arterial injury and hence ISR in arteries with low to moderate levels of
disease where predilation of stenosis is not required. Although it has been shown
that IH is not significantly greater after high pressure implantation of traditional BMS
97
compared to low pressure implantation (Hoffmann et al., 2001); no such
investigation has been carried out for bare-metal or drug-eluting magnesium stent
platforms. Lower radial stiffness and higher elastic recoil were observed for the
WE43 stent, compared to the 316L stent, in the deployment experiments (Figure
4.10(b)). Elastic recoil of a stent permits an immediate re-narrowing of the stented
artery upon balloon deflation and represents a negative clinical result. The relatively
high elastic recoil observed in this work for the WE43 stent in the MCA may be
attributed to the influences of the stent geometry, the relatively low elastic modulus
of WE43 and the moderate SER used in the experiments. These findings are in
agreement with those reported in clinical trials of the Biotronik Magic stent (AMS-1),
where post-dilation was frequently used to obtain larger lumen diameter, owing to
the relatively low radial stiffness and high acute recoil of the stent (Erbel et al.,
2007). Later versions of the Biotronik DREAMS stent used a non-commercial form
of the WE43 alloy and had a revised stent geometry that reduced elastic recoil
below 3% (Waksman, 2011).
Post-dilation and over-expansion presents a greater fracture risk for
magnesium stents compared to stents manufactured from traditional materials, such
as 316L, due to the increased likelihood of localised material fracture in magnesium
alloys (Grogan et al., 2012; Poncin and Proft, 2003). In this study, trial deployment
tests were carried out to investigate the risk of fracture for the magnesium stent.
Stents were deployed using both 3.5 mm and 4.0 mm angioplasty balloons,
representing SERs of 2.05 and 2.3, respectively. SEM analysis revealed a number
of fractured struts for a SER of 2.3. In all cases, the stent successfully expanded to
a SER of 2.05 (Figure 4.11(a)), with no fractures observed, which compared
favourably to the SER (1.5) observed for crimped AZ31 stents in a similar study,
where some strut fractures were reported (Wu et al., 2012). Therefore, all the
subsequent deployment testing carried out in this study used 3.5 mm angioplasty
98
balloons, as described in section 3.3.
In the third test (cyclic strain amplitude), prior to the application of the static
pressure loads, the elastomeric MCA acted to reduce the diameter of the stent that
was deployed within it. Under the influence of physiological pressure loads, the
diameter of both the MCA and the stent were observed to increase. It was noted that
contact was maintained at all times during the tests between the outer stent surface
and the inner MCA surface. In mechanics therefore, the stented MCA construct
could be represented by two springs of different stiffness connected in series. The
stent was significantly stiffer and likely followed a near-linear pressure-displacement
curve under physiological pressure loads. The MCA was significantly more
compliant and exhibited a non-linear hyperelastic response under pressure loads in
the physiological range (Colombo et al., 2010). Significantly higher cyclic strain
amplitude levels were observed for MCAs stented with the WE43 stent, compared to
the 316L stent (Figure 4.10(d)). This may be attributed to the influence of the lower
elastic modulus of WE43 compared to 316L and the significant differences that
existed in the geometrical design of the two stents. The WE43 stent had a greater
inner surface area, compared to the 316L stent, upon which static pressure acted to
deform the stent radially. This may have contributed to the higher cyclic strain
amplitude levels observed for the WE43 stent. Another limitation of the test was the
inability of the videoextensometer to measure the ID or OD of the stent directly by
visualisation due to the opacity of the MCA material. Therefore changes in the
diameter of the stent (and hence cyclic strain amplitude) were inferred from
measurements of the MCA OD which was not ideal. Small measurement errors in
the apparent stent OD were likely introduced due to thickness changes in the walls
of the MCA while under the influence of the two different pressure loads.
Compliant magnesium stent designs that facilitate high cyclic strain
amplitude levels that are closer to the normal physiological range (5-8%) (Draney et
99
al., 2002), and thus maintain the anti-proliferative effect of cyclic strain amplitude on
SMCs (Colombo et al. 2012), could offer an effective strategy to reduce ISR rates.
The subsequent corrosion-induced loss of mechanical integrity of magnesium stents
(Figure 4.19(f)) should produce a further temporal increase in the cyclic strain
amplitude. Upon completion of stent absorption, reconstitution of normal vessel
vasomotor function can be expected; thus removing many of the known stimuli for
IH and ST (Jabara et al., 2009). It has been suggested that current BAS may not be
ideal for patients with severe amounts of lesion calcification due to their limited
deliverability scaffolding properties (Kern, 2012). Conceptually, an ideal candidate
group for magnesium stents and other types of BAS may be young patients with
small plaque burden and focal disease or those requiring treatment of vulnerable
plaques prior to plaque rupture (Sabate, 2014).
5.3
Corrosion Experiments
The micro-galvanic corrosion attack observed in this work (Figure 4.13(c)) and, the
tendency toward a uniform localised corrosion that progressed in a layer by layer
manner (Figure 4.13(f)), is in agreement with results reported for larger samples of
WE43 in similar SBF compositions in the literature (Kalb et al., 2012). However, the
steady-state corrosion rates (0.021-0.027 mg cm-2 h-1) observed in this work (Figure
4.18(b)) were between eight and twenty orders of magnitude lower than those
reported for larger WE43 specimens (Rettig and Virtanen, 2009; Zhang et al., 2012)
and may be partly attributed to the increased resistance to galvanic corrosion
afforded by materials processing of the precursor tubes (Ge et al., 2013; Wang et
al., 2007). Further, the surface treatments imparted a magnesium phosphate
surface layer with reduced impurity levels (Nwaogu et al., 2009) and low surface
roughness (Gray-Munro et al., 2009; Walter et al., 2013) which significantly
improved corrosion resistance. This layer may have contributed to the low corrosion
rate observed for specimens with 0% plastic strain (Figure 4.12) though some
100
degradation of the layer was observed after 1 d immersion (Figure 4.13(a)). The
layer was significantly more degraded on specimens with plastic strain after 1 d
(Figure 4.14(a)).
Mass loss measurements, which are applicable to materials exhibiting nearuniform corrosion, revealed two different trends among the specimens (Figure 4.12).
Specimens with 0% plastic strain exhibited a low initial corrosion rate and a largely
linear rate thereafter; while higher levels of plastic resulted in larger increases in
mass loss in the first 2 d of immersion and moderate increases in steady-state
corrosion rate thereafter. The increased mass loss due to plastic strain may be
attributed to the likely damage caused to the brittle surface oxide and a deformationinduced acceleration in micro-galvanic attack combined with hydrogen embrittlement
and stress corrosion cracking (SCC). Damage and dissolution of the surface oxide
and severe pitting attack accompanied by cracking of the corrosion layer (Figure
4.14(a-b)) led to greater levels of material removal from the surface (Figure 4.14(c))
and increased corrosion on laser cut edges (Figure 4.14(e)). These findings are in
agreement with those from previous mechanical-electrochemical studies which
showed that plastic strain (5.5%) induced significant increases in both the pitting
susceptibility of grains (one order of magnitude) and the cathodic reactions at grain
boundaries (two order of magnitude) of an aluminium alloy in NaCl solution (Krawiec
et al., 2012). Similar findings have been reported for magnesium alloys that were
subjected to compression loading (Snir et al., 2012).
The corrosion-induced reduction in the thickness of stent struts (Figure
4.18(c)) was greatest in plastic hinge regions (Figure 4.18(d)) and contributed to a
46% reduction in radial stiffness after 60 h (Figure 4.19(f)) despite a total mass loss
of only 28% for the stent. The layer of corrosion products that are formed on the
specimen surfaces appeared to offer little protection to the substrate and thus
corrosion rates remained largely linear (Figure 4.18(b)). The composition of the
101
corrosion products, as determined by EDX analysis, was rich in Ca, O, P, Mg, Cl
and C, and was similar to that reported for WE43 specimens in similar solutions
(Kalb et al., 2012; Rettig and Virtanen, 2009). The shape of the mass loss curve and
the loss of approximately 25-30% stent mass after 60 h, due to uniform microgalvanic and SCC corrosion processes, is in agreement with results reported for an
electropolished AZ31 stent (Wu et al., 2012).
The corrosion-induced reduction in the stress-strain response and
mechanical properties (UTS, εu, E and σy) observed in experiment A (Figure 4.15),
likely resulted from the reduction in dogbone specimen cross-section and the
influence of stress concentrations in pitted regions and, was similar to the reduced
stress-strain responses reported for loaded and non-loaded magnesium specimens
(Bowen et al., 2012; Winzer et al., 2008). Compared to specimens with 0% plastic
strain, specimens with plastic strain exhibited a further reduction in mechanical
properties, most notably for εu, E and σy. It was concluded that the presence of
plastic strain increased the micro-galvanic corrosion activity and induced
intergranular and transgranular SCC (Figure 4.17). Similar SCC failure modes have
been reported for RE magnesium alloys after slow strain rate tests in NaCl solution
(Kannan et al., 2008). Magnesium alloys are susceptible to a number of stressmediated SCC processes which may lead to the rupture of the surface film, H2 gas
ingress and embrittlement and crack initiation and crack growth leading to fracture
(Winzer et al., 2005a).
The corrosion-induced reduction in mechanical strength observed for
specimens with 0% plastic strain (Figure 4.15(d)) was less rapid than that reported
previously for AZ31 and Mg–6Zn alloy specimens in similar studies (Grogan et al.,
2011; Zhang et al., 2010) and, may be attributed to the relatively uniform reduction
in specimen cross-section seen in this work. Consequently, a mass loss of 40% for
dogbone specimens and 28% for stents, corresponded to a 50% reduction in
102
specimen mechanical strength, compared to a mass loss of only 10-20% in similar
studies (Grogan et al., 2011; Zhang et al., 2010). By comparison, the pitting attack
observed in this work (Figure 4.13(c)) was less localised, owing in part to the
relatively low pitting factor (defined as the ratio of the deepest penetration to the
average penetration) of WE43 compared to other magnesium alloys (Kalb et al.,
2012).
While no significant differences were observed in the corrosion behaviour on
the inner and outer walls of specimens, the corrosion attack observed on laser cut
edges had a different morphology (Figure 4.13(g)) and progressed at a higher rate
when compared to the other surfaces, particularly in specimens with plastic strain
(Figure 4.14(d-e)). Since laser cutting may influence surface roughness, alloy
microstructure and expose higher index crystal planes of lower corrosion resistance
(Liu et al., 2008), in the absence of subsequent surface modification or coating,
higher corrosion rates can be expected on laser-cut edges; though a detailed study
has not been carried out for magnesium stents in this regard.
The results of experiment A and B suggest that corrosion behaviour was
strongly influenced by alloy microstructure and the magnitude of deployment strains
in this study. The findings demonstrate the potential for improvements in the
mechanical and corrosion behaviour of magnesium stents through improvements in
material processing and surface treatments and, optimisation of stent designs in
order to minimise deployment strains. Based on these findings, it is likely that
uniform deployed configurations and high strut apposition rates are critical to ensure
uniform and controlled corrosion of magnesium stent struts in vivo, with an
accompanying controlled reduction in corrosion-induced mechanical integrity, owing
to the influence of complex blood-tissue interactions on the corrosion behaviour of
magnesium alloys (Pierson et al., 2011; Wittchow et al., 2013).
103
The macroscopic tensile stress-strain response of non-corroded dogbone
specimens (Figure 4.15) was shown to be different to that reported for bulk
specimens of WE43 (Gu et al., 2010). Compared to bulk specimens with larger grain
sizes, dogbone specimens exhibited decreases in σy, modest decreases in UTS but
no significant differences in εu and εf. These findings are largely in agreement with
those from experimental (Guo et al., 2011) and numerical (Grogan et al., 2014b)
studies that examined the influence of grain size on the mechanical properties of
non-corroded magnesium materials. The differences in the mechanical properties of
dogbone and bulk specimens may be attributed to influences such as, the finegrained microstructure of the precursor tube (Ge et al., 2013) and size effects
resulting from the small specimen geometry and its processing. Magnesium
specimens with small grain size have a larger grain boundary fraction and
consequently release local strain energy faster, thereby exhibiting lower E values
(Moitra, 2013).
Studies of the corrosion behaviour of pure magnesium wires in static SBF
and rat artery models suggest that in vitro corrosion rates (as denoted by a
combined metric) are approximately three times greater than those observed in vivo
(Bowen et al., 2013). The lower in vivo corrosion rates observed for magnesium
alloys have been attributed to influences such as the adsorption of proteins on the
implant surface, the availability of oxygen (Feyerabend et al., 2012), the levels of
blood flow in nearby tissues (Willbold et al., 2013), the suppression of microgalvanic
corrosion due to encapsulation (Zainal Abidin et al., 2013) and the influence of
macrophages which can release acid digesting substances (Wittchow et al., 2013).
The in vitro corrosion rate observed in this study for the WE43 stent was lower than
those previously reported for WE43 specimen in similar SBF solutions. However, the
likely in vivo lifetime of the stent would be far too short to scaffold a coronary artery
for the six-to-nine month period required, as was the clinical experience with the
104
(WE43) bare-metal Biotronik AMS-1 (Erbel et al., 2007). Future improvements in
magnesium metallurgy and materials processing, combined with more effective
surface treatments and stent coating strategies should permit better control of
degradation rates and allow the lifetime of magnesium stents to be prolonged. Given
the high susceptibility of magnesium alloys to galvanic corrosion, accelerated in vitro
corrosion testing via electrochemical testing techniques, such as open circuit
potential and cyclic potentiodynamic polarisation testing (Gerber, 2008; Nascimento
and Zeddies, 2007), should provide useful methods for rapid evaluation of the
corrosion behaviour of different magnesium alloys for biomedical applications.
Biodegradable polymers such as poly (l-lactic acid) (PLLA), poly (Lcaprolactone) (PCL) and poly (glycolic acid) (PGA) have been approved for
numerous clinical applications (O’Brien and Carroll, 2009). More recently, the use of
these polymers as protective coatings intended to increase the corrosion resistance
of biodegradable magnesium has been explored. PCL and PLLA display good
adhesion and cytocompatibility properties, and offer significant reductions in the
corrosion rate of the magnesium substrate (Xu and Yamamoto, 2012). In a rat
model, poly(1,3-trimethylene carbonate) (PTMC) coatings reduced the corrosion
rate by three orders and one order of magnitude compared to bare and PCL coated
Mg-Zn-Mn alloy specimens, respectively (Wang et al., 2013). A PLGA coating was
used on the Biotronik AMS-3 stent. In animal trials, a mixture of 85% lactide and
15% glycolide provided the lowest intimal area at 28 days and 180 days, and the
highest endothelialisation score at 28 days (Wittchow et al., 2013). Chen et al.
(Chen et al., 2011) noted a special interaction between high purity magnesium and
both PCL and PLA coatings under dynamic SBF flow conditions, which had a
deleterious effect on corrosion performance. Such interactions between polymer
coating and the magnesium substrate could pose challenges to the future
development of drug-eluting magnesium stents.
105
5.4
Corrosion Modelling
The peak strain (~27%) predicted by the FE model for the deployed stent was
significantly greater than the experimentally observed value of failure strain for
WE43 (~20%). Strain in the FE model was predicted in the circumferential direction
(first principal strain) which was the direction most relevant to the stent deployment
analysis. It should be noted that the strain state in the deployed stent was more
complex than that induced experimentally by simple uniaxial tensile loading of the
dogbone specimens. Therefore, it is suggested that the use of a ductile failure
criterion, such as Oyane’s Criterion, within the FE model would facilitate better
prediction of material failure in future stent deployment analyses.
Observations from the corrosion experiments showed that uniform microgalvanic corrosion and SCC were likely to be the dominant corrosion processes for
both dogbone and stent specimens. The results also suggested a plastic strain
mediated corrosion attack for both specimens, as has been observed in previous
studies of magnesium alloys (Snir et al., 2012; Wittke et al., 2014). Therefore, an
element-specific plastic strain parameter (φe) was introduced in this work in order to
account for the effects of plastic strain induced corrosion in the modelling
framework. The plastic strain corrosion model, which was calibrated based on the
results of the corrosion experiments, was capable of predicting the experimentally
observed mass loss and the corrosion-induced reduction in stress-strain response of
the dogbone and stent specimens, whereas the uniform corrosion model failed to
capture the observed effects. The use of a non-linear relationship between εp1 and φe
for elements with εp1 values above of 8.4%, rather than a simple linear relationship,
in the corrosion model better captured the strain-mediated corrosion attack at plastic
hinge regions (Figure 4.25). Although only the proportion of elements in the model
having εp1 values was modest (4%), the increased corrosion rates prescribed for
these elements in the plastic strain model led to significantly greater reductions in
106
the predicted radial stiffness for the stent, without significantly increasing the overall
mass loss of the stent, when compared to the linear corrosion model. This approach
(i.e. the use of a non-linear relationship between εp1 and φe) was in contrast to
previous studies where a ductile failure condition was employed by setting D = 1 for
elements in which strains exceeded those observed experimentally at the UTS
(Gastaldi et al., 2011; Grogan et al., 2011). Thus, in these models, elements may be
immediately removed from the mesh at the start of the corrosion simulation leading
to the prediction of instantaneous mass loss at a simulation time of zero.
The plastic strain parameter (φe) used in this work accounted for both the
effects of corrosion and plastic strain (strain hardening) in a single parameter. It may
be possible to develop other parameters for use in the corrosion model which
delineate the individual influences of plastic strain and corrosion on specimen
mechanical integrity and mass loss. Figure 5.1(a) shows the typical stress-strain
response for a specimen induced with 2.7% plastic strain (curve 1) and, the
subsequent response for the same specimen prior to corrosion (curve 2) and after
corrosion (curve 3). The difference between the response observed at curve 1 and 2
indicated the loss of mechanical integrity (reduced ductility) due to plastic strain
while that observed at curve 2 and 3 indicated the loss of mechanical integrity due to
corrosion alone. As shown in Figure 5.1(b), the mass loss due to plastic strain
(green curve) could be delineated by subtracting the mass loss curves for 0% and
2.7% plastic strain.
This use of the element exposure parameter (κe) which scaled the corrosion
rate of the element according to the number of exposed element faces, allowed the
model to predict the rounding of initially sharp strut corners which is expected for a
diffusion-controlled corrosion process (Figure 4.25). This feature could also be
modified to include the effects of other corrosion phenomena that may be influenced
by the number of exposed element faces, such as the growth of corrosion pits.
107
Although the effects of pitting corrosion were not included in his work, the results of
the experiments suggested that the pitting tendency of the WE43 was low when
compared to that in previous studies where it was the primary corrosion process
(Grogan et al., 2011). The model developed in this work is deterministic in nature.
Probabilistic models are useful as they can be used to handle variability and
uncertainty which are inherent in many corrosion processes (MacDonald, 1994).
Figure 5.1 (a) Typical stress-strain response for a specimen induced with 2.7%
plastic strain (curve 1) and, the subsequent response (inset) for the same specimen
prior to corrosion (curve 2) and after corrosion (curve 3), (b) Mass loss curve (green)
that is driven by plastic strain alone
Simulation of stent performance may result in large and computationallyexpensive FE models owing to strong material and geometrical nonlinearities and,
complex multiple contact phenomena. The RUC corrosion model developed in this
work had good computational efficiency, requiring approximately nine CPU hours on
a single quad-core Intel i7 (2.6 GHz) on a desktop computer. The RUC stent model
offered significant reductions in model size and solution time which is especially
useful in the preliminary stages of stent design where many cycles of iteration
between the simulation and parameter selection are necessary (Xia et al., 2007).
108
5.5
Summary and Limitations
A number of limitations of this study should be highlighted. No post-cut annealing
was used to relieve residual stresses (Poncin and Proft, 2003) in the specimens
following laser cutting. However, the precursor tubes were pre-annealed prior to
laser cutting. The heat generated during laser cutting was minimised by using a
pulsed laser source, high pressure shield gas, and cutting paths and laser
parameters intended to minimise heat effects in the machined parts.
Although the surface treatments employed in this work are likely to differ
from those used for the Biotronik AMS-1 (Erbel et al., 2007), they provided effective
and inexpensive methods for functionalising the laser cut stents. The suitability of
the processing route used was demonstrated in bench tests where all stents were
expanded successfully to a SER of 2.1 and exhibited a uniform deployed geometry
with no evidence of strut fracture.
While the thickness of the dogbone specimen was equal to that of the stent
strut (140 µm), its width (1.85 mm) was approximately 1.5 orders of magnitude
greater. Therefore, the mechanical and corrosion behaviour observed for the
dogbone specimens was indicative of behaviour at that size-scale. However, when
compared to the existing studies for magnesium stents in the literature, the
specimen size, materials processing and surface treatments used in this work
represent a significant step towards elucidating the precise behaviour of magnesium
stent struts. Furthermore, the use of finished WE43 stents in the experiments of this
work to elucidate device performance characteristics represents a significant
contribution to knowledge.
The experimental performance of the stents was determined in MCAs that
mimicked only the passive mechanical properties of healthy coronary arteries under
static pressure conditions, though the results provided useful data regarding the
109
comparative performance of WE43 and 316L stents. The in vitro conditions used in
the corrosion experiments yielded useful fundamental information regarding the
corrosion behaviour of the stent and the alloy material but they cannot fully account
for the complex tissue interactions that may influence the performance of
magnesium devices in vivo (Pierson et al., 2011). However, it is noted that the
corrosion experiments in this work were carried out under sterile conditions in a cell
culture incubator; only a few such studies have been reported for magnesium
(Kirkland et al., 2010).
The influence of stent crimping, compressive loading and dynamic (fatigue)
loading on corrosion behaviour was not examined in this work. The corrosion model
was calibrated in experiments that examined corrosion behaviour at three plastic
strain values (0%, 2.7%, 8.4%) and at intervals of twenty four hours. The first
principal plastic strain (εp1) which acted parallel to the applied uniaxial tensile loading
in the dogbone specimen, was employed as the mediator of the corrosion in the
corrosion models. It is noted that the plastic strain state in the stent was non-uniaxial
owing to the complex loading pattern that is associated with the stent deployment
process though the model was still capable of capturing the experimentally observed
behaviour. Linear rather than continuous functions were used in the representation
of the mass loss curves and in the diffusion-controlled corrosion of elements in the
corrosion model.
110
6. Conclusions
6.1
Main Findings
The main aim of this study was the determination of the mechanical and corrosion
performance of an absorbable magnesium stent using experimental and numerical
analysis. A processing route was identified by which viable balloon-expandable
coronary stents were manufactured from precursor tubes of the WE43 alloy using a
combination of laser cutting and chemical and plasma-based surface treatments.
The physical and chemical characteristics of the stent surfaces produced were
revealed through SEM, AFM and EDX analyses. To the author’s knowledge, this is
the first study to fully describe the fabrication of WE43 stents and to describe the
surface treatment of magnesium stents by plasma processing (RIE).
While limited data regarding the performance of the Biotronik AMS-1 stent
(also WE43) has been reported in clinical trials (Erbel et al., 2007), this is the first
experimental study to report on the mechanical and corrosion performance of a
WE43 stent. It is also the first study to demonstrate the critical role of deploymentinduced plastic strain on the subsequent corrosion behaviour and temporal
scaffolding ability of a magnesium stent. By way of novel corrosion experiments,
plastic strain was shown to significantly increase corrosion rates, particularly at
plastic hinge regions, thus leading to a deformation-induced acceleration of the
underlying corrosion processes. The likely primary corrosion process was identified
as uniform micro-galvanic corrosion in non-deformed specimens and as uniform
micro-galvanic corrosion combined with SCC and hydrogen embrittlement for
specimens with plastic strain. Damage to the brittle surface oxide caused by plastic
strain was also thought to play a significant role in accelerating the corrosion attack.
Relatively modest mass losses at hinge regions were shown to induce significant
reductions in the temporal radial stiffness of the stent.
111
The corrosion rates observed in this work were found to be significantly
lower, with a reduced tendency toward corrosion processes such as pitting
corrosion, when compared to many of the existing corrosion studies of magnesium
alloys. The rate of mass loss and the nature of the underlying corrosion processes
observed for the stent was comparable with those of an electropolished AZ31 stent
in a previous study (Wu et al., 2012). When compared to the mechanical properties
reported for bulk WE43 specimens in the literature, the non-corroded dogbone
specimens used in this work exhibited significantly lower mechanical properties,
particularly with respect to E, σy and UTS. The observed effects were strongly linked
to the material microstructure and the surface properties of the specimens and thus
these findings highlight the vital importance of conducting testing on the stent itself
or a sample of the stent in its final form and finish. The experimental data presented
in this work, regarding the mechanical and corrosion performance of WE43 stents
and specimens, will serve as input in future computational models of stent
deployment, corrosion analysis and stent optimisation studies.
In the deployment experiments, the WE43 stent deployed at a significantly
lower range of balloon pressures than the commercially-available 316L stent. This
could potentially be beneficial in terms of reducing arterial injury and hence ISR in
arteries with low to moderate levels of disease where predilation of stenosis may be
avoided. However, this requires further investigation. The WE43 stent was found to
have a lower initial radial stiffness that was observed to decrease with immersion
time in the corrosion experiments. Non-corroded WE43 stents were shown to
support higher levels of cyclic strain amplitude which plays a crucial role in reducing
ISR, owing to the anti-proliferative influence of higher cyclic strain amplitude on
SMCs (Colombo et al., 2009). These findings highlight the need for improved
understanding of the biomechanical environment in stented vessels. In the future,
modulation of the cyclic strain amplitude in vessels stented with compliant
112
magnesium stent designs that have steerable corrosion kinetics should provide a
method to improve the ISR outcomes observed with current stents.
The need for corrosion models that explore the interaction of plastic strain
and plastic damage with corrosion damage has been recognised as one of the
current challenges in degradable material modelling (Gastaldi et al., 2011). A FE
based plastic strain mediated phenomenological corrosion model was developed in
this work and was calibrated based on the results of the corrosion experiments. It
was found to be capable of predicting the experimentally observed plastic strain
mediated mass loss profile and the corrosion-induced reduction in the radial
stiffness of the stent over time. To the author’s knowledge, the results presented
here represent the first experimental calibration of a plastic strain mediated
corrosion model of a corroding magnesium stent.
The results of this study provide new insights into the performance of an
absorbable magnesium stent and the fundamental corrosion processes that lead to
the temporal degradation of the stent. The findings will help direct future design
efforts towards the minimisation of plastic strain during device manufacture,
deployment and in-service, in order to reduce corrosion rates and prolong the
mechanical integrity of magnesium devices. Improvements in the corrosion
performance and biocompatibility of magnesium stents are needed and may be
achieved through improved stent design, advances in metallurgy and materials
processing, better surface treatments and coating strategies (Kitabata et al., 2014).
FE modelling is a powerful tool for device evaluation and corrosion
simulations and is a requirement for FDA approval of many medical devices, such
as coronary stents. Although the model developed in this work was validated and
applied to the case of a WE43 stent, it can be readily adapted in order to study the
behaviour of other alloys and devices. The model can also be used to examine the
113
effects of different materials processing, alloy compositions, and surface treatments
on overall device performance. It provides a flexible modelling framework which can
be adapted to explore other potentially relevant corrosion processes, such as pitting
and crevice corrosion, through modification of the damage law (Eqn. 3.4).
6.2
Future Work
The influence of the surface treatments used in this study on the various surface
properties which effect the biocompatibility of magnesium (Yang et al., 2011) require
further investigation. Although a bare-metal magnesium stent platform, similar to the
that developed in this work, demonstrated good biocompatibility in clinical trials
(Erbel et al., 2007), the toxicology and metabolic pathway of the RE elements used
in the WE43 alloy require further investigation (Gu and Zheng, 2010).
Biocompatibility testing should include cytotoxicity, haemolysis and platelet adhesion
testing, cell viability assays and additional corrosion analysis performed on
specimens with materials processing and surface treatments that are relevant to
stents. The work presented in this study should be advanced by carrying out
experimental mechanical testing of strut-sized magnesium specimens, as has been
investigated previously for traditional stent materials such as 316L (Murphy et al.,
2003). However, the determination of specimen mass loss by gravimetric
measurement at these size-scales represents a significant challenge. Hence,
alternative approaches to determine specimen corrosion rates may be required such
as Inductively Coupled Plasma Mass Spectrometry (ICP-MS) which can be used to
measure the concentration of magnesium ions released into solution by corroding
magnesium specimens (Schinhammer et al., 2010).
Drug-elution studies have shown that long-lasting antiproliferative effects do
not require sustained drug release (Garg & Serruys, 2010). These findings suggest
that effective magnesium BAS could be composed of an absorbable metal
114
backbone coated with an antiproliferative drug, rather than requiring drug wells or
the inclusion of antiproliferative agents in the backbone material (Topol, 2008). The
widespread adoption of magnesium stents requires improvements in device efficacy
which could be achieved through the development of a biocompatible coating
technology that permits control of drug delivery and device corrosion rate.
Since the study of plastic damage was the original application of the CDM
approach, future CDM-based models of corrosion damage could contribute to
exploring the interaction of plastic strain with corrosion damage for a range of
absorbable magnesium devices. The various components of plastic strain that are
induced during the processing of precursor tubes and stent manufacturing, as well
as those that result from fatigue loading in vivo, could be used in future CDM models
as the impetus for corrosion within the model. One approach might be to use
deformation-induced changes in electrochemical measurements in the magnesium
alloy as mediators of corrosion behaviour in the model. A number of studies have
explored this relationship between deformation-induced electrochemical changes
and corrosion behaviour in magnesium alloys (Snir et al., 2012; Wittke et al., 2014).
In the future, advanced imaging modalities and lower-cost screening tests
may enable identification and pre-emptive treatment with BAS of patients with
vulnerable plaques in order to reduce morbidity and mortality (Kukreja et al., 2009).
The unique properties of magnesium devices will lead to new treatment options and
improved outcomes for patients and should play a role in revolutionising medical
devices (Farraro et al., 2013). Tissue engineering strategies may be combined with
magnesium, such as the addition of biological agents to expedite healing of hard
and soft tissues. Such devices could find utility in a host of implants such as
coronary and ureteral stents (Lock et al., 2012), fracture fixation devices, ligament
and tendon reconstruction implants (Walton and Cotton, 2007), sutures and
ligatures (Seitz et al., 2011) and as bone substitutes (Krüger et al., 2013).
115
References
Abbas, G., Liu, Z., Skeldon, P., 2005. Corrosion behaviour of laser-melted
magnesium alloys. Appl. Surf. Sci. 247, 347–353.
Abderrazak, K., Kriaa, W., Ben Salem, W., Mhiri, H., Lepalec, G., Autric, M., 2009.
Numerical and experimental studies of molten pool formation during an
interaction of a pulse laser (Nd:YAG) with a magnesium alloy. Opt. &
Laser Technol. 41, 470–480.
Agema, W.R.P., Monraats, P.S., Zwinderman, A.H., De Winter, R.J., Tio, R. a,
Doevendans, P. a F.M., Waltenberger, J., De Maat, M.P.M., Frants, R.R.,
Atsma, D.E., Van Der Laarse, A., Van Der Wall, E.E., Jukema, J.W., 2004.
Current PTCA practice and clinical outcomes in The Netherlands: the real
world in the pre-drug-eluting stent era. Eur. Heart J. 25, 1163–70.
doi:10.1016/j.ehj.2004.05.006
Albaladejo, P., Samama, C.M., 2010. Patients under anti-platelet therapy. Best
Pract. Res. Clin. Anaesthesiol. 24, 41–50. doi:10.1016/j.bpa.2009.09.008
Alqahtani, A., suwaidi, J., Mohsen, M., 2013. Stent fracture: How frequently is it
recognized? Hear. Views 14, 72–81.
Alvarez-Lopez, M., Pereda, M.D., del Valle, J. a, Fernandez-Lorenzo, M., GarciaAlonso, M.C., Ruano, O. a, Escudero, M.L., 2010. Corrosion behaviour of AZ31
magnesium alloy with different grain sizes in simulated biological fluids. Acta
Biomater. 6, 1763–71. doi:10.1016/j.actbio.2009.04.041
Amira, S., Dubé, D., Tremblay, R., Ghali, E., 2008. Influence of the microstructure
on the corrosion behavior of AXJ530 magnesium alloy in 3.5% NaCl solution.
Mater. Charact. 59, 1508–1517. doi:10.1016/j.matchar.2008.01.018
Anis, R.R., Karsch, K.R., 2006. The future of drug eluting stents. Heart 92, 585–8.
doi:10.1136/hrt.2005.068288
Ansys, 2010. Ansys Parametric Design Language (APDL) User Guide, Release
13.0.
Aoki, J., Nakazawa, G., Tanabe, K., Hoye, A., Yamamoto, H., Nakayama, T.,
Onuma, Y., Higashikuni, Y., Otsuki, S., Yagishita, A., Yachi, S., Nakajima, H.,
Hara, K., 2007. Incidence and clinical impact of coronary stent fracture after
sirolimus-eluting stent implantation. doi:10.1002/ccd.20950
Arakawa, K., Isoda, K., Sugiyabu, Y., Fukuda, M., Nishizawa, K., Shibuya, T.,
Nakamura, H., 1998. Intimal proliferation after stenting reflected by increased
stent-to-vessel cross-sectional area ratio: serial intravascular ultrasound study.
J. Cardiol. 32, 379–389.
Argade, G.R., Panigrahi, S.K., Mishra, R.S., 2012. Effects of grain size on the
corrosion resistance of wrought magnesium alloys containing neodymium.
Corros. Sci. 58, 145–151. doi:10.1016/j.corsci.2012.01.021
Atrens, A., Liu, M., Zainal Abidin, N.I., 2011. Corrosion mechanism applicable to
biodegradable magnesium implants. Mater. Sci. Eng. B 176, 1609–1636.
Avedesian, M.M., Baker, H., 1999. Magnesium & Magnesium Alloys. ASM
International.
Bailey, S.R., 2009. DES Design: Theoretical Advantages and Disadvantages of
Stent Strut Materials, Design, Thickness, and Surface Characteristics. J. Interv.
Cardiol. 22, S3–S17. doi:10.1111/j.1540-8183.2009.00449.x
116
Berlin, J., 2011. Analysis of Boron with Energy Dispersive Spectrometry. Imaging
Microsc. 13, 19–21.
Birrer,
R.B.,
Shallash,
A.J.,
2002.
Selected
Topics :
Toxicology
HYPERMAGNESEMIA-INDUCED FATALITY FOLLOWING EPSOM SALT
GARGLES 22, 185–188.
Bosiers, M., Deloose, K., Verbist, J., Peeters, P., 2005. First Clinical Application of
Absorbable Metal Stents in the Treatment of Critical Limb Ischemia: 12-month
results. Vasc. Dis. Manag. 2, 86–91.
Bosiers, M., Peeters, P., D’Archambeau, O., Hendriks, J., Pilger, E., Düber, C.,
Zeller, T., Gussmann, A., Lohle, P.N.M., Minar, E., Scheinert, D., Hausegger,
K., Schulte, K.-L., Verbist, J., Deloose, K., Lammer, J., 2009. AMS INSIGHT-absorbable metal stent implantation for treatment of below-the-knee critical
limb ischemia: 6-month analysis. Cardiovasc. Intervent. Radiol. 32, 424–35.
doi:10.1007/s00270-008-9472-8
Bowen, P.K., Drelich, J., Goldman, J., 2013. A new in vitro-in vivo correlation for
bioabsorbable magnesium stents from mechanical behavior. Mater. Sci. Eng.
C. Mater. Biol. Appl. 33, 5064–70. doi:10.1016/j.msec.2013.08.042
Bowen, P.K., Drelich, J., Goldman, J., 2014. Magnesium in the murine artery :
Probing the products of corrosion. Acta Biomater. 10, 1475–83.
doi:10.1016/j.actbio.2013.11.021
Bowen, P.K., Gelbaugh, J. a, Mercier, P.J., Goldman, J., Drelich, J., 2012. Tensile
testing as a novel method for quantitatively evaluating bioabsorbable material
degradation. J. Biomed. Mater. Res. B. Appl. Biomater. 100, 2101–13.
doi:10.1002/jbm.b.32775
Briguori, C., Takahiro Nishida, Milena Adamian, Remo Albiero, Angelo Anzuini,
Carlo Di Mario, Antonio Colombo, Nishida, T., Adamian, M., Albiero, R.,
Anzuini, A., Mario, C. Di, Colombo, A., 2000. Coronary stenting versus balloon
angioplasty in small coronary artery with complex lesions. Catheter.
Cardiovasc. Interv. 50, 390–397.
Castagna, M.T., Mintz, G.S., Leiboff, B.O., Ahmed, J.M., Mehran, R., Satler, L.F.,
Kent, K.M., Pichard, A.D., Weissman, N.J., 2001. The contribution of
“mechanical” problems to in-stent restenosis: An intravascular ultrasonographic
analysis of 1090 consecutive in-stent restenosis lesions. Am. Heart J. 142,
970–974.
Celik, T., Iyisoy, A., Jata, B., Yuksel, C.U., Isik, E., 2008. Stent fracture: A new villain
of the village. Int. J. Cardiol. 138, 100–101.
Chen, Y., Song, Y., Zhang, S., Li, J., Zhao, C., Zhang, X., 2011. Interaction between
a high purity magnesium surface and PCL and PLA coatings during dynamic
degradation. Biomed. Mater. 6, 025005. doi:10.1088/1748-6041/6/2/025005
Chu, P.K., Chen, J.Y., Wang, L.P., Huang, N., 2002. Plasma-surface modification of
biomaterials. Mater. Sci. Eng. … 36, 143–206.
Colombo, A., Cahill, P.A., Lally, C., 2009. The amplitude rather than mean strain
determines the proliferative and apoptotic capacity of vascular smooth muscle
cells in vitro. Communications 81, 2009.
Colombo, A., Guha, S., Mackle, J.N., Cahill, P. a, Lally, C., 2012. Cyclic strain
amplitude dictates the growth response of vascular smooth muscle cells in
vitro: role in in-stent restenosis and inhibition with a sirolimus drug-eluting stent.
Biomech. Model. Mechanobiol. 12, 671–683. doi:10.1007/s10237-012-0433-4
117
Colombo, A., Karvouni, E., 2000. Biodegradable Stents : “Fulfilling the Mission and
Stepping Away”. Circulation 102, 371–373.
Colombo, A., Zahedmanesh, H., Toner, D.M., Cahill, P.A., Lally, C., 2010. A method
to develop mock arteries suitable for cell seeding and in-vitro cell culture
experiments. J. Mech. Behav. Biomed. Mater. 3, 470–477.
Connolley, T., McHugh, P.E., Bruzzi, M., 2005. A review of deformation and fatigue
of metals at small size scales. Fatigue Fract. Eng. Mater. Struct. 28, 1119–
1152. doi:10.1111/j.1460-2695.2005.00951.x
Coy, A.E., Viejo, F., Skeldon, P., Thompson, G.E., 2010. Susceptibility of rare-earthmagnesium alloys to micro-galvanic corrosion. Corros. Sci. 52, 3896–3906.
doi:10.1016/j.corsci.2010.08.006
Dangas, G.D., Claessen, B.E., Caixeta, A., Sanidas, E. a, Mintz, G.S., Mehran, R.,
2010. In-stent restenosis in the drug-eluting stent era. J. Am. Coll. Cardiol. 56,
1897–907. doi:10.1016/j.jacc.2010.07.028
Dangas, G.D., Serruys, P.W., Kereiakes, D.J., Hermiller, J., Rizvi, A., Newman, W.,
Sudhir, K., Smith, R.S., Cao, S., Theodoropoulos, K., Cutlip, D.E., Lansky, A.J.,
Stone, G.W., 2013. Meta-analysis of everolimus-eluting versus paclitaxeleluting stents in coronary artery disease: final 3-year results of the SPIRIT
clinical trials program (Clinical Evaluation of the Xience V Everolimus Eluting
Coronary Stent System in the Treatment of P. JACC. Cardiovasc. Interv. 6,
914–22. doi:10.1016/j.jcin.2013.05.005
De Scheerder, I., Sohier, J., Froyen, L., Van Humbeeck, J., Verbeken, E., 2001.
Biocompatibility of Coronary Stent Materials: Effect of Electrochemical
Polishing. doi:10.1002/1521-4052(200102)32:2<142::aid-mawe142>3.0.co;2-n
Demir, A.G., Previtali, B., Biffi, C.A., 2013. Fibre Laser Cutting and Chemical
Etching of AZ31 for Manufacturing Biodegradable Stents. Adv. Mater. Sci. Eng.
2013, 1–11.
Demir, A.G., Previtali, B., Colombo, D., Ge, Q., Vedani, M., Petrini, L., Wu, W., Biffi,
C.A., 2012. Fiber laser micromachining of magnesium alloy tubes for
biocompatible and biodegradable cardiovascular stents 823730–823739.
doi:10.1117/12.910131
Deshpande, K.B., 2010. Validated numerical modelling of galvanic corrosion for
couples: Magnesium alloy (AE44)–mild steel and AE44–aluminium alloy
(AA6063)
in
brine
solution.
Corros.
Sci.
52,
3514–3522.
doi:10.1016/j.corsci.2010.06.031
Deshpande, K.B., 2011. Numerical modeling of micro-galvanic corrosion.
Electrochim. Acta 56, 1737–1745. doi:10.1016/j.electacta.2010.09.044
Di Mario, C., Ferrante, G., 2008. Biodegradable drug-eluting stents: promises and
pitfalls. Lancet 371, 873–874.
Di Mario, C., Griffiths, H., Goktekin, O., Peeters, N., Verbist, J., Bosiers, M.,
Deloose, K., Heublein, B., Rohde, R., Kasese, V., Ilsley, C., Erbel, R., 2004.
Drug-eluting bioabsorbable magnesium stent. J. Interv. Cardiol. 17, 391–395.
doi:JOIC4081 [pii] 10.1111/j.1540-8183.2004.04081.x
Dietzel, W., Pfuff, M., Winzer, N., 2010. Testing and mesoscale modelling of
hydrogen assisted cracking of magnesium. Eng. Fract. Mech. 77, 257–263.
doi:10.1016/j.engfracmech.2009.07.009
Draney, M.T., Herfkens, R.J., Hughes, T.J.R., Pelc, N.J., Wedding, K.L., Zarins,
C.K., Taylor, C. a., 2002. Quantification of Vessel Wall Cyclic Strain Using Cine
118
Phase Contrast Magnetic Resonance Imaging. Ann. Biomed. Eng. 30, 1033–
1045. doi:10.1114/1.1513566
Drenth, D.J., Winter, J.B., Veeger, N.J.G.M., Monnink, S.H.J., van Boven, A.J.,
Grandjean, J.G., Mariani, M. a., Boonstra, P.W., 2002. Minimally invasive
coronary artery bypass grafting versus percutaneous transluminal coronary
angioplasty with stenting in isolated high-grade stenosis of the proximal left
anterior descending coronary artery: Six months’ angiographic and clinical
follow-u.
J.
Thorac.
Cardiovasc.
Surg.
124,
130–135.
doi:10.1067/mtc.2002.122525
Drynda, A., Deinet, N., Braun, N., Peuster, M., 2009. Rare earth metals used in
biodegradable magnesium-based stents do not interfere with proliferation of
smooth muscle cells but do induce the upregulation of inflammatory genes. J.
Biomed. Mater. Res. A 91, 360–9. doi:10.1002/jbm.a.32235
Duraiswamy, N., Schoephoerster, R.T., Moreno, M.R., Moore, J.E., 2007. Stented
Artery Flow Patterns and Their Effects on the Artery Wall. Annu. Rev. Fluid
Mech. 39, 357–382. doi:10.1146/annurev.fluid.39.050905.110300
Early, M., Lally, C., Prendergast, P.J., Kelly, D.J., 2009. Stresses in peripheral
arteries following stent placement: a finite element analysis. Comput. Methods
Biomech. Biomed. Engin. 12, 25–33. doi:10.1080/10255840802136135
Edelman, E.R., Rogers, C., 1998. Pathobiologic Responses to Stenting. Am. J.
Cardiol. 81, 4E–6E.
Erbel, R., Di Mario, C., Bartunek, J., Bonnier, J., de Bruyne, B., Eberli, F.R., Erne,
P., Haude, M., Heublein, B., Horrigan, M., Ilsley, C., Bose, D., Koolen, J.,
Luscher, T.F., Weissman, N., Waksman, R., 2007. Temporary scaffolding of
coronary arteries with bioabsorbable magnesium stents: a prospective, nonrandomised multicentre trial. Lancet 369, 1869–1875. doi:S01406736(07)60853-8 [pii] 10.1016/S0140-6736(07)60853-8
Erne, P., Schier, M., Resink, T.J., 2006. The road to bioabsorbable stents: reaching
clinical reality? Cardiovasc Interv. Radiol 29, 11–16. doi:10.1007/s00270-0040341-9
Fang, G., Ai, W., Leeflang, S., Duszczyk, J., Zhou, J., 2013. Multipass cold drawing
of magnesium alloy minitubes for biodegradable vascular stents. Mater. Sci.
Eng. C. Mater. Biol. Appl. 33, 3481–8. doi:10.1016/j.msec.2013.04.039
Farè, S., Ge, Q., Vedani, M., Vimercati, Ig., Gastaldi, D., Francesco Migliavacca;
Lorenza Petrini; Stefano Trasatti, 2010. Evaluation of material properties and
design requirements for biodegradable magnesium stents. Rev. Mater. 15, 96–
103.
Farraro, K.F., Kim, K.E., Woo, S.L.-Y., Flowers, J.R., McCullough, M.B., 2013.
Revolutionizing orthopaedic biomaterials: The potential of biodegradable and
bioresorbable magnesium-based materials for functional tissue engineering. J.
Biomech. doi:10.1016/j.jbiomech.2013.12.003
Feyerabend, F., Drücker, H., Laipple, D., Vogt, C., Stekker, M., Hort, N., Willumeit,
R., 2012. Ion release from magnesium materials in physiological solutions
under different oxygen tensions. J. Mater. Sci. Mater. Med. 23, 9–24.
doi:10.1007/s10856-011-4490-5
Feyerabend, F., Fischer, J., Holtz, J., Witte, F., Willumeit, R., Drücker, H., Vogt, C.,
Hort, N., 2010. Evaluation of short-term effects of rare earth and other
elements used in magnesium alloys on primary cells and cell lines. Acta
Biomater. 6, 1834–1842.
119
Frey, T., Noster, U., Scholtes, B., Uggowitzer, P.J., 2005. Microstructural influences
on the cyclic deformation behaviour of magnesium wrought alloys, in: H.
Kaufmann (Ed.), Proc. of the 2nd Int. Conf. on Light Metals Technology.
Springer Verlag, pp. 149 – 154.
Friedrich, H.E., Mordike, B.L. (Eds.), 2006. Magnesium Technology: Metallurgy,
Design Data, Applications, eBook. ed. Springer Berlin Heidelberg.
Fuentes, J.C., Salmon, A. a, Silver, M. a, 2006. Acute and chronic oral magnesium
supplementation: effects on endothelial function, exercise capacity, and quality
of life in patients with symptomatic heart failure. Congest. Heart Fail. 12, 9–13.
Fung, Y.C., 1993. Biomechanics. Mechanical Properties of Living Tissues , 2nd ed.
Springer-Verlag, New York.
Galvin, E., Morshed, M.M., Cummins, C., Daniels, S., Lally, C., MacDonald, B.J.,
2013. Surface Modification of Absorbable Magnesium Stents by Reactive Ion
Etching. J. Plasma Chem. Process. 33, 1137–1152. doi:10.1007/s11090-0139477-1
Gastaldi, D., Sassi, V., Petrini, L., Vedani, M., Trasatti, S., Migliavacca, F., 2011.
Continuum damage model for bioresorbable magnesium alloy devices -Application to coronary stents. J. Mech. Behav. Biomed. Mater. 4, 352–365.
Ge, Q., Dellasega, D., Demir, A.G., Vedani, M., 2013. The processing of ultrafinegrained Mg tubes for biodegradable stents. Acta Biomater. 9, 8604–10.
doi:10.1016/j.actbio.2013.01.010
Ge, Q., Vedani, M., Vimercati, G., 2011. Extrusion of Magnesium Tubes for
Biodegradable Stent Precursors. Mater. Manuf. Process. 27, 140–146.
doi:10.1080/10426914.2011.560231
Gerber, I., 2008. Metallic Medical Implants : Electrochemical Characterization of
Corrosion Processes 35–40.
Go, A.S., Mozaffarian, D., Roger, V.L., Benjamin, E.J., Berry, J.D., Borden, W.B.,
Bravata, D.M., Dai, S., Ford, E.S., Fox, C.S., Franco, S., Fullerton, H.J.,
Gillespie, C., Hailpern, S.M., Heit, J. a, Howard, V.J., Huffman, M.D., Kissela,
B.M., Kittner, S.J., Lackland, D.T., Lichtman, J.H., Lisabeth, L.D., Magid, D.,
Marcus, G.M., Marelli, A., Matchar, D.B., McGuire, D.K., Mohler, E.R., Moy,
C.S., Mussolino, M.E., Nichol, G., Paynter, N.P., Schreiner, P.J., Sorlie, P.D.,
Stein, J., Turan, T.N., Virani, S.S., Wong, N.D., Woo, D., Turner, M.B., 2013.
Heart disease and stroke statistics--2013 update: a report from the American
Heart
Association.
Circulation
127,
e6–e245.
doi:10.1161/CIR.0b013e31828124ad
Gogas, B., Farooq, V., Onuma, Y., Serruys, P., 2012. The ABSORB bioresorbable
vascular scaffold: an evolution or revolution in interventional cardiology. Hell. J
Cardiol 53, 301–309.
Gogas, B.D., McDaniel, M., Samady, H., King, S.B., 2014. Novel Drug-Eluting
Stents For Coronary Revascularization. Trends Cardiovasc. Med.
doi:10.1016/j.tcm.2014.07.004
Gray-Munro, J.E., Seguin, C., Strong, M., 2009. Influence of surface modification on
the in vitro corrosion rate of magnesium alloy AZ31. J. Biomed. Mater. Res. A
91, 221–30. doi:10.1002/jbm.a.32205
Grogan, J.A., Leen, S.B., McHugh, P.E., 2012. Comparing coronary stent material
performance on a common geometric platform through simulated bench
testing.
J.
Mech.
Behav.
Biomed.
Mater.
12,
129–38.
doi:10.1016/j.jmbbm.2012.02.013
120
Grogan, J.A., Leen, S.B., McHugh, P.E., 2013. Influence of statistical size effects on
the plastic deformation of coronary stents. J. Mech. Behav. Biomed. Mater. 20,
61–76. doi:10.1016/j.jmbbm.2012.12.008
Grogan, J.A., Leen, S.B., McHugh, P.E., 2014a. A physical corrosion model for
bioabsorbable
metal
stents.
Acta
Biomater.
10,
2313–2322.
doi:10.1016/j.actbio.2013.12.059
Grogan, J.A., Leen, S.B., McHugh, P.E., 2014b. Computational micromechanics of
bioabsorbable magnesium stents. J. Mech. Behav. Biomed. Mater. 34, 93–105.
doi:10.1016/j.jmbbm.2014.01.007
Grogan, J.A., O’Brien, B.J., Leen, S.B., McHugh, P.E., 2011. A corrosion model for
bioabsorbable metallic stents. Acta Biomater. 7, 3523–3533.
Gu, X.N., Zheng, Y.F., 2010. A review on magnesium alloys as biodegradable
materials. Front. Mater. Sci. China 4, 111–115.
Gu, X.N., Zhou, W.R., Zheng, Y.F., Cheng, Y., Wei, S.C., Zhong, S.P., Xi, T.F.,
Chen, L.J., 2010. Corrosion fatigue behaviors of two biomedical Mg alloys AZ91D and WE43 - In simulated body fluid. Acta Biomater. 6, 4605–4613.
Gunde, P., Hänzi, A.C., Sologubenko, A.S., Uggowitzer, P.J., 2011. High-strength
magnesium alloys for degradable implant applications. Mater. Sci. Eng. A 528,
1047–1054.
Gundogan, B., Tan, A., Farhatnia, Y., Alavijeh, M.S., Cui, Z., Seifalian, A.M., 2014.
Bioabsorbable stent quo vadis: a case for nano-theranostics. Theranostics 4,
514–33. doi:10.7150/thno.8137
Guo, L., Chen, Z., Gao, L., 2011. Effects of grain size, texture and twinning on
mechanical properties and work-hardening behavior of AZ31 magnesium
alloys. Mater. Sci. Eng. A 528, 8537–8545. doi:10.1016/j.msea.2011.07.076
Guo, L.F., Yue, T.M., Man, H.C., 2005. Excimer laser surface treatment of
magnesium alloy WE43 for corrosion resistance improvement. J. Mater. Sci.
40, 3531–3533. doi:10.1007/s10853-005-2888-5
Guy, L., 2013. PRECLINICAL STUDIES OF UPCOMING BIORESORBABLE
SCAFFOLDS Pre-clinical evaluation of the magnesium scaffold, in: EuroPCR.
Halfmann, H., Bibinov, N., Wunderlich, J., Awakowicz, P., 2007. A double inductively
coupled plasma for sterilization of medical devices. J. Phys. D. Appl. Phys. 40,
4145–4154. doi:10.1088/0022-3727/40/14/008
Hänzi, A.C., Gerber, I., Schinhammer, M., Löffler, J.F., Uggowitzer, P.J., 2010. On
the in vitro and in vivo degradation performance and biological response of new
biodegradable
Mg-Y-Zn
alloys.
Acta
Biomater.
6,
1824–33.
doi:10.1016/j.actbio.2009.10.008
Hänzi, A.C., Gunde, P., Schinhammer, M., Uggowitzer, P.J., 2009. On the
biodegradation performance of an Mg-Y-RE alloy with various surface
conditions in simulated body fluid. Acta Biomater. 5, 162–171.
Hara, H., Nakamura, M., Palmaz, J.C., Schwartz, R.S., 2006. Role of stent design
and coatings on restenosis and thrombosis. Adv. Drug Deliv. Rev. 58, 377–86.
doi:10.1016/j.addr.2006.01.022
Hartwig, a, 2001. Role of magnesium in genomic stability. Mutat. Res. 475, 113–21.
Hassel, T., Bach, F.W., Golovko, A.N., 2006. Production and Properties of Small
Tubes Made From MgCa0.8 for Application as Stent in Biomedical Science, in:
Kainer, K.U. (Ed.), 7th International Conference on Magnesium Alloys and
121
Their Applications. Wiley-VCH, Dresden, pp. 432–437.
Haude, M., Erbel, R., Erne, P., Verheye, S., Degen, H., Böse, D., Vermeersch, P.,
Wijnbergen, I., Weissman, N., Prati, F., Waksman, R., Koolen, J., 2013. Safety
and performance of the drug-eluting absorbable metal scaffold (DREAMS) in
patients with de-novo coronary lesions: 12 month results of the prospective,
multicentre, first-in-man BIOSOLVE-I trial. Lancet 381, 836–844.
doi:10.1016/S0140-6736(12)61765-6
Hermawan, H., Dubé, D., Mantovani, D., 2010a. Degradable metallic biomaterials:
design and development of Fe–Mn alloys for stents. J. Biomed. … 1–10.
Hermawan, H., Dubé, D., Mantovani, D., 2010b. Developments in metallic
biodegradable stents. Acta Biomater. 6, 1693–1697.
Hermawan, H., Mantovani, D., Dubé, D., Alamdari, A., 2008. Iron-Manganese: new
class of metallic degradable biomaterials prepared by powder metallurgy.
Powder Metall. 51, 38–45.
Heublein, B., Rohde, R., Kaese, V., Niemeyer, M., Hartung, W., Haverich, A., 2003.
Biocorrosion of magnesium alloys: a new principle in cardiovascular implant
technology? Heart 89, 651–656.
Hoffmann, R., Haager, P., Mintz, G.S., Kerckhoff, G., Schwarz, R., Franke, A., vom
Dahl, J., Hanrath, P., 2001. The impact of high pressure vs low pressure stent
implantation on intimal hyperplasia and follow-up lumen dimensions; results of
a randomized trial. Eur. Heart J. 22, 2015–24. doi:10.1053/euhj.2001.2609
Hoffmann, R., Radke, P.W., Ortlepp, J.R., Haager, P.K., Blindt, R., Iofina, E.,
Franke, A., Langenberg, R., Weber, C., Hanrath, P., 2004. Intravascular
ultrasonic comparative analysis of degree of intimal hyperplasia produced by
four different stents in the coronary arteries. Am. J. Cardiol. 94, 1548–1550.
Holzapfel, G. a., Sommer, G., Regitnig, P., 2004. Anisotropic Mechanical Properties
of Tissue Components in Human Atherosclerotic Plaques. J. Biomech. Eng.
126, 657. doi:10.1115/1.1800557
Holzapfel, G., Gasser, T., Ogden, R., 2000. A New Constitutive Framework for
Arterial Wall Mechanics and a Comparative Study of Material Models. J. Elast.
61, 1–48. doi:10.1023/a:1010835316564
Holzapfel, G.A., Ogden, R.W., 2006. Mechanics of biological tissues. SpringerVerlag, Berlin.
Holzapfel, G.A., Sommer, G., Gasser, C.T., Regitnig, P., 2005. Determination of
layer-specific mechanical properties of human coronary arteries with
nonatherosclerotic intimal thickening and related constitutive modeling. Am. J.
Physiol.
Hear.
Circ.
Physiol.
289,
H2048–H2058.
doi:10.1152/ajpheart.00934.2004
Hong, D., Saha, P., Chou, D.-T., Lee, B., Collins, B.E., Tan, Z., Dong, Z., Kumta,
P.N., 2013. In vitro degradation and cytotoxicity response of Mg-4% Zn-0.5%
Zr (ZK40) alloy as a potential biodegradable material. Acta Biomater. 9, 8534–
47. doi:10.1016/j.actbio.2013.07.001
Hornberger, H., Virtanen, S., Boccaccini, A.R., 2012. Biomedical coatings on
magnesium alloys - a review. Acta Biomater. doi:10.1016/j.actbio.2012.04.012
Hruby, A., McKeown, N.M., 2013. Bioactive Food as Dietary Interventions for
Diabetes, Bioactive Food as Dietary Interventions for Diabetes. Elsevier.
doi:10.1016/B978-0-12-397153-1.00032-9
Hunt, B., Poston, L., Schachter, M., Halliday, A., 2002. An introduction to vascular
122
biology; From basic science to clinical practice, Second. ed. Cambridge Press,
UK.
Jabara, R., Pendyala, L., Chen, J., Chronos, N., 2009. Bioabsorbable Stents: The
Future Is Near, A review of the breakthroughs and challenges of bioabsorbable
stents as a potential solution to the risks associated with available drug-eluting
stents. Card. Interv. Today June/July, 50–53.
Jain, A., Duygulu, O., Brown, D.W., Tomé, C.N., Agnew, S.R., 2008. Grain size
effects on the tensile properties and deformation mechanisms of a magnesium
alloy,
AZ31B,
sheet.
Mater.
Sci.
Eng.
A
486,
545–555.
doi:10.1016/j.msea.2007.09.069
Jia, J.X., Song, G., Atrens, A., 2006. Influence of geometry on galvanic corrosion of
AZ91D
coupled
to
steel.
Corros.
Sci.
48,
2133–2153.
doi:10.1016/j.corsci.2005.08.013
Ju, F., Xia, Z., Sasaki, K., 2008. On the finite element modelling of balloonexpandable stents. J. Mech. Behav. Biomed. Mater. 1, 86–95.
Kainer, K.U., 2003. Magnesium - Proceedings of the 7th International Conference
on Magnesium Alloys and Their Applications (My Ref. Book 2). Wiley-VCH
Verlag, Weinheim.
Kalb, H., Rzany, A., Hensel, B., 2012. Impact of microgalvanic corrosion on the
degradation morphology of WE43 and pure magnesium under exposure to
simulated
body
fluid.
Corros.
Sci.
57,
122–130.
doi:10.1016/j.corsci.2011.12.026
Kandzari, D.E., Leon, M.B., Meredith, I., Fajadet, J., Wijns, W., Mauri, L., 2013. Final
5-year outcomes from the Endeavor zotarolimus-eluting stent clinical trial
program: comparison of safety and efficacy with first-generation drug-eluting
and bare-metal stents. JACC. Cardiovasc. Interv. 6, 504–12.
doi:10.1016/j.jcin.2012.12.125
Kandzari, D.E., Tcheng, J.E., Zidar, J.P., 2002. Coronary artery stents: Evaluating
new designs for contemporary percutaneous intervention. Catheter.
Cardiovasc. Interv. 56, 562–576. doi:10.1002/ccd.10248
Kannan, M.B., 2010. Influence of microstructure on the in-vitro degradation
behaviour of magnesium alloys. Mater. Lett. 64, 739–742.
Kannan, M.B., Dietzel, W., Blawert, C., Atrens, A., Lyon, P., 2008. Stress corrosion
cracking of rare-earth containing magnesium alloys ZE41, QE22 and Elektron
21 (EV31A) compared with AZ80. Mater. Sci. Eng. A 480, 529–539.
Kastrati, A., Mehili, J., Dirschinger, J., 2001. Intracoronary stenting and angiographic
results: strut thickness effect on restenosis outcomes (ISAR-STEREO) trial.
ACC Curr. J. Rev. 10, 59.
Kern, M., 2012. Bioabsorbable Stents - Where are we now? Cath Lab Dig. 20, 4–6.
Kim, G.-H., Kim, C.-I., 2007. Dry etching of magnesium oxide thin films by using
inductively coupled plasma for buffer layer of MFIS structure. Thin Solid Films
515, 4955–4959. doi:10.1016/j.tsf.2006.10.087
Kim, J.-S., Jang, I.-K., Fan, C., Kim, T.H., Kim, J.-S., Park, S.M., Choi, E.-Y., Lee,
S.H., Ko, Y.-G., Choi, D., Hong, M.-K., Jang, Y., 2009. Evaluation in 3 months
duration of neointimal coverage after zotarolimus-eluting stent implantation by
optical coherence tomography: the ENDEAVOR OCT trial. JACC. Cardiovasc.
Interv. 2, 1240–7. doi:10.1016/j.jcin.2009.10.006
Kirkland, N.T., Lespagnol, J., Birbilis, N., Staiger, M.P., 2010. A survey of bio123
corrosion rates of magnesium alloys. Corros. Sci. 52, 287–291.
Kitabata, H., Waksman, R., Warnack, B., 2014. Bioresorbable metal scaffold for
cardiovascular application: Current knowledge and future perspectives.
Cardiovasc. Revasc. Med. 15, 109–116. doi:10.1016/j.carrev.2014.01.011
Kraitzer, A., Kloog, Y., Zilberman, M., 2008. Approaches for prevention of
restenosis. doi:10.1002/jbm.b.30974
Krawiec, H., Szklarz, Z., Vignal, V., 2012. Influence of applied strain on the
microstructural corrosion of AlMg2 as-cast aluminium alloy in sodium chloride
solution. Corros. Sci. 65, 387–396. doi:10.1016/j.corsci.2012.08.047
Krüger, R., Seitz, J.-M., Ewald, A., Bach, F.-W., Groll, J., 2013. Strong and tough
magnesium wire reinforced phosphate cement composites for load-bearing
bone replacement. J. Mech. Behav. Biomed. Mater. 20, 36–44.
doi:10.1016/j.jmbbm.2012.12.012
Kukreja, N., Onuma, Y., Serruys, P.W., 2009. Future Directions of Drug-Eluting
Stents.
J.
Interv.
Cardiol.
22,
S96–S105.
doi:10.1111/j.15408183.2009.00455.x
Kutniy, K. V, Papirov, I.I., Tikhonovsky, M.A., Pikalov, A.I., Sivtzov, S. V,
Pirozhenko, L.A., Shokurov, V.S., Shkuropatenko, V.A., 2009. Influence of
grain size on mechanical and corrosion properties of magnesium alloy for
medical implants Einfluss der Korngröße auf die mechanischen und
korrosionstechnischen Eigenschaften von Magnesiumlegierungen für
medizinische
Implantate.
Materwiss.
Werksttech.
40,
242–246.
doi:10.1002/mawe.200900434
Lally, C., Kelly, D.J., Prendergast, P.J., 2006. Stents, in: Wiley Encyclopedia of
Biomedical
Engineering.
John
Wiley
&
Sons,
Inc.
doi:10.1002/9780471740360.ebs1142
Lamaka, S.V., Montemor, M.F., Galio, a. F., Zheludkevich, M.L., Trindade, C., Dick,
L.F., Ferreira, M.G.S., 2008. Novel hybrid sol–gel coatings for corrosion
protection of AZ31B magnesium alloy. Electrochim. Acta 53, 4773–4783.
doi:10.1016/j.electacta.2008.02.015
Levine, G.N., Bates, E.R., Blankenship, J.C., Bailey, S.R., Bittl, J. a., Cercek, B.,
Chambers, C.E., Ellis, S.G., Guyton, R. a., Hollenberg, S.M., Khot, U.N.,
Lange, R. a., Mauri, L., Mehran, R., Moussa, I.D., Mukherjee, D., Nallamothu,
B.K., Ting, H.H., 2011. 2011 ACCF/AHA/SCAI Guideline for Percutaneous
Coronary Intervention: Executive Summary. J. Am. Coll. Cardiol. 58, 2550–
2583. doi:10.1016/j.jacc.2011.08.006
Lewington, S., Clarke, R., Qizilbash, N., Peto, R., Collins, R., 2003. Age-specific
relevance of usual blood pressure to vascular mortality. Lancet 361, 1391–
1392.
Li, J., Chen, R., Ma, Y., Ke, W., 2013. Effect of Zr modification on solidification
behavior and mechanical properties of Mg–Y–RE (WE54) alloy. J. Magnes.
Alloy. 1, 346–351. doi:10.1016/j.jma.2013.12.001
Li, J., Zheng, F., Qiu, X., Wan, P., Tan, L., Yang, K., 2014. Finite element analyses
for optimization design of biodegradable magnesium alloy stent. Mater. Sci.
Eng. C 42, 705–714. doi:10.1016/j.msec.2014.05.078
Li, N., Zhang, H., Ouyang, H., 2009. Shape optimization of coronary artery stent
based on a parametric model. Finite Elem. Anal. Des. 45, 468–475.
Li, N., Zheng, Y., 2013. Novel Magnesium Alloys Developed for Biomedical
124
Application: A Review. J.
doi:10.1016/j.jmst.2013.02.005
Mater.
Sci.
Technol.
29,
489–502.
Lim, H., Hur, G., Kim, S., Kim, Y., Kwon, S., 2008. Coronary Stent Fracture:
Detection with 64-Section Multidetector CT Angiography in Patients and in
Vitro1. Radiology 249, 810–819.
Liu, L.J., Schlesinger, M., 2009. Corrosion of magnesium and its alloys. Corros. Sci.
51, 1733–1737. doi:10.1016/j.corsci.2009.04.025
Liu, M., Qiu, D., Zhao, M.-C., Song, G., Atrens, A., 2008. The effect of
crystallographic orientation on the active corrosion of pure magnesium. Scr.
Mater. 58, 421–424.
Liu, S.Y., Hu, J.D., Yang, Y., Guo, Z.X., Wang, H.Y., 2005. Microstructure analysis
of magnesium alloy melted by laser irradiation. Appl. Surf. Sci. 252, 1723–
1731. doi:10.1016/j.apsusc.2005.03.110
Lock, J.Y., Draganov, M., Whall, A., Dhillon, S., Upadhyayula, S., Vullev, V.I., Liu,
H., 2012. Antimicrobial properties of biodegradable magnesium for next
generation ureteral stent applications. Conf. Proc. IEEE Eng. Med. Biol. Soc.
2012, 1378–81. doi:10.1109/EMBC.2012.6346195
Lunder, O., Lein, J.E., Hesjevik, S.M., Aune, T.K., Nisancioglu, K., 1994. Corrosion
morphologies on magnesium alloy AZ 91. Mater. Corros. 45, 331–340.
Ma, J., Thompson, M., Zhao, N., Zhu, D., 2014. Similarities and differences in
coatings for magnesium-based stents and orthopaedic implants. J. Orthop.
Transl. doi:10.1016/j.jot.2014.03.004
Macdonald, D.D., 1994. Performance Comparison Between a Statistical Model a
Deterministic Model , and an Artificial Neural Network Model for Predicting
Damage From Pitting Corrosion. J. Res. Natl. Inst. Stand. Technol. 99.
Magnesium Elektron, 2007. Material Datasheet WE43B, Magnesium Elektron,
http://www.magnesium-elektron.com/, February 2007.
Martini, F.H., 2006. Fundamentals of Anatomy and Physiology, 7th ed. Pearson
Education, San Francisco USA.
Mathers, T.W., Beckstrand, R.L., 2009. Oral magnesium supplementation in adults
with coronary heart disease or coronary heart disease risk. J. Am. Acad. Nurse
Pract. 21, 651–7. doi:10.1111/j.1745-7599.2009.00460.x
Mercado, N., Boersma, E., Wijns, W., Gersh, B.J., Morillo, C.A., de Valk, V., van Es,
G.-A., Grobbee, D.E., Serruys, P.W., 2001. Clinical and quantitative coronary
angiographic predictors of coronary restenosis: A comparative analysis from
the balloon-to-stent era. J. Am. Coll. Cardiol. 38, 645–652.
Migliavacca, F., Petrini, L., Colombo, M., Auricchio, F., Pietrabissa, R., 2002.
Mechanical behavior of coronary stents investigated through the finite element
method. J. Biomech. 35, 803–811.
Moitra, A., 2013. Grain size effect on microstructural properties of 3D
nanocrystalline magnesium under tensile deformation. Comput. Mater. Sci. 79,
247–251. doi:10.1016/j.commatsci.2013.05.051
Montoya, R., Iglesias, C., Escudero, M.L., García-Alonso, M.C., 2014. Modeling in
vivo corrosion of AZ31 as temporary biodegradable implants. Experimental
validation in rats. Mater. Sci. Eng. C. Mater. Biol. Appl. 41, 127–33.
doi:10.1016/j.msec.2014.04.033
Moravej, M., Prima, F., Fiset, M., Mantovani, D., 2010a. Electroformed iron as new
125
biomaterial for degradable stents: Development process and structureproperties relationship. Acta Biomater. 6, 1726–1735.
Moravej, M., Purnama, A., Fiset, M., Couet, J., Mantovani, D., 2010b. Electroformed
pure iron as a new biomaterial for degradable stents: In vitro degradation and
preliminary cell viability studies. Acta Biomater. 6, 1843–1851.
Morshed, M.M., Alam, M.M., Daniels, S.M., 2011. Moisture Removal from Natural
Jute Fibre by Plasma Drying Process. Plasma Chem. Plasma Process. 32,
249–258. doi:10.1007/s11090-011-9340-1
Morton, A.C., Crossman, D., Gunn, J., 2004. The influence of physical stent
parameters upon restenosis. Pathol. Biol. (Paris). 52, 196–205.
doi:10.1016/j.patbio.2004.03.013
Murphy, B.P., Savage, P., McHugh, P.E., Quinn, D.F., 2003. The Stress–Strain
Behavior of Coronary Stent Struts is Size Dependent. Ann. Biomed. Eng. 31,
686–691. doi:10.1114/1.1569268
Nakamura, Y., Suzuki, Y., Watanabe, Y., 1996. Effect of oxygen plasma etching on
adhesion between polyimide films and metal. Thin Solid Films 290-291, 367–
369.
Nascimento, M., Zeddies, M., 2007. Magnesium and its alloys as degradable
biomaterials: corrosion studies using potentiodynamic and EIS electrochemical
techniques. Materials (Basel). 10.
Nichols, W.W., O’Rourke, M.F., Vlachopoulos, C., 2005. McDonald’s blood flow in
arteries: theoretical, experimental, and clinical principles, 6th ed. Hodder
Arnold, New York.
Nishimoto, K.T., 2003. A study of plasma etching for use in active metals.
Massachusetts Institute of Technology.
Nishinari, M., Shimohama, T., Tojo, T., Shiono, T., Shinagawa, H., Kameda, R.,
Aoyama, N., Izumi, T., 2013. Two-week interval optical coherence tomography:
imaging evidence on neointimal coverage completion after implantation of the
Endeavor zotarolimus-eluting stent. Catheter. Cardiovasc. Interv. 82, E871–8.
doi:10.1002/ccd.25136
Nishio, S., Kosuga, K., Igaki, K., Okada, M., Kyo, E., Tsuji, T., Takeuchi, E.,
Inuzuka, Y., Takeda, S., Hata, T., Takeuchi, Y., Kawada, Y., Harita, T., Seki, J.,
Akamatsu, S., Hasegawa, S., Bruining, N., Brugaletta, S., de Winter, S.,
Muramatsu, T., Onuma, Y., Serruys, P.W., Ikeguchi, S., 2012. Long-Term (>10
Years) clinical outcomes of first-in-human biodegradable poly-l-lactic acid
coronary
stents:
Igaki-Tamai
stents.
Circulation
125,
2343–53.
doi:10.1161/CIRCULATIONAHA.110.000901
Nwaogu, U.C., Blawert, C., Scharnagl, N., Dietzel, W., Kainer, K.U., 2009. Influence
of inorganic acid pickling on the corrosion resistance of magnesium alloy AZ31
sheet. Corros. Sci. 51, 2544–2556. doi:10.1016/j.corsci.2009.06.045
O’Brien, B., Carroll, W., 2009. The evolution of cardiovascular stent materials and
surfaces in response to clinical drivers: A review. Acta Biomater. 5, 945–958.
O’Riordan, M., 2006. Disappearing act: Bioabsorbable magnesium-alloy stent safe
to implant. theHeart.org.
Oberhofer, G., Gese, H., Groß, M., Kühling, M., Seidel, D., 2006. Numerical
Analysis of the Balloon Dilatation Process Using the Explicit Finite Element
Method for the Optimization of a Stent Geometry, in: Proceedings of LS-DYNA
Anwenderforum. Ulm, pp. 35–46.
126
Ong, A.T.L., Hoye, A., Aoki, J., van Mieghem, C.A.G., Rodriguez Granillo, G.A.,
Sonnenschein, K., Regar, E., McFadden, E.P., Sianos, G., van der Giessen,
W.J., de Jaegere, P.P.T., de Feyter, P., van Domburg, R.T., Serruys, P.W.,
2005. Thirty-day incidence and six-month clinical outcome of thrombotic stent
occlusion after bare-metal, sirolimus, or paclitaxel stent implantation. J. Am.
Coll. Cardiol. 45, 947–953.
Onuma, Y., Serruys, P.W., 2011. Bioresorbable Scaffold: The Advent of a New Era
in Percutaneous Coronary and Peripheral Revascularization? Circulation 123,
779–797. doi:10.1161/circulationaha.110.971606
Pachla, W., Mazur, a., Skiba, J., Kulczyk, M., Przybysz, S., 2012. Wrought
Magnesium Alloys ZM21, ZW3 and WE43 Processed by Hydrostatic Extrusion
with Back Pressure. Arch. Metall. Mater. 57. doi:10.2478/v10172-012-0050-3
Palmerini, T., Biondi-Zoccai, G., Della Riva, D., Mariani, A., Sabaté, M., Smits, P.C.,
Kaiser, C., D’Ascenzo, F., Frati, G., Mancone, M., Genereux, P., Stone, G.W.,
2014. Clinical outcomes with bioabsorbable polymer- versus durable polymerbased drug-eluting and bare-metal stents: evidence from a comprehensive
network
meta-analysis.
J.
Am.
Coll.
Cardiol.
63,
299–307.
doi:10.1016/j.jacc.2013.09.061
Peuster, M., Hesse, C., Schloo, T., Fink, C., Beerbaum, P., von Schnakenburg, C.,
2006. Long-term biocompatibility of a corrodible peripheral iron stent in the
porcine descending aorta. Biomaterials 27, 4955–4962.
Peuster, M., Wohlsein, P., Brugmann, M., Ehlerding, M., Seidler, K., Fink, C.,
Brauer, H., Fischer, A., Hausdorf, G., 2001. A novel approach to temporary
stenting: degradable cardiovascular stents produced from corrodible metalresults 6-18 months after implantation into New Zealand white rabbits. Heart
86, 563–569.
Pierson, D., Edick, J., Tauscher, A., Pokorney, E., Bowen, P., Gelbaugh, J., Stinson,
J., Getty, H., Lee, C.H., Drelich, J., Goldman, J., 2011. A simplified in vivo
approach for evaluating the bioabsorbable behavior of candidate stent
materials. J. Biomed. Mater. Res. Part B Appl. Biomater. 100, 58–67.
doi:10.1002/jbm.b.31922
Poncin, P., Proft, J., 2003. Stent Tubing: Understanding the Desired Attributes.
Proc. ASM Conf. Mater. Process. Med. Devices 253–259.
Ratner, B.D., Hoffman, A.S., Schoen, F.J., Lemons, J.E., 1996. Biomaterials
science: an introduction to materials in medicine. Academic Press, San Diego.
Raulea, L.V., Goijaerts, A.M., 2001. Size effects in the processing of thin metal
sheets. … Mater. Process. … 115, 44–48. doi:10.1016/S0924-0136(01)007701
Rettig, R., Virtanen, S., 2009. Composition of corrosion layers on a magnesium rareearth alloy in simulated body fluids. J. Biomed. Mater. Res. A 88, 359–69.
doi:10.1002/jbm.a.31887
Sabate, M., 2014. Absorb STEMI - TROF II: The use of Bioresorbable Scaffolds in
Acute STEMI, in: EuroPCR Online.
Sabir, M., Liu, E.B., Li, Z., Zheng, Y.F., Li, L., 2012. Magnesium Alloy Stent
Expansion Behavior Simulated by Finite Element Method. Appl. Mech. Mater.
232, 697–700. doi:10.4028/www.scientific.net/AMM.232.697
Sammel, A.M., Chen, D., Jepson, N., 2013. New generation coronary stent
technology--is the future biodegradable? Heart. Lung Circ. 22, 495–506.
doi:10.1016/j.hlc.2013.02.008
127
Sangiorgi, G., Melzi, G., Agostoni, P., Cola, C., Clementi, F., Romitelli, P., Virmani,
R., Colombo, A., 2007. Engineering aspects of stents design and their
translation into clinical practice. Ann. Ist. Super. Sanita 43, 89–100.
Santhanakrishnan, S., Kumar, N., Dendge, N., Choudhuri, D., Katakam, S.,
Palanivel, S., Vora, H.D., Banerjee, R., Mishra, R.S., Dahotre, N.B., 2013.
Macro- and Microstructural Studies of Laser-Processed WE43 (Mg-Y-Nd)
Magnesium Alloy. Metall. Mater. Trans. B 44, 1190–1200. doi:10.1007/s11663013-9896-7
Savage, M.P., Fischman, D.L., Schatz, R.A., Teirstein, P.S., Leon, M.B., Baim, D.,
Ellis, S.G., Topol, E.J., Hirshfeld, J.W., Cleman, M.W., 1994. Long-term
angiographic and clinical outcome after implantation of a balloon-expandable
stent in the native coronary circulation. Palmaz-Schatz Stent Study Group. J
Am Coll Cardiol 24, 1207–1212.
Savage, P., O’Donnell, B.P., McHugh, P.E., Murphy, B.P., Quinn, D.F., 2004.
Coronary stent strut size dependent stress-strain response investigated using
micromechanical finite element models. Ann. Biomed. Eng. 32, 202–11.
Schinhammer, M., Hänzi, A.C., Löffler, J.F., Uggowitzer, P.J., 2010. Design strategy
for biodegradable Fe-based alloys for medical applications. Acta Biomater. 6,
1705–1713.
Schulz, C., Herrmann, R. a, Beilharz, C., Pasquantonio, J., Alt, E., 2000. Coronary
stent symmetry and vascular injury determine experimental restenosis. Heart
83, 462–7.
Schwartz, R.S., Holmes Jr, D.R., Topol, E.J., 1992. The restenosis paradigm
revisited: An alternative proposal for cellular mechanisms. J. Am. Coll. Cardiol.
20, 1284–1293.
Scott, N. a, 2006. Restenosis following implantation of bare metal coronary stents:
pathophysiology and pathways involved in the vascular response to injury. Adv.
Drug Deliv. Rev. 58, 358–76. doi:10.1016/j.addr.2006.01.015
Seitz, J.-M., Eifler, R., Stahl, J., Kietzmann, M., Bach, F.-W., 2012. Characterization
of MgNd2 alloy for potential applications in bioresorbable implantable devices.
Acta Biomater. 8, 3852–64. doi:10.1016/j.actbio.2012.05.024
Seitz, J.-M., Utermöhlen, D., Wulf, E., Klose, C., Bach, F.-W., 2011. The
Manufacture of Resorbable Suture Material from Magnesium – Drawing and
Stranding of Thin Wires. Adv. Eng. Mater. 13, 1087–1095.
doi:10.1002/adem.201100152
Serruys, P.W., de Jaegere, P., Kiemeneij, F., Macaya, C., Rutsch, W., Heyndrickx,
G., Emanuelsson, H., Marco, J., Legrand, V., Materne, P., Belardi, J., Sigwart,
U., Colombo, A., Goy, J.J., van den Heuvel, P., Delcan, J., Morel, M., 1994. A
Comparison of Balloon-Expandable-Stent Implantation with Balloon
Angioplasty in Patients with Coronary Artery Disease. N. Engl. J. Med. 331,
489–495. doi:doi:10.1056/NEJM199408253310801
Sianos, G., Hofma, S., Ligthart, J.M.R., Saia, F., Hoye, A., Lemos, P.A., Serruys,
P.W., 2004. Stent fracture and restenosis in the drug-eluting stent era.
Catheter. Cardiovasc. Interv. 61, 111–116. doi:10.1002/ccd.10709
Snir, Y., Ben-Hamu, G., Eliezer, D., Abramov, E., 2012. Effect of compression
deformation on the microstructure and corrosion behavior of magnesium alloys.
J. Alloys Compd. 528, 84–90. doi:10.1016/j.jallcom.2012.03.010
Song, G., 1999. Corrosion mechanisms of magnesium alloys. Adv. Eng. Mater. 11–
33.
128
Song, G., Atrens, a., 2003. Understanding Magnesium Corrosion—A Framework for
Improved
Alloy
Performance.
Adv.
Eng.
Mater.
5,
837–858.
doi:10.1002/adem.200310405
Song, G., Atrens, a., 2007. Recent Insights into the Mechanism of Magnesium
Corrosion and Research Suggestions. Adv. Eng. Mater. 9, 177–183.
doi:10.1002/adem.200600221
Song, G., Atrens, A., 1999. Corrosion mechanisms of magnesium alloys. Adv. Eng.
Mater.
1,
11–33.
doi:10.1002/(SICI)1527-2648(199909)1:1<11::AIDADEM11>3.3.CO;2-E
Song, G., Atrens, A., Dargusch, M., 1999. Influence of microstructure on the
corrosion of diecast AZ91D. Corros. Sci. 41, 138–162.
Song, G.L., 2011. Corrosion of Magnesium Alloys. Woodhead Publishing, Oxford
Philadelphia, PA.
Song, G.-L., Xu, Z., 2012. Crystal orientation and electrochemical corrosion of
polycrystalline Mg. Corros. Sci. 63, 100–112. doi:10.1016/j.corsci.2012.05.019
Stack, R.E., Califf, R.M., Phillips, H.R., 1988. Interventional cardiac catheterization
at Duke Medical Center. Am J Cardiol .
Staiger, M.P., Pietak, A.M., Huadmai, J., Dias, G., 2006. Magnesium and its alloys
as orthopedic biomaterials: a review. Biomaterials 27, 1728–1734.
Sternberg, K., Gratz, M., Koeck, K., Mostertz, J., Begunk, R., Loebler, M.,
Semmling, B., Seidlitz, A., Hildebrandt, P., Homuth, G., Grabow, N., Tuemmler,
C., Weitschies, W., Schmitz, K.-P., Kroemer, H.K., 2011. Magnesium used in
bioabsorbable stents controls smooth muscle cell proliferation and stimulates
endothelial cells in vitro. J. Biomed. Mater. Res. Part B Appl. Biomater. 100B,
41–50. doi:10.1002/jbm.b.31918
Stone, G., Ellis, S., Cox, D., 2004. A polymer-based, paclitaxel-eluting stent in
patients with coronary artery disease. … Engl. J. … 350, 221–231.
Sun, W., Wang, L., Wu, T., Liu, G., 2014. An arbitrary Lagrangian–Eulerian model
for modelling the time-dependent evolution of crevice corrosion. Corros. Sci.
78, 233–243. doi:10.1016/j.corsci.2013.10.003
Svilaas, T., van der Horst, I.C.C., Zijlstra, F., 2007. A quantitative estimate of baremetal stenting compared with balloon angioplasty in patients with acute
myocardial infarction: angiographic measures in relation to clinical outcome.
Heart 93, 792–800. doi:10.1136/hrt.2006.093740
Tamai, H., Igaki, K., Kyo, E., Kosuga, K., Kawashima, A., Matsui, S. et al., 2000.
Initial and 6-month results of biodegradable poly-l-lactid acid coronary sents in
humans. J Am Hear. Assoc.
Thamida, S.K., 2012. Modeling and simulation of galvanic corrosion pit as a moving
boundary
problem.
Comput.
Mater.
Sci.
65,
269–275.
doi:10.1016/j.commatsci.2012.07.029
Topol, E.J., 2008. Textbook of Interventional Cardiology, 5th Ed. ed. Elsevier Inc.,
Philadelphia.
Transparency Market Research, 2014. Coronary Stents Market (Bare Metal Stents,
Drug Eluting Stents and Bio-absorbable Stents) - Global Industry Analysis,
Size, Share, Pipeline Overview, Growth, Trends and Forecast, 2013 - 2019.
Ulvrová, M., Labrosse, S., Coltice, N., Råback, P., Tackley, P.J., 2012. Numerical
modelling of convection interacting with a melting and solidification front:
129
Application to the thermal evolution of the basal magma ocean. Phys. Earth
Planet. Inter. 206-207, 51–66. doi:10.1016/j.pepi.2012.06.008
Valente, T., 2001. Grain boundary effects on the behavior of WE43 magnesium
castings. Corrosion 67–69.
Vedani, M., Ge, Q., Wu, W., Petrini, L., 2012. Texture effects on design of Mg
biodegradable stents. Int. J. Mater. Form. 7, 31–38. doi:10.1007/s12289-0121108-5
Vernhet, H., Demaria, R., Juan, J.M., Oliva-Lauraire, M.C., Sénac, J.P., Dauzat, M.,
2001. Changes in Wall Mechanics After Endovascular Stenting in the Rabbit
Aorta. Am. J. Roentgenol. 176, 803–807.
Vernhet, H., Juan, J.M., Demaria, R., Oliva-Lauraire, M.C., Sénac, J.P., Dauzat, M.,
2000. Acute Changes in Aortic Wall Mechanical Properties after Stent
Placement in Rabbits. J. Vasc. Interv. Radiol. 11, 634–638.
Waksman, R., 2011. The Biotronik Bioabsorbable Magnesium Scaffold DREAMS,
in: TCTAP 2011. Seoul, Korea.
Waksman, R., Erbel, R., Di Mario, C., Bartunek, J., de Bruyne, B., Eberli, F.R., Erne,
P., Haude, M., Horrigan, M., Ilsley, C., Böse, D., Bonnier, H., Koolen, J.,
Lüscher, T.F., Weissman, N.J., 2009. Early- and Long-Term Intravascular
Ultrasound and Angiographic Findings After Bioabsorbable Magnesium Stent
Implantation in Human Coronary Arteries. J. Am. Coll. Cardiol. Cardiovasc.
Interv. 2, 312–320.
Waksman, R., Pakala, R., Kuchulakanti, P.K., Baffour, R., Hellinga, D., Seabron, R.,
Tio, F.O., Wittchow, E., Hartwig, S., Harder, C., Rohde, R., Heublein, B.,
Andreae, A., Waldmann, K.H., Haverich, A., 2006. Safety and efficacy of
bioabsorbable magnesium alloy stents in porcine coronary arteries. Catheter
Cardiovasc Interv 68, 607–609. doi:10.1002/ccd.20727
Walter, R., Kannan, M.B., 2010. Influence of surface roughness on the corrosion
behaviour of magnesium alloy. Mater. Des. 32, 2350–2354.
Walter, R., Kannan, M.B., He, Y., Sandham, a., 2013. Effect of surface roughness
on the in vitro degradation behaviour of a biodegradable magnesium-based
alloy. Appl. Surf. Sci. 279, 343–348. doi:10.1016/j.apsusc.2013.04.096
Walton, C. a., Horstemeyer, M.F., Martin, H.J., Francis, D.K., 2014. Formulation of a
macroscale corrosion damage internal state variable model. Int. J. Solids
Struct. 51, 1235–1245. doi:10.1016/j.ijsolstr.2013.12.007
Wang, H., Estrin, Y., Fu, H., Song, G., Zúberová, Z., 2007. The Effect of PreProcessing and Grain Structure on the Bio-Corrosion and Fatigue Resistance
of
Magnesium
Alloy
AZ31.
Adv.
Eng.
Mater.
9,
967–972.
doi:10.1002/adem.200700200
Wang, J., He, Y., Maitz, M.F., Collins, B., Xiong, K., Guo, L., Yun, Y., Wan, G.,
Huang, N., 2013. A surface-eroding poly(1,3-trimethylene carbonate) coating
for fully biodegradable magnesium-based stent applications: toward better
biofunction, biodegradation and biocompatibility. Acta Biomater. 9, 8678–89.
doi:10.1016/j.actbio.2013.02.041
Wang, X., Cui, F., Li, J., Zhao, X., 2009. [Mechanical analysis on a new type of
biodegradable magnesium-alloy stent]. Sheng Wu Yi Xue Gong Cheng Xue Za
Zhi 26, 338–341.
Weiß, S., Szymczak, H., Meißner, A., 2009. Fatigue and endurance of coronary
stents. doi:10.1002/mawe.200800409
130
Welt, F.G.P., Rogers, C., 2002. Inflammation and Restenosis in the Stent Era.
Arterioscler.
Thromb.
Vasc.
Biol.
22,
1769–1776.
doi:10.1161/01.ATV.0000037100.44766.5B
Wiersma, S.A., Taylor, D., 2004. A critical distance approach applied to microscopic
316L biomedical specimens under fatigue loading, in: The 15th European
Conference of Fracture-Advanced Fracture Mechanics for Life and Safety
Assessments. Sweden.
Willbold, E., Kalla, K., Bartsch, I., Bobe, K., Brauneis, M., Remennik, S., Shechtman,
D., Nellesen, J., Tillmann, W., Vogt, C., Witte, F., 2013. Biocompatibility of
rapidly solidified magnesium alloy RS66 as a temporary biodegradable metal.
Acta Biomater. 9, 8509–17. doi:10.1016/j.actbio.2013.02.015
Williams, D., 2006. New interests in magnesium. Med Device Technol 17, 9–10.
Winzer, N., Atrens, a, Dietzel, W., Raja, V., Song, G., Kainer, K., 2008.
Characterisation of stress corrosion cracking (SCC) of Mg–Al alloys. Mater.
Sci. Eng. A 488, 339–351. doi:10.1016/j.msea.2007.11.064
Winzer, N., Atrens, a., Song, G., Ghali, E., Dietzel, W., Kainer, K.U., Hort, N.,
Blawert, C., 2005a. A Critical Review of the Stress Corrosion Cracking (SCC)
of
Magnesium
Alloys.
Adv.
Eng.
Mater.
7,
659–693.
doi:10.1002/adem.200500071
Winzer, N., Atrens, A., Dietzel, W., Song, G., Kainer, K.U., 2007. Stress corrosion
cracking in magnesium alloys: Characterization and prevention. Jom 59, 49–
53. doi:10.1007/s11837-007-0104-6
Winzer, N., Atrens, A., Song, G., Ghali, E., Dietzel, W., Kainer, K.U., Hort, N.,
Blawert, C., 2005b. A Critical Review of the Stress Corrosion Cracking (SCC)
of
Magnesium
Alloys.
Adv.
Eng.
Mater.
7,
659–693.
doi:10.1002/adem.200500071
Wittchow, E., Adden, N., Riedmüller, J., Savard, C., Waksman, R., Braune, M.,
2013. Bioresorbable drug-eluting magnesium-alloy scaffold: design and
feasibility in a porcine coronary model. EuroIntervention 8, 1441–50.
doi:10.4244/EIJV8I12A218
Witte, F., 2010. The history of biodegradable magnesium implants: A review. Acta
Biomater. 6, 1680–1692.
Witte, F., Hort, N., Vogt, C., Cohen, S., Kainer, K.U., Willumeit, R., Feyerabend, F.,
2008. Degradable biomaterials based on magnesium corrosion. Curr. Opin.
Solid State Mater. Sci. 12, 63–72. doi:10.1016/j.cossms.2009.04.001
Wittke, P., Klein, M., Walther, F., 2014. Corrosion Fatigue Behaviour of Creepresistant Magnesium Alloy Mg-4Al-2Ba-2Ca. Procedia Eng. 74, 78–83.
doi:10.1016/j.proeng.2014.06.228
Wu, Q., Zhu, S., Wang, L., Liu, Q., Yue, G., Wang, J., Guan, S., 2012. The
microstructure and properties of cyclic extrusion compression treated Mg-Zn-YNd alloy for vascular stent application. J. Mech. Behav. Biomed. Mater. 8, 1–7.
doi:10.1016/j.jmbbm.2011.12.011
Wu, W., Chen, S., Gastaldi, D., Petrini, L., 2012. Experimental data confirm
numerical modeling on degradation process of magnesium alloys stents. Acta
Biomater. doi:10.1016/j.actbio.2012.10.035.
Wu, W., Gastaldi, D., Yang, K., Tan, L., Petrini, L., Migliavacca, F., 2011. Finite
element analyses for design evaluation of biodegradable magnesium alloy
stents in arterial vessels. Mater. Sci. Eng. B 176, 1733–1740.
131
Wu, W., Petrini, L., Gastaldi, D., Villa, T., Vedani, M., Lesma, E., Previtali, B.,
Migliavacca, F., 2010. Finite Element Shape Optimization for Biodegradable
Magnesium Alloy Stents. Ann. Biomed. Eng. 38, 2829–2840.
doi:10.1007/s10439-010-0057-8
Xia, Z., Ju, F., Sasaki, K., 2007. A general finite element analysis method for balloon
expandable stents based on repeated unit cell (RUC) model. Finite Elem. Anal.
Des. 43, 649–658.
Xin, Y., Huo, K., Tao, H., Tang, G., Chu, P.K., 2008. Influence of aggressive ions on
the degradation behavior of biomedical magnesium alloy in physiological
environment. Acta Biomater. 4, 2008–2015.
Xu, L., Yamamoto, A., 2012. Characteristics and cytocompatibility of biodegradable
polymer film on magnesium by spin coating. Colloids Surf. B. Biointerfaces 93,
67–74. doi:10.1016/j.colsurfb.2011.12.009
Xue, Y., Horstemeyer, M., Mcdowell, D., Elkadiri, H., Fan, J., 2007. Microstructurebased multistage fatigue modeling of a cast AE44 magnesium alloy. Int. J.
Fatigue 29, 666–676. doi:10.1016/j.ijfatigue.2006.07.005
Yang, J., Cui, F., Lee, I.S., 2011. Surface modifications of magnesium alloys for
biomedical applications. Ann. Biomed. Eng. 39, 1857–1871.
Yang, J., Cui, F., Lee, I.S., Wang, X., 2010. Plasma surface modification of
magnesium alloy for biomedical application. Surf. Coatings Technol. 205,
S182–S187. doi:10.1016/j.surfcoat.2010.07.045
Yang, L., Zhang, E., 2009. Biocorrosion behavior of magnesium alloy in different
simulated fluids for biomedical application. Mater. Sci. Eng. C 29, 1691–1696.
doi:10.1016/j.msec.2009.01.014
Yang, Z., Li, J.P., Zhang, J.X., Lorimer, G.W., Robson, J., 2008. Review on
Research and Development of Magnesium Alloys. Acta Metall. Sin. (English
Lett. 21, 313–328.
Yi, C.H., Jeong, C.H., Lee, Y.H., Ko, Y.W., Yeom, G.Y., 2004. Oxide surface
cleaning by an atmospheric pressure plasma. Surf. Coatings Technol. 177-178,
711–715. doi:10.1016/j.surfcoat.2003.08.011
Yusuf, S., Hawken, S., Ounpuu, S., Dans, T., Avezum, A., Lanas, F., McQueen, M.,
Budaj, A., Pais, P., Varigos, J., Lisheng, L., 2004. Effect of potentially
modifiable risk factors associated with myocardial infarction in 52 countries (the
INTERHEART
study):
case-control
study.
Lancet
364,
937–52.
doi:10.1016/S0140-6736(04)17018-9
Yutani, C., Imakita, M., Ishibashi-Ueda, H., Tsukamoto, Y., Nishida, N., Ikeda, Y.,
1999. Coronary atherosclerosis and interventions: Pathological sequences and
restenosis. Pathol. Int. 49, 273–290. doi:10.1046/j.1440-1827.1999.00861.x
Zahedmanesh, H., John Kelly, D., Lally, C., 2010. Simulation of a balloon
expandable stent in a realistic coronary artery--Determination of the optimum
modelling strategy. J. Biomech. 43, 2126–2132.
Zainal Abidin, N.I., Rolfe, B., Owen, H., Malisano, J., Martin, D., Hofstetter, J.,
Uggowitzer, P.J., Atrens, A., 2013. The in vivo and in vitro corrosion of highpurity magnesium and magnesium alloys WZ21 and AZ91. Corros. Sci. 75,
354–366. doi:10.1016/j.corsci.2013.06.019
Zeng, R., Dietzel, W., Witte, F., Hort, N., Blawert, C., 2008. Progress and Challenge
for Magnesium Alloys as Biomaterials. Adv. Eng. Mater. 10, B3–B14.
doi:10.1002/adem.200800035
132
Zeng, R., Zhang, J., Huang, W., Dietzel, W., 2006. Review of studies on corrosion of
magnesium alloys. Trans. Nonferrous Met. Soc. China 16, 763–771.
Zhang, J., Niu, X., Qiu, X., Liu, K., Nan, C., Tang, D., Meng, J., 2009. Effect of
yttrium-rich misch metal on the microstructures, mechanical properties and
corrosion behavior of die cast AZ91 alloy. J. Alloys Compd. 471, 322–330.
doi:10.1016/j.jallcom.2008.03.089
Zhang, S., Zhang, X., Zhao, C., Li, J., Song, Y., Xie, C., Tao, H., Zhang, Y., He, Y.,
Jiang, Y., Bian, Y., 2010. Research on an Mg-Zn alloy as a degradable
biomaterial. Acta Biomater. 6, 626–40. doi:10.1016/j.actbio.2009.06.028
Zhang, X., Yuan, G., Mao, L., Niu, J., Ding, W., 2012. Biocorrosion properties of asextruded Mg–Nd–Zn–Zr alloy compared with commercial AZ31 and WE43
alloys. Mater. Lett. 66, 209–211. doi:10.1016/j.matlet.2011.08.079
Zhang, Y., Bourantas, C. V, Farooq, V., Muramatsu, T., Diletti, R., Onuma, Y.,
Garcia-Garcia, H.M., Serruys, P.W., 2013. Bioresorbable scaffolds in the
treatment of coronary artery disease. Med. Devices (Auckl). 6, 37–48.
doi:10.2147/MDER.S22547
Zhao, H.D., Qin, G.W., Ren, Y.P., Pei, W.L., Chen, D., Guo, Y., 2011. The maximum
solubility of Y in α-Mg and composition ranges of Mg24Y5−x and Mg2Y1−x
intermetallic phases in Mg–Y binary system. J. Alloys Compd. 509, 627–631.
doi:10.1016/j.jallcom.2010.09.120
Zhao, M.-C., Liu, M., Song, G.-L., Atrens, A., 2008. Influence of pH and chloride ion
concentration on the corrosion of Mg alloy ZE41. Corros. Sci. 50, 3168–3178.
Zheng, Y.F., Gu, X.N., Witte, F., 2014. Biodegradable metals. Mater. Sci. Eng. R
Reports 77, 1–34. doi:10.1016/j.mser.2014.01.001
Zhou, W., Shan, D., Han, E.-H., Ke, W., 2008. Structure and formation mechanism
of phosphate conversion coating on die-cast AZ91D magnesium alloy. Corros.
Sci. 50, 329–337. doi:10.1016/j.corsci.2007.08.007
Zhou, W.R., Zheng, Y.F., Leeflang, M. a, Zhou, J., 2013. Mechanical property,
biocorrosion and in vitro biocompatibility evaluations of Mg-Li-(Al)-(RE) alloys
for future cardiovascular stent application. Acta Biomater. 1–11.
doi:10.1016/j.actbio.2013.01.032
Zohdi, T.I., Holzapfel, G. a, Berger, S. a, 2004. A phenomenological model for
atherosclerotic plaque growth and rupture. J. Theor. Biol. 227, 437–43.
doi:10.1016/j.jtbi.2003.11.025
Zorion Medical, 2014. ZMED Device Information [WWW Document]. URL
http://www.zorionmedical.com/ (accessed 7.10.13).
133
Appendix A
A.1 Compositional analysis
Three specimens (n=3) of the WE43 precursor tube were acid digested using a
Milestone Microwave Digestion System and analysed using Inductively Coupled
Plasma Mass Spectrometry (ICP-MS) (7500 ICP-MS, Agilent Technologies) in order
to determine the compositional content of magnesium and the three main alloying
elements (Y, Zr & Nd). The weight % of magnesium and the alloying elements was
found to be consistent with that specified by the manufacturer of WE43 (Magnesium
Elektron, 2007), as shown in Table A.1.
Table A.1 Elemental composition (average weight %) of WE43 tube specimen
determined by ICP-MS
A.2 Additional Data from Corrosion Experiment A
Figure A.1 Stress-strain curves (terminating at the UTS) for dogbone specimens
based on the results of experiment A, (a) 0% PS (b) 2.7% PS and (c) 8.4% PS.
Error bars represent 95% confidence intervals derived from T-table.
134
A.3 APDL Implementation of Corrosion Model
!=====================================================================
!----- Plastic strain-mediated corrosion algorithm V5.1 ----------!----- Written by: Emmet Galvin, Date: 10/07/14 -------------------!----- All text to the right of “!” are comments -----------------------!=====================================================================
!--------------- Set control parameters for corrosion algorithm ----------------------finish
/prep7
! Enter preprocessor
k_U=8.1667e-3
! Corrosion kinetic parameter (hr^-1)
Delta_t=1
! Time step (hr)
tMax=60
! Maximum solution time (hr)
ItNum=tMax/Delta_t+1
! Set number of iterations
ItNum=nint(ItNum)
! Find nearest integer value of ItNum
INL=ItNum+1
delta_U=0.015
! Material characteristic length (mm)
DeDt=delta_U*k_U*Delta_t
esel,s,type,,1
*get,J,elem,,count
esel,all
! Get the number of elements (J) in
! the selected set
!--------------- Create element characteristic length (L_e) array -------------------*del,L_e
*dim,L_e,array,J
! Create array for L_e
*do,EE,1,J
*get,V_e,elem,EE,volu
! Get element volume
L_e(EE)=(V_e)**(1/3)
! Get L_e and write
*enddo
! to L_e array
!--------------- Create damage (D) array ------------------------------------------------*del,De1
*dim,De1,array,J,INL
! Create array for De1
*do,EE,1,J
De1(EE,1)=0
! Initialise damage parameter
De1(EE,2)=0
*enddo
!--------------- Create results array ------------------------------------------------------*del,MML
*dim,MML,array,INL,4
! Create results array
MML(1,1)=0
! Iteration number
MML(1,2)=0
! Solution time (hr)
MML(1,3)=J
! Number of live elements
MML(1,4)=0
! Model mass loss(%) from DES
!--------------- Create K array --------------------------------------------------------------*del,K
*dim,K,array,J,9
! K (no. of exposed element faces) array
*do,EE,1,J
*do,xx,1,9
K(EE,xx)=0
*enddo
*enddo
! Initialise K parameter
esel,s,,,allbcelem
135
*get,J3,elem,0,count
i=0
xx=1
*do,xx,1,J3
ff=elnext(i)
K(ff,7)=1
i=ff
*enddo
alls
! Set K(ff,7)=1 for elements
! on symmetry edges
!--------------- Create current element strain (CES) array ----------------------------*del,CES
*dim,CES,array,J
*do,EE,1,J
CES(EE)=0
! CES array records element first
*enddo
! principal plastic strain
!--------------- Create DES (current Element Damage) array -----------------------*del,DES
*dim,DES,array,J
! DES array records element damage (D)
*do,EE,1,J
DES(EE)=0
*enddo
!--------------- Create recent death (RDM) array ----------------------------------------*del,RDM
*dim,RDM,array,J,INL
! RDM array records element death
*do,EE,1,J
*do,zz,1,INL
RDM(EE,zz)=0
*enddo
*enddo
!--------------- Create material model number (MMN) array ------------------------*del,MMN
*dim,MMN,array,J
*do,EE,1,J
MMN(EE)=1
! Set initial material model
*enddo
! for all elements to 1
alls
!--------------- Create plastic strain parameter (Phi) array --------------------------!--------------- to scale DeDt based on current element strain (CES) ------------*DIM,mult1,TABLE,6,6,1, , ,
*TAXIS,mult1(1,1,,),, 0,0.027,0.084,0.12,0.168,0.35
! Plastic strain
*TAXIS,mult1(1,1,,),2,24,48,96,144,192,240
! Time (hr)
mult1(1,1)=0.368, 2.190, 5.830, 25.00, 70.00, 700.0
mult1(1,2)=1.287, 1.265, 1.264, 6.000, 12.00, 20.00
mult1(1,3)=1.000, 0.632, 0.630, 1.500, 2.000, 3.000
mult1(1,4)=1.035, 1.092, 1.149, 1.150, 1.250, 1.350
mult1(1,5)=0.862, 1.149, 1.093, 1.093, 1.093, 1.093
mult1(1,6)=0.785, 1.095, 1.483, 1.800, 2.000, 2.200
!--------------- Simulate Stent Deployment & Recoil ----------------------------------finish
! Finish preprocessor
/solu
! Enter solution
nlgeom,on
! Activate large deformation controls
solcon,on
! Activate solution controls
136
sf,stentinner,pres,0.95
nsubst,50,100,20
outres,all,last
time,0.9
solve
! Simulate stent expansion
sf,stentinner,pres,0
nsubst,20,20,20
outres,all,all
time,1.0
solve
! Simulate stent elastic recoil
!--------------- Update CES array ---------------------------------------------------------finish
! Finish solution
/post1
! Enter postprocessor
etable,epl,eppl,1
! create etable 'epl' of first
*do,EE,1,J
! principal plastic strain
*get,pps,elem,EE,etab,epl
CES(EE)=pps
! update CES array
*enddo
I=2
parsav, all, params1, ,
parsav,scalar,scalparams,,
finish
/solu
antype,,restart,,,continue
parres, new, params1,,
parres, change, scalparams,,
nsubst,1,1,1
outres,all,last,
alls
save
! Save scalar parameters to text file
! CES data is lost after restart so
! must read from text file
! Save last result set only
!--------------- Corrosion Algorithm Iterative cycle ----------------------------------*do,I,2,ItNum
! I varies from 2 to ItNum
CT=I-1
/gopr
! Corrosion Time (hr)
! Reactivate suppressed printout
ST=Delta_t*(CT)
*if,ST,ge,0,and,ST,le,24,then
y=24
*elseif,ST,gt,24,and,ST,le,48,then
y=48
*elseif,ST,gt,48,and,ST,le,96,then
y=96
*elseif,ST,gt,96,and,ST,le,144,then
y=144
*elseif,ST,gt,144,and,ST,le,192,then
y=192
*elseif,ST,gt,192,and,ST,le,240,then
y=240
*endif
! Solution Time (hr)
! y mediates mult1 array column
! selection
!--------------- Update Ke array -----------------------------------------------esel,s,live
! Select live elements
esel,r,type,,1
137
*do,EE,1,J
! Find number of neighbouring elements
*do,fn,1,6
*if,eladj(EE,fn),lt,1,then
K(EE,fn)=0
*else
*get,ELS1,elem,eladj(EE,fn),attr,live
*if,eladj(EE,fn),ge,1,and,ELS1,eq,1,then
K(EE,fn)=1
*else
K(EE,fn)=0
*endif
*endif
*enddo
K(EE,8)=6-(K(EE,1)+K(EE,2)+K(EE,3)+K(EE,4)+K(EE,5)+K(EE,6)+K(EE,7))
! K(EE,8)= number of exposed elem faces
*if,K(EE,8),eq,0,then
! Set Ke parameter based on K(EE,8)
K(EE,9)=0.000
*elseif,K(EE,8),eq,1,then
K(EE,9)=1.000
*elseif,K(EE,8),eq,2,then
K(EE,9)=1.289
*elseif,K(EE,8),eq,3,then
K(EE,9)=2.182
*elseif,K(EE,8),eq,4,then
K(EE,9)=2.667
*elseif,K(EE,8),eq,5,then
K(EE,9)=2.805
*elseif,K(EE,8),eq,6,then
K(EE,9)=2.949
*endif
*enddo
!--------------- Corrosion process--------------------------------------------esel,s,live
esel,r,type,,1
nsle,r
nsel,r,ext
nsel,u,,,allbcnode
esln,r
! Select external elems for corrosion
*do,EE,1,J
*get,ESS,elem,EE,esel
! Get elem selection status (ESS)
*if,ESS,eq,1,then
*get,ELS,elem,EE,attr,live
*endif
! Element life status
*if,ESS,eq,1,and,ELS,eq,1,then
Phi=mult1(CES(EE),y)
! Set Phi from mult1 array
De1(EE,I)=De1(EE,I-1)+((DeDt*Phi*K(EE,9))/L_e(EE))
*else
De1(EE,I)=De1(EE,I-1)
*endif
!------- Change material model number or kill element -----*if,De1(EE,I),ge,0,and,De1(EE,I),le,0.1,then
/wait,0
*elseif,De1(EE,I),ge,0.1,and,De1(EE,I),lt,0.2,then
138
MPCHG,2,EE
*elseif,De1(EE,I),ge,0.2,and,De1(EE,I),lt,0.3,then
MPCHG,3,EE
*elseif,De1(EE,I),ge,0.3,and,De1(EE,I),lt,0.4,then
MPCHG,4,EE
*elseif,De1(EE,I),ge,0.4,and,De1(EE,I),lt,0.5,then
MPCHG,5,EE
*elseif,De1(EE,I),ge,0.5,and,De1(EE,I),lt,0.6,then
MPCHG,6,EE
*elseif,De1(EE,I),ge,0.6,and,De1(EE,I),lt,0.7,then
MPCHG,7,EE
*elseif,De1(EE,I),ge,0.7,and,De1(EE,I),lt,0.8,then
MPCHG,8,EE
*elseif,De1(EE,I),ge,0.8,and,De1(EE,I),lt,0.9,then
MPCHG,9,EE
*elseif,De1(EE,I),ge,0.9,and,De1(EE,I),lt,0.999,then
MPCHG,10,EE
*elseif,De1(EE,I),ge,0.999,then
De1(EE,I)=1
ekill,EE
! If D>1 then kill element
RDM(EE,I)=1 ! Record element death in RDM array
*endif
DES(EE)=De1(EE,I)
! Update DES
! calculation
*enddo
array
for
mass
!--------------- Update Results Array ---------------------------------------------------esel,s,live
esel,r,type,,1
*get,NoE,elem,0,count
! Get the number of live elements
*vscfun,DESSUM,sum,DES(1) ! Sum Damage (D) for all elements
MML(I,1)=CT
MML(I,2)=Delta_t*(CT)
MML(I,3)=NoE
MML(I,4)=100*(DESSUM/J)
! Iteration
! Solution time (hr)
! Number of live elements
! % Mass loss based on D
solve
save
! *** SOLVE ***
finish
/post1
set,last
rsys,1
! Finish solution
! Enter postprocessor
! Cylindrical results co-ord system
!--------------- Update CES array ------------------------------------------------esel,s,type,,1
etable,eps,eppl,1
! Set up etable 'eps' for
*do,EE,1,J
! 1st principal plastic strain
*get,pps,elem,EE,etab,eps
! Write plastic strains to CES
CES(EE)=pps
! array
*enddo
!--------------- Update MMN array -----------------------------------------------*do,EE,1,J
*get, EMMN, elem, EE, attr, mat
MMN(EE)=EMMN
*enddo
!--------------- Get stent outer diameter ----------------------------------------
139
loss
cmsel,s,stentouter
nsort,u,x
*get,uxmax%I%,sort, ,max
alls
!--------------- Corrosion plot-----------------------------------------------------esel,s,type,,1
/cont,1,20,0,0.1,1
! Set values for contour plot
etable,CVPH,s,x
! Set sx as corrosion value
*vput,De1(1,I),elem,1,etab,CVPH
! placeholder
esel,s,live
! Select live elements
esel,r,type,,1
/efacet,1
/title,Corrosion %MML(I,2)%[hr] (Delta_t=%Delta_t%[hr])
pletab,CVPH
! Plot element table
/view,1,1,1,1
/ang,1
/auto,1
/rep,fast
/input,0plots,txt
alls
parsav, all, params%I%, ,
finish
/solu
! Save scalar parameters
antype,,restart,,,continue
parres, new, scalparams,,
parres, new, params%I%, ,
! Restart analysis
! Read scalar parameters
*do,EE,1,J
*if,RDM(EE,I),eq,1,then
ekill,EE
*endif
*enddo
! Remove killed elements
I=I+1
parsav, all, params%I%, ,
parsav,scalar,scalparams,,
! Save scalar parameters
*enddo
*mwrite,MML,filenam1,tsv,,jik,5,ItNum,1
(5F12.6)
! Write results to .tsv file
*msg,ui
**Solution completed**
finish
/out
/mail, --,emmet.galvin3@mail.dcu.ie,0plots,txt
finish
/eof
140
! Send message to GUI
! Alert user by email
! End of file
A.4 Corrosion Model Convergence Studies
A.4.1 Dogbone Specimen Model
Owing to the simplicity of the dogbone specimen model geometry and the
deformation induced by the uniaxial load, it was found that the predicted nodal
displacement, element stress and element strain results were all converged even
when a relatively large Le was used. Therefore, a mesh sensitivity study was
performed, as shown in Figure A.2(a), in which the predicted mass loss results were
converged to within approximately 3% when using Le = 0.012 mm. Using this Le
value, a suitable time step of 1 hour was identified in a similar iterative process, as
shown in Figure A.2(b).
Figure A.2 (a) Plot of predicted dogbone specimen mass loss over time for varying
average element size (Le) values, (b) Plot of predicted dogbone specimen mass loss
over time for varying time step (∆t) values
A.4.2 Stent Model
For the RUC stent model, predicted nodal displacement, element stress and strain
and radial stiffness results were all found to have converged within 2% when using
141
Le = 0.015. Therefore, based on the results of the convergence study shown in
Figure A.3, Le and ∆t values of 0.012 mm and 1 hour were selected, respectively.
Figure A.3 (a) Plot of predicted stent mass loss over time for varying average
element size (Le) values, (b) Plot of predicted stent mass loss over time for varying
time step (∆t) values
142